Description
1. Commercial Introduction
The silent air compressor is the heart of the 285xx range of pressure-relieving mattress systems: it ensures precise pressure regulation and discreet, silent operation, ideal for hospitals, care homes, clinics and home care. It is available in two references (28565 and 28568) to suit both air bubble mattresses and mattresses with interchangeable cells or alternating pressure, and can be hung from any type of hospital bed frame using two adjustable hooks. It is a class I medical device, which: consult your technical-commercial advisor at Cahnos for terms and delivery times. For large orders you can request a quote.
2. Technical explanation of the equipment
This compressor is a special silent-running pump designed for the prevention and treatment of pressure ulcers (bedsores). Pressure is regulated simply and precisely using a control on the front panel, allowing the inflation level of the surface to be adjusted to the patient’s needs. The unit can be hung from the hospital bed frame thanks to its two adjustable hooks and is compatible with bubble mattresses (bubble mattress) or alternating pressure cell mattresses. The cyclic inflation and deflation of the cells periodically redistributes the contact points between the patient and the support surface, promoting blood circulation, relieving load-bearing areas and reducing prolonged compression of the tissues. There are two versions differentiated by their cycle time: 28565 (6-minute cycle, for the 28564 bubble mattress) and 28568 (12-minute cycle, for the 28567 interchangeable cell mattress).
3. Technical Specifications
GM28565 · Compressor for bubble mattress (Ref. 28564)
| Measurements | 25 x 12 x 10 cm |
|---|---|
| Operating voltage | 110-220 V, 50-60 Hz |
| Working pressure | 40-100 mmHg |
| Air flow rate | 6-7 litres/min |
| Cycle time | 6 min |
| Housing | TUV-CE IEC/EN 60601-1; 60601-1-2 |
GM28568 · Compressor for interchangeable cell mattress (Ref. 28567)
| Measurements | 25 x 12 x 10 cm |
|---|---|
| Operating voltage | 110-220 V, 50-60 Hz |
| Working pressure | 40-100 mmHg |
| Air flow rate | 6-7 litres/min |
| Cycle time | 12 min |
| Housing | TUV-CE IEC/EN 60601-1; 60601-1-2 |
Available references
| Reference | Model | Sales unit |
|---|---|---|
| GM28565 | 28564 | 1 pc |
| GM28568 | 28567 | 1 pc |
Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | 1 pc |
| Device type | Medical device (Medical device) |
| Device class | I |
| CND | V080701 |
4. Equivalences and Cross-References
References 28565 and 28568 correspond to the manufacturer’s official codes for the compressor/air pump systems in the pressure-relieving surface range, without transcriptions or normalisations. Reference 28565 is the compressor for the 28566 air bubble mattress kit (made up of 28564 + 28565), while reference 28568 is the compressor for the 28569 interchangeable cell mattress kit (made up of 28567 + 28568). No validated 1:1 equivalence table cross-referencing these references with specific codes from other manufacturers of compressors for pressure-relieving surfaces has been found in reliable sources; any substitution in tenders or protocols must be validated field by field (measurements, voltage, working pressure, flow rate, cycle time and regulatory classification) with the technical-commercial department at Cahnos. Within the Cahnos catalogue there are complementary products from the same range, comparable by category but not interchangeable 1:1 with this compressor: the 28566 kit (bubbles, 200 x 90 x 7 cm, load 135 kg), the 28569 kit (20+1 cells, 200 x 86 x 9.5 cm, load 140 kg) and the 5010E alternating pressure kit (Ref. 28575, 198 x 86 x 12.8 cm, load 150 kg).
5. Selection tips and variants of interest
Select the compressor according to the pressure-relieving surface you are going to equip: Ref. 28565 (6-min cycle) is intended for the air bubble mattress ref. 28564, indicated for stage I in short-term and home therapies; Ref. 28568 (12-min cycle) corresponds to the interchangeable cell mattress ref. 28567, recommended for the prevention and treatment of pressure ulcers in stages I and II in medium-term therapies. Connect the compressor to the 110-220 V mains, adjust the pressure using the front control respecting the working range of 40-100 mmHg and suspend it from the bed frame using the two adjustable hooks. Verify the compatibility of the mattress and that the patient’s weight does not exceed the maximum load of the associated mattress (135 kg for the bubble model, 140 kg for the cell model). For complete fittings, hospitalisation projects or supply contracts, contact the technical-commercial team at Cahnos for special conditions and a formalised quote.







Reviews