Description
1. Introduction
The intravenous safety catheter with wings and injection port Nipro SafeTouch 18G x 32 mm (ref. NIC-SF-1832PWP-ET) is the solution your nursing service needs to achieve fast, reliable and fully protected peripheral vascular access. It combines the comfort of fixation wings, the versatility of a Y-injection port and a passive safety mechanism that protects the professional against accidental needlestick injuries, one of the most frequent occupational risks in the healthcare environment. Supplied in a box of 50 sterile units, it is the usual choice in hospitalisation, emergency departments, outpatient clinics and transfusion services when a large gauge and high flow rate are required.
2. Technical Explanation
This peripheral safety catheter incorporates a passive and irreversible activation system: when the needle is withdrawn after venipuncture, the stainless steel protector advances automatically and covers the tip, with the lock confirmed by an audible click, without requiring any additional manoeuvre from the professional and without changing the puncture technique. The flexible wings allow a firm grip and precise control during insertion, as well as facilitating fixation of the catheter to the patient’s limb once cannulated. The Y-injection port with universal luer connector allows administration of medication boluses or flushing of the system without disconnecting the main line, reducing unnecessary manipulations and the risk of contamination. The thermosensitive and radiopaque polyurethane cannula softens at body temperature, adapting to the venous bed, minimising kinking and maintaining a guaranteed minimum flow rate of 81 ml/min. The stainless steel needle with triple bevel and silicone coating reduces penetration force and patient discomfort. It is manufactured latex-free, DEHP-free and PVC-free.
3. Technical Specifications
| Manufacturer reference | NIC-SF-1832PWP-ET |
| Brand / line | Nipro SafeTouch, catheter with wings and injection port |
| Gauge | 18G (green) |
| Size (diameter x length) | 1.3 x 32 mm (1 1/4″) |
| Minimum flow rate | ≥ 81 ml/min |
| Cannula material | Thermosensitive, radiopaque polyurethane (PUR) |
| Needle material | Stainless steel SUS-304, triple bevel, silicone-coated |
| Safety system | Passive and irreversible, with audible click lock (SafeTouch) |
| Design | Colour-coded flexible wings + Y injection port with universal Luer connector |
| Sterility | Sterile, single-use |
| Composition | Latex-free, DEHP and PVC-free |
| Presentation | Box of 50 units |
| Indications | Intravenous infusion, injection and transfusion |
| Regulatory classification | CE Class IIa medical device (Rule 7, MDD 93/42/EEC); compliant with ISO 10555-5, ISO 7864 and ISO 23908 |
Full range available in the same manufacturer’s line (catheter with wings and injection port), with references verified in the official technical data sheet:
| Reference | Gauge / Size | Minimum flow rate |
| NIC-SF-1650PWP-ET | 16G x 2″ (1.7 x 50 mm) | ≥ 162 ml/min |
| NIC-SF-1832PWP-ET | 18G x 1 1/4″ (1.3 x 32 mm) | ≥ 81 ml/min |
| NIC-SF-1850PWP-ET | 18G x 2″ (1.3 x 50 mm) | ≥ 81 ml/min |
| NIC-SF-2032PWP-ET | 20G x 1 1/4″ (1.1 x 32 mm) | ≥ 49 ml/min |
| NIC-SF-2050PWP-ET | 20G x 2″ (1.1 x 50 mm) | ≥ 49 ml/min |
| NIC-SF-2225PWP-ET | 22G x 1″ (0.8 x 25 mm) | ≥ 29 ml/min |
| NIC-SF-2232PWP-ET | 22G x 1 1/4″ (0.8 x 32 mm) | ≥ 29 ml/min |
| NIC-SF-2419PWP-ET | 24G x 3/4″ (0.7 x 19 mm) | ≥ 17 ml/min |
| NIC-SF-2425PWP-ET | 24G x 1″ (0.7 x 25 mm) | ≥ 17 ml/min |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references this product with specific codes from other brands; consult the technical department before use in tenders. At category level, the product is comparable with peripheral safety catheters with injection port and wings from B. Braun (Introcan Safety line), BD (Nexiva and Insyte Autoguard lines) and Vygon (Leader-Cath); before substituting in an internal protocol, the professional must verify in each brand’s technical data sheet: equivalent gauge, cannula length, material (polyurethane versus other polymers), type of safety mechanism (active/passive), presence of injection port and presentation per box.
5. Selection tips and variants of interest
The 18G gauge with 32 mm length is an ideal choice for medium-duration access in adult patients with medium-calibre veins or for situations requiring a high flow rate (≥ 81 ml/min): surgery, emergency departments, administration of blood and blood products, or high-viscosity fluids. For paediatric patients or those with fragile veins, a smaller gauge is recommended (22G x 25 mm, ref. NIC-SF-2225PWP-ET, or 24G from the same line). For very high-flow transfusions, choose the 16G x 50 mm (ref. NIC-SF-1650PWP-ET, flow rate 162 ml/min). If your protocol requires an injection port, always check that the code includes the PWP suffix (Y-port), as the SafeTouch line offers variants without a port (P/PW suffix) that are not interchangeable 1:1. Always purchase presentations of 50 units to ensure correct stock rotation and availability in hospitalisation, emergency and operating theatre services.




Reviews
There are no reviews yet.