Description
1. Introduction
The rechargeable linear cutting stapler WZCTS-A-80M is a mechanical suturing instrument designed for open surgery and demanding resection procedures. Its double-row titanium staple system with integrated cutting allows the surgeon to suture and divide tissue in a single action, reducing operating theatre time and improving anastomotic safety.
It is supplied sterile and ready to use, with an included load of 80 staples (variant M, 3.8 mm) indicated for medium-thick tissue. Its reloadable design allows the handle to be used with up to five loads in the same procedure, making it a very competitive cost-per-procedure option for hospitals, clinics and purchasing centres.
2. Technical explanation: how it works
The stapler places two interlocking rows of titanium staples on each side of the cutting line (four rows in total) while an integrated blade divides the tissue between both double rows simultaneously. The result is an anastomosis or incision closure whose stapled line is approximately 50% longer than the cutting line, a safety margin that reduces tension at the ends and minimises the risk of anastomotic leakage.
The main anvil beam is made of reinforced stainless steel, ensuring uniform distribution of compression force across the entire suture line and consistent staple formation even in thick tissue. The reloading system incorporates a positioning device at the distal end and an acoustic signal that confirms correct seating of the load before firing.
The M variant (3.8 mm staple height) is especially indicated for medium and medium-thick tissues: stomach, intestine and parenchyma. Typical applications: creation of the incision closure site, reconstruction of the digestive tract and resection procedures in general, thoracic, colorectal, bariatric, urology and hepatobiliary surgery.
3. Technical Specifications
3.1. Features of this stapler (WZCTS-A-80M)
| Factory reference | WZCTS-A-80 (variant M) |
| SKU Cahnos | WZCTS-A-80M |
| Manufacture | Own (Cahnos) |
| Staple count | 80 units |
| Staple diameter | 0.20 mm (φ 0.2 ± 0.02 mm) |
| Staple height | 3.8 mm (± 0.2 mm) — M variant (medium-thick tissue) |
| Staple material | Titanium |
| Material of the main anvil beam | Reinforced stainless steel |
| Staple arrangement | Double interlocking line on each side of the cutting line |
| System | Simultaneous cutting and suturing |
| Loading system | Reloadable, up to 5 reloads in the same surgery |
| Compatible cartridge | WZCTS-P1-80M |
| Approx. total device length | 280 mm (± 4.0 mm) |
| Reload length | 81 mm (± 2.0 mm) |
| Use | Single-use cartridge; handle reusable in the same procedure |
| Sterilisation | Sterilised by radiation (sterile) |
| Certifications | CE, ISO 13485, SGS; AEMPS registration (Spain) |
| Applications | Digestive, abdominal, colorectal, bariatric, thoracic and hepatobiliary surgery |
3.2. WZCTS-A range (manufacturer’s catalogue data)
| Model | Number of staples | Diameter (mm) | Height (mm) | Approx. total length (mm) |
|---|---|---|---|---|
| WZCTS-A-60 L / M / H | 60 | 0.2 ± 0.02 | 2.5 / 3.8 / 4.2 ± 0.2 | 265 ± 4.0 |
| WZCTS-A-80 L / M / H (this product: variant M) | 80 | 0.2 ± 0.02 | 3.8 (M variant) | 280 |
| WZCTS-A-100 L / M / H | 100 | 0.2 ± 0.02 | 2.5 / 3.8 / 4.2 ± 0.2 | 323 ± 4.0 |
Note: the staple height indicated by the manufacturer in its official table is 3.8 mm (± 0.2 mm) for the M model. The replacement load corresponding to this unit is reference WZCTS-P1-80M.
3.3. Compatible spare cartridge family
| Spare part model | Length (± 2.0 mm) | Staple height (± 0.2 mm) L/M/H | Diameter (± 0.02 mm) | Staple count |
|---|---|---|---|---|
| WZCTS-P1-60 L / M / H | 61 mm | 2.5 / 3.8 / 4.2 | φ 0.2 | 60 |
| WZCTS-P1-80 L / M / H (reload for this device) | 81 mm | 3.8 (M variant) | φ 0.2 | 80 |
| WZCTS-P1-100 L / M / H | 101 mm | 2.5 / 3.8 / 4.2 | φ 0.2 | 100 |
4. Equivalences and Cross-References
The code WZCTS-A-80 (variant M) has been verified against documentation from the manufacturer and authorised distributors. It is confirmed as an identical official reference to the factory reference; the Cahnos SKU WZCTS-A-80M corresponds to that official manufacturer reference.
| Manufacturer / Brand | Family / Model | Reference | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Catalogue Cahnos | Reloadable linear cutting stapler | WZCTS-A-80 (variant M) | Identical (same factory reference) | Same staple geometry 0.2 × 3.8 mm and 80 staples. SKU Cahnos: WZCTS-A-80M. |
| Catalogue Cahnos | Reloadable linear cutting stapler, height 4.2 mm | WZCTS-A-80H | Direct equivalent (same family) | Same capacity (80 staples) and diameter; the staple height changes to 4.2 mm (very thick tissue). |
| Catalogue Cahnos | Compatible reload | WZCTS-P1-80M | Direct equivalent (official replacement) | Single-use cartridge with the same 80 staples 0.2 × 3.8 mm for the WZCTS-A-80 handle. |
| Medtronic (Covidien) | Endo GIA™ Tri-Staple™ | e.g. EGIA60AMT (60 mm, medium/thick) | Comparable by category, not interchangeable 1:1 | Same clinical logic for medium-thick tissue, but different anchoring system and anvil; it does not fit the WZCTS handle. 60 mm platform: the number of staples per load differs. |
| Johnson & Johnson (Ethicon) | ECHELON FLEX™ / ENDOPATH™ GST | e.g. GST60D (60 mm, gold, 3.8 mm) | Comparable by category, not interchangeable 1:1 | Staple height 3.8 mm equivalent to the M variant; proprietary platform not compatible with the WZCTS series. |
| Purple Surgical | Ultimate Reloadable Linear Cutter | According to manufacturer’s catalogue | Comparable by category, not interchangeable 1:1 | Same category (rechargeable linear cutter with staggered double titanium rows and simultaneous cutting). Verify staple heights and load system. |
Important notice: there is no inter-brand equivalence table validated by the manufacturer that cross-references this reference with other brands. Before replacing this reference with another brand in a tender or substitution, consult the technical department and verify the mechanical compatibility of loads and geometry.
5. Selection tips and variants of interest
- Staple height by tissue: L variant (2.5 mm) for fine tissue (lung, vascular); M variant (3.8 mm) for medium-thick tissue (stomach, intestine, parenchyma) — this unit is the M variant; H variant (4.2 mm) for very thick tissues (bariatric and colorectal surgery).
- Line length: the number of staples (60 / 80 / 100) determines the length of the suture line. This 80-staple unit is indicated for medium-long lines and extensive resections.
- Use and reloading: each load is single-use; the reloadable handle accepts up to 5 changes in the same procedure. Always verify the integrity of the sterile packaging before each use.
- Compatibility: use only loads from the WZCTS-A series (60/80/100), such as the replacement load WZCTS-P1-80M. Do not use loads from other brands without prior technical validation.
- Variants at Cahnos: same family in other capacities and heights: WZCTS-A-80H (4.2 mm), WZCTS-A-60M, WZCTS-A-100M and WZCTS-A-100H. Replacement loads available: WZCTS-P1-60M, WZCTS-P1-60H, WZCTS-P1-80H, WZCTS-P1-100M and WZCTS-P1-100H, to cover all operating theatre needs.
To equip a surgical department, we recommend combining this stapler with its replacement loads WZCTS-P1-80M and planning stock according to the volume of firings per procedure. Sterile product, with single-use load; verify the integrity of the sterile packaging before each use.