Description
1. Commercial Introduction
Safety in a Magnetic Resonance Imaging (MRI) service is non-negotiable. The MRI test magnet CAHIRM8002 is the most widely used manual screening tool at access control points for zones III and IV: a powerful portable neodymium (rare-earth) magnet that allows you to reliably check whether a patient, companion, professional or device has ferromagnetic properties before approaching the scanner’s magnetic field. Lightweight, robust, intuitive to handle and supplied with a padded case as standard, it is the perfect ally for radiologists, senior diagnostic imaging technicians (TSID), nursing staff and facility safety officers who want to prevent the well-known projectile effect and ensure the safety of people and equipment.
2. Technical explanation of the equipment
This test magnet is made of neodymium, a rare-earth material with extremely high energy density capable of generating a surface magnetic field greater than 3,200 Gauss (0.32 T), sufficient to detect even weakly ferromagnetic objects that would go unnoticed in a visual inspection. At about 2 cm from the surface, the intensity remains around 1,200 Gauss, allowing effective verification without direct contact. Its ergonomic design, with a grip handle and rigid high-visibility red housing with silver warning lettering, ensures comfortable and safe handling during prolonged checks. The magnet is securely housed in a casing with bearings at the base that allow it to be easily released if it accidentally becomes attached to a large metal object (stretcher, oxygen tank, wheelchair, etc.). Neodymium retains its magnetic strength for decades (estimated loss of approximately 1% every 100 years), making it a long-lasting investment. Important: it is a pre-screening device, it must never be taken into the scanning room (zone IV) or used on patients during the scan.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Reference (SKU) | CAHIRM8002 |
| Product type | Portable test magnet for ferromagnetic screening in MRI |
| Magnetic material | Neodymium (rare earth) |
| Magnetic field strength | Greater than 3,200 Gauss (0.32 T) at surface |
| Strength at 2 cm from the surface | Approximately 1,200 Gauss |
| Intended application | Detection of ferromagnetic properties in objects, devices, auxiliary equipment and belongings before access to MRI zones III and IV |
| Design | High-visibility red housing with silver warning lettering, ergonomic handle and release bearings at the base |
| Approximate dimensions | 14 x 7.6 x 8.9 cm (height x width x length) |
| Weight | Approx. 0.9 kg |
| Included accessory | Padded storage and transport case |
| Magnetic durability | Loss of strength of approx. 1% every 100 years |
| Safety | Not suitable for use inside the MRI room (zone IV) |
| Presentation | Unit |
4. Equivalences and Cross-References
This test magnet functionally corresponds to the industry standard reference SM-100 (manufacturer MRImed / Z&Z Medical) and to the 20-1212 from Steiner Enterprises, the family of rare-earth test magnets for MRI safety screening most widely used on the international market, developed in collaboration with Dr. Emanuel Kanal. In the Cahnos catalogue, the CAHIRM8002 test magnet is part of the MRI safety accessories family and is complemented by the following products:
| Reference | Product | Related items |
|---|---|---|
| CAHIRM8002 | MRI Test Magnet 3,200 Gauss | Reference product (this item) |
| CAHIRM6689 | Portable high-sensitivity metal detector | Complementary: contactless scanning at the entrance |
| CAHIRM8010 | MR-safe ferromagnetic metal detector for zone 4 | Complementary: verification inside the MRI room |
| CAHIRM8022 | Portable Sensitive metal detector for MRI | Functional alternative with digital technology and docking station |
| R-MN-01 | Metal detector for MRI room | Functional alternative with high-capacity rechargeable batteries |
5. Selection tips and variants of interest
To set up a comprehensive safety protocol in your Magnetic Resonance service, consider the following recommendations: control point – place the test magnet at the entrance to zone III, before the person or device approaches the magnetic field, and combine it with an access sweep detector (CAHIRM6689) for a quick contactless check; individual check – use the test magnet for specific objects (implants with questionable documentation, dressing material, keys, coins, auxiliary equipment); in-room reinforcement – if your centre needs to check within the room itself (zone 4), supplement it with the MR-safe detector CAHIRM8010; digital technology – for services that prefer fully digital systems with a charging station, consider reference CAHIRM8022 or R-MN-01; storage – always keep the magnet in its padded case when not in use and periodically check its attractive force with a known reference object. Our technical-commercial team at Cahnos can advise you on the comprehensive safety provision for your Magnetic Resonance service and on standardising references with other manufacturers.





Reviews
There are no reviews yet.