Description
1. Commercial Introduction
The disposable linear surgical stapler WZLS-A-30H is the mechanical suture solution that your healthcare centre needs for open surgery procedures on the digestive tract, resection of viscera and closure of incisions. Its single-use design eliminates reprocessing costs and guarantees a new, sterile instrument with full stapling capacity in each intervention.
Manufactured with high-strength titanium staples, it offers a firm and homogeneous seal that reduces postoperative bleeding and shortens suture times compared with manual technique. Its value for money, ergonomics and traceable reference make it a common choice in general and digestive surgery operating theatres, with a cost per procedure significantly lower than that of the major houses.
2. Technical explanation of the equipment
This device applies two interlocking and staggered lines of titanium staples to the tissue, fixing both edges with a homogeneous distribution of tension and effective haemostatic sealing. The staggered staple configuration (not opposed) reduces point pressure on the wall and minimises the risk of postoperative bleeding and anastomotic leakage.
The stapler combines manual and automatic actuation with audible signals that confirm correct closure of the anvil and release of the load: the surgeon perceives when stapling has been completed, avoiding incomplete firings. The narrow-design anvil reduces the footprint on healthy peripheral tissue, favouring preservation of local blood supply and cleaner healing.
The green 4.5 mm load is especially indicated for thick or oedematous tissue: thickened gastric or intestinal wall, pulmonary parenchyma, closure of stumps in resections and side-to-side anastomoses. Being single-use, the risk of cross-infection is eliminated and the sharpness of the mechanism is ensured in each application.
3. Technical Specifications
| Specification | Value |
| Model / reference | WZLS-A-30H |
| Device type | Disposable linear stapler (open surgery, non-cutting) |
| Staple line length | 30 mm |
| Staple wire diameter | 0.28 mm |
| Staple height | 4.5 mm |
| Load identification colour | Green (thick tissue) |
| Number of staples per unit | 13 |
| Staple rows | 2 staggered (interlocked) rows of titanium |
| Staple material | Titanium (imported wire) |
| Anvil | Narrow design, protects peripheral tissue |
| Actuation | Manual and automatic with audible signals |
| Use | Single-use (disposable), single fire |
| Sterilisation | Ethylene oxide (EO), individually packaged |
| Composition | Latex-free and DEHP-free |
| Expiry | 3 years from the date of manufacture |
| Presentation | Sterile unit in individual blister |
| Application | Closure and suture of incisions, resection of viscera and reconstructive surgery of the digestive tract |
Table of the WZLS-A family (factory references): the series combines four stapling line lengths (30, 45, 60 and 90 mm) with two staple heights (3.8 mm blue for standard tissue and 4.5 mm green for thick tissue), allowing the instrument to be adjusted to each surgical requirement.
| Model | Length (mm) | Staple height (mm) | Colour | No. of staples |
| WZLS-A-30L | 30 | 3.8 | Blue | 13 |
| WZLS-A-30H (this model) | 30 | 4.5 | Green | 13 |
| WZLS-A-45L | 45 | 3.8 | Blue | 19 |
| WZLS-A-45H | 45 | 4.5 | Green | 19 |
| WZLS-A-60L | 60 | 3.8 | Blue | 25 |
| WZLS-A-60H | 60 | 4.5 | Green | 25 |
| WZLS-A-90L | 90 | 3.8 | Blue | 39 |
| WZLS-A-90H | 90 | 4.5 | Green | 39 |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references the WZLS-A series with the references of the major Western houses. The code WZLS-A-30H corresponds to the manufacturer’s factory nomenclature; it has not been found in independent public catalogues, so its traceability must be confirmed with the technical department before its use in specifications or tenders.
The following references are offered as a functional category comparison verified against each manufacturer’s sources, and in no case as 1:1 interchangeable (dimensions, material, handle system and regulatory classification vary):
| Manufacturer / brand | Family / model | Reference | Equivalence level | Technical differences / observations |
| Haida Medical (Changzhou) | HSN disposable linear stapler | HSN-30L | Comparable by category (not interchangeable 1:1) | Same measurement configuration (30 mm, 13 titanium staples) and disposable concept; without official cross-validation |
| Medtronic (Covidien) | DST Series™ TA™ 30 | TA3035S (instrument) / TA3035L (reload) | Comparable by category (not interchangeable 1:1) | 30 mm, double row, titanium; blue load 3.5 mm versus green 4.5 mm; reloadable (up to 7 reloads) versus disposable; different number of staples |
| Ethicon (Johnson & Johnson MedTech) | PROXIMATE™ Reloadable Linear TX | XR30B / TX30B | Comparable by category (not interchangeable 1:1) | 30 mm reloadable; wire 0.23 mm versus 0.28 mm; B-Form technology and higher price segment; confirm green reload height with the manufacturer |
| Volkmann Med | Reloadable Linear Stapler | 30–90 mm series (code to be confirmed) | Comparable by category (not interchangeable 1:1) | Reloadable (not disposable); the standard load heights 3.8 mm and thick tissue 4.5 mm match the WZLS-A series |
| Purple Surgical | Ultimate Linear Stapler | 30–90 mm series (code to be confirmed) | Comparable by category (not interchangeable 1:1) | Reloadable up to 8 firings; typical reloads of 3.5 mm and 4.8 mm that do not exactly match 3.8/4.5 mm |
There is no equivalence table validated by the manufacturer for this series; consult the technical department before using it as a substitute in tenders or closed purchasing protocols.
5. Selection tips and variants of interest
- Choice of staple height: the green 4.5 mm load (this model) is designed for thick, oedematous or parenchymatous tissue (thickened gastric or intestinal wall, pulmonary parenchyma). For standard or thin-walled tissue select the blue 3.8 mm load (WZLS-A-30L).
- Choice of suture line length: the 30 mm model is used for short closures, pedicles and areas of limited access; the 45 and 60 mm versions cover anastomoses and medium transections, and the 90 mm version is reserved for wide transections of parenchyma or stomach.
- Prior checks: confirm the wall thickness with the compression instrument before firing, check the integrity of the sterile packaging and the expiry date, and do not reuse or re-sterilise the device.
- Range variants: if your department requires stapling with simultaneous cutting, consult the family of reloadable linear cutting staplers WZCTS-A (60/80/100 mm) and their spare parts; for laparoscopic approaches, the range of endoscopic linear staplers and articulated spare loads.
For tenders: since there is no official cross-reference table from the manufacturer, validate each equivalence with our technical department and provide the technical data sheets of the comparables before submitting a bid.