Description
1. Introduction
The LN2 back-up system (ref. FIO-ACC-2797) is the original accessory from Fiocchetti Scientific S.r.l. designed for the ULTRAFREEZER series of ultra-low temperature freezers. It is the ideal solution for laboratories, biobanks, hospitals and the pharmaceutical industry that need to guarantee the integrity of their temperature-sensitive samples even in the event of a power cut or a failure of the refrigeration system.
This accessory is supplied as a factory pre-installation for connection to liquid nitrogen (LN2) cylinders. Fiocchetti configures each unit to order, so the product is sold on quotation: our technical-commercial team will advise you on correctly sizing the assembly according to your installation.
2. Technical explanation
In ultra-low temperature applications, loss of cold due to a power failure can irreversibly compromise the quality of stored samples. The LN2 back-up system acts as an autonomous emergency system: when a drop in the power supply is detected (or a temperature deviation outside the set limits), the system releases liquid nitrogen into the compartment, harnessing the cryogenic effect of LN2 (≈ -196 °C) to maintain the temperature inside the compartment at around -60 °C for as long as the LN2 cylinder supplies gas.
The unit includes the pre-installation for connection of LN2 cylinders (cylinders not included) and is integrated into the original ultra-low temperature freezer, so that activation occurs automatically and without staff intervention. This back-up is especially recommended in facilities with critical storage, cold chain traceability protocols, areas with an unstable power grid or contingency procedures during unattended hours.
3. Technical specifications and compatibility
| Brand / Manufacturer | Fiocchetti / Fiocchetti Scientific S.r.l. |
| Reference (SKU Cahnos) | FIO-ACC-2797 |
| Accessory type | Emergency cryogenic back-up system (factory configuration) |
| Cryogenic agent | Liquid nitrogen (LN2) |
| Actuation mode | Automatic on power failure; maintains ≈ -60 °C until cylinder is exhausted |
| Includes | Pre-installation for connection to LN2 cylinders (cylinders not included) |
| Compatible series | ULTRAFREEZER (vertical and horizontal configurations) |
| Supply method | Custom-configured on the equipment; sold on quotation |
Compatible models of the ULTRAFREEZER series
| Model | Configuration | Manufacturer reference |
| ULTRAFREEZER 120 | Vertical | ULV0121 |
| ULTRAFREEZER 240 | Vertical | ULV0241 |
| ULTRAFREEZER 360 | Vertical | ULV0361 |
| ULTRAFREEZER 400 | Horizontal | ULH0401 |
| ULTRAFREEZER 480 | Vertical | ULV0481 |
| ULTRAFREEZER 570 | Horizontal | ULH0571 |
| ULTRAFREEZER 600 | Vertical | ULV0601 |
| ULTRAFREEZER 710 | Vertical | ULV0711 |
| ULTRAFREEZER 820 | Vertical | ULV0821 |
| ULTRAFREEZER 1000 | Vertical | ULV01001 |
4. Equivalences and Cross-References
The product is identified in the Cahnos catalogue with the reference FIO-ACC-2797 and corresponds to the LN2 back-up system in the manufacturer’s catalogue (fiocchetti.com). No direct 1:1 equivalences with cryogenic back-up systems from other ultra-low temperature freezer manufacturers (PHCbi, Thermo Scientific, B Medical Systems, etc.) have been verified in reliable sources: it is a proprietary accessory of internal configuration for Fiocchetti equipment, comparable only at the level of functional category and not interchangeable without technical validation. To guarantee full compatibility with your equipment, the original Fiocchetti accessory, configured together with the corresponding ultra-low temperature freezer, must be used.
Within the Fiocchetti accessories ecosystem for the ULTRAFREEZER series available at Cahnos, this system is complemented by the 220V/50-60 Hz power supply version (FIO-ACC-2783) and the PT100 catheter with free contacts (FIO-ACC-2796) for external temperature monitoring. As a gas back-up alternative, Fiocchetti also offers the CO2 back-up system.
5. Selection tips and variants of interest
Before configuring your order, confirm that your unit appears in the table of compatible ULTRAFREEZER models (120, 240, 360, 400, 480, 570, 600, 710, 820 and 1000). The LN2 back-up system is especially recommended when your protocol requires maintaining the temperature below -60 °C during a prolonged outage, when the power grid is unstable, or in installations with critical storage during unattended hours. Note that liquid nitrogen cylinders are not included and must be provided for in the laboratory gas supply.
To complete the protection of your installation, consider combining this system with the PT100 catheter (FIO-ACC-2796), the 220V/50-60 Hz power supply option (FIO-ACC-2783) or the CO2 back-up system. Contact the technical-commercial team at Cahnos indicating the exact model of your ultra-low temperature freezer to receive a custom-configured proposal and a personalised quotation, without obligation.