Description
1. Complete percutaneous tracheostomy kit with subglottic suction
The percutaneous dilation kit Ref. TP90-38 is the complete solution for establishing tracheal access by puncture and progressive dilation in critically ill adult patients, following the Ciaglia single dilator technique. It incorporates a tracheostomy tube of 8 mm with fixed flange and specific subglottic suction lumen, a feature especially valued in intensive care units for the prevention of ventilator-associated pneumonia (VAP).
All the material needed for the procedure is presented in a single sterile, ready-to-use kit, which reduces preparation times, minimises the risk of missing items and facilitates standardisation of the process in ICU, emergency and operating theatre. A single order, a single code: the guarantee that the team has all the instruments with full traceability.
2. Technical explanation of the procedure
Percutaneous dilation tracheostomy using the Ciaglia technique is performed by tracheal puncture with an introducer needle, passage of a guidewire and progressive dilation of the stoma with a single conical-profile 38 Fr dilator, which allows the access to be created in a single step. This design minimises tissue trauma compared with open surgical techniques and favours a faster, safer bedside procedure, without the need to move the critically ill patient.
The 8 mm tracheostomy tube incorporates an additional lumen dedicated to subglottic suction, which allows aspiration of secretions accumulated above the cuff, reducing the risk of micro-aspiration and ventilation-associated respiratory infections. The fixed flange facilitates stable and predictable fixation of the tube, while the introducer with locking system maintains position during insertion until correct placement in the trachea.
3. Technical specifications and kit contents
Kit contents
| Component | Specification |
|---|---|
| Tracheotomy cannula | 8 mm tube with fixed flange and subglottic suction lumen |
| Dilator | 38 Fr, conical profile for single-step dilation |
| Tube introducer | With locking system and conical tip |
| Guidewire | J-tip nitinol 0.05″ (0.89 mm), kink-resistant |
| Guide catheter | Included for guiding the dilator |
| Scalpel | Safety No. 15 with retractable protection |
| Syringe | 10 ml |
| Introducer needle | 16 G |
| Fixation band | With Velcro closure |
| Table cover | For instruments |
| Sterilisation | Ethylene oxide (ETO), validated process |
Technical features of the tube
| Parameter | Value |
|---|---|
| Internal tube diameter | 8.0 mm |
| Flange type | Fixed |
| Subglottic suction | Yes, dedicated lumen |
| Connector | Universal 15 mm (15M) |
| Use | Single-use (sterile) |
| Materials | Latex-free and phthalate-free |
| Classification | Medical device class IIa |
Ciaglia kit family with fixed flange and subglottic suction (TP90-3x series)
| Reference | Tube diameter | Flange | Subglottic suction | Dilator |
|---|---|---|---|---|
| TP90-37 | 7 mm | Fixed | Yes | 38 Fr |
| TP90-38 (this product) | 8 mm | Fixed | Yes | 38 Fr |
| TP90-39 | 9 mm | Fixed | Yes | 38 Fr |
4. Equivalences and Cross-References
The Ciaglia single dilator technique with a 38 Fr dilator is a widely established standard in percutaneous tracheostomy. For comparison:
- Comparable at category level: Cook Medical Ciaglia Blue Rhino G2 (percutaneous single dilator introducer set). Same one-step dilation principle and 38 Fr dilator, but it is a set from a different manufacturer with different components and presentation, so it is not interchangeable 1:1.
- Comparable at category level: TRACOE experc (REF 520), dilation kit for percutaneous tracheotomy with nitinol Seldinger wire and hydrophilic dilator. Same clinical indication, different component configuration.
- Comparable at category level: percutaneous dilation tracheotomy kits with a single 38 Fr dilator from other manufacturers in the sector (single dilator technique).
There is no equivalence table validated by the manufacturer that allows this reference to be cross-referenced with specific codes from other suppliers; it is recommended to consult the technical department before using this datasheet in a tendering process or stock replacement.
5. Selection tips and variants of interest
To choose the appropriate kit, consider the tube diameter according to the patient and the clinical indication: the 8 mm variant (TP90-38) is the reference option for medium-build adults; the 9 mm variant (TP90-39) is reserved for medium-to-large build adults or those requiring greater ventilatory flow, and the 7 mm variant (TP90-37) for smaller adults or more limited anatomical situations.
If the patient requires prolonged mechanical ventilation, the subglottic suction of this series is especially recommended for its role in preventing ventilator-associated pneumonia. If an adjustable flange or a reinforced tube (helical anti-kink reinforcement) is preferred, consult the rest of the Ciaglia family available in the store (TP90-0x series with adjustable flange and TP90-1x reinforced), as well as the standard versions without subglottic suction (TP90-27, TP90-28 and TP90-29). All kits are single-use: dispose of them after the procedure as healthcare waste according to local regulations.






Reviews