Description
1. Complete and sterile kit for percutaneous airway access
The percutaneous dilatation kit Ref. TP02-18 is the complete single-use solution for establishing a safe airway quickly, minimally invasively and in a controlled manner in adult patients. It follows the Griggs technique with forceps over a guidewire and incorporates a reinforced 8 mm tracheostomy tube with adjustable cervical flange, a combination particularly valued in intensive care units for its structural resistance to kinking and its adaptation to variable cervical anatomies.
All necessary material is presented in a single sterile kit, ready to use: one order, one code and the assurance that the team has all the instruments with full traceability. Preparation time is reduced, the risk of forgetting material is minimised and standardisation of the procedure is facilitated in ICU, emergency departments, operating theatre and airway units.
2. Technical explanation of the equipment and procedure
Percutaneous dilatational tracheostomy (PDT) according to the Griggs technique is an evolution of the Ciaglia technique: it is performed at the bedside, usually under bronchoscopic control, without the need to transfer the critically ill patient to the operating theatre. The procedure begins with tracheal puncture using the 16 G introducer needle, continues with the passage of the J-tip guidewire (atraumatic tip that minimises tissue damage) and ends with progressive dilatation of the stoma using Howard-Kelly type dilatation forceps, which advance over the guidewire opening the orifice in a single movement, followed by advancement of the pre-mounted cannula on the introducer with locking system and conical tip, which avoids steps and maintains tube position during insertion.
The tube incorporates reinforced helical winding anti-kinking, low-pressure balloon, universal 15 mm (15M) connector compatible with the ventilator circuit, radiopaque line for radiological control and an adjustable cervical flange in transparent plastic that allows fixation to be adapted to different anatomies and correct visualisation of the stoma. All components are single-use, latex-free and phthalate-free.
3. Technical specifications and kit contents
Kit contents (Ref. TP02-18)
| Component | Specification |
|---|---|
| Tracheotomy cannula | Reinforced 8 mm tube with adjustable flange, helical anti-kinking winding |
| Dilatation forceps | Griggs type (Howard-Kelly), dilatation over guidewire |
| Tracheostomy tube introducer | Locking system and conical tip for safe and progressive insertion |
| Dilator | 14 Fr for progressive dilatation of the stoma |
| Guidewire | J-tip 0.05\”, NiTiNol core resistant to kinking |
| Guide catheter | For guiding the dilator and the tube |
| Safety scalpel | No. 15, with retractable shield and fixed blade |
| Syringe | 10 ml |
| Introducer needle | 16 G |
| Fixation band | Velcro-type closure |
| Instrument table cover | Sterile field cover |
| Stoma accessories | Tracheal compress 9 × 9 cm (three-layer foam), gauze swabs 10 × 10 cm and sterile lubricating gel |
Technical data for the tube and kit
| Parameter | Specification |
|---|---|
| Technique | Percutaneous dilatational tracheostomy, Griggs technique with forceps |
| Internal tube diameter | 8.0 mm (exact outer diameter and length according to the tube technical data sheet) |
| Tube type | Reinforced, with anti-kink helical reinforcement |
| Flange type | Adjustable, cervical, flexible, transparent plastic |
| Cuff | Low pressure |
| Connector | Universal 15 mm (15M) |
| Radiological control | Radiopaque line |
| Dilatation forceps | Included (Griggs technique) |
| Subglottic suction | No (see variants with subglottic lumen in the series) |
| Use | Single-use (disposable) |
| Sterilisation | Ethylene oxide (ETO), process validated according to EN 11135-1; desorption in accordance with ISO 10993-7 |
| Presentation | Sterile, individually packaged in TYVEK pouch in accordance with EN ISO 11607-1/2 |
| Materials | Latex-free and phthalate-free in all components |
| Regulatory classification | CE Class IIa medical device (Directive 93/42/EEC) |
| Indication | Percutaneous tracheostomy, cricothyrotomy, coniotomy or laryngotomy |
Family of Griggs kits with adjustable flange in Cahnos (TP02 series)
| Reference | Diameter | Tube type | Forceps |
|---|---|---|---|
| TP02-07 | 7 mm | Standard | Yes |
| TP02-08 | 8 mm | Standard | Yes |
| TP02-09 | 9 mm | Standard | Yes |
| TP02-17 | 7 mm | Reinforced | Yes |
| TP02-18 (this product) | 8 mm | Reinforced | Yes |
| TP02-19 | 9 mm | Reinforced | Yes |
| TP02-27 | 7 mm | Standard | No |
| TP02-28 | 8 mm | Standard | No |
| TP02-29 | 9 mm | Standard | No |
| TP02-37 | 7 mm | Reinforced | No |
| TP02-38 | 8 mm | Reinforced | No |
| TP02-39 | 9 mm | Reinforced | No |
4. Equivalences and Cross-References
The Griggs technique with dilatation forceps over a guidewire is a widely established standard in percutaneous tracheostomy. The best-known references are listed below, classified according to the level of verification achieved; none should be treated as an exact 1:1 equivalence.
Direct range equivalent (Cahnos catalogue): references TP02-17 (7 mm) and TP02-19 (9 mm) share the exact configuration with this kit (reinforced tube, adjustable flange and dilatation forceps) and only differ in diameter: they are direct range equivalents, not interchangeable 1:1 depending on the patient.
Comparable at category level, not interchangeable 1:1: Portex (ICU Medical / Smiths Medical) percutaneous kits from the GRIGGS family, such as the 100/891 series with Blue Line Ultra tube and dilatation forceps (reference 100/891/080 corresponds to the 8.0 mm size). Same functional family (Griggs technique, tube with flange and introducer), but different manufacturer, materials and component geometry.
Comparable at category level, not interchangeable 1:1: TRACOE experc kit (REF 520, TRACOE / Atos Medical) completed with flanged cannula. Same clinical indication and procedural approach, but different component configuration and dilatation mechanics (single Ciaglia-type dilator versus forceps). Also comparable at category level are Cook Medical percutaneous sets (Ciaglia and Melker families, with single Blue Rhino-type dilator) and DEAS/ACCEX Deaflux line kits with adjustable flange.
There is no manufacturer-validated equivalence table allowing this reference to be cross-matched with specific codes from other suppliers; it is recommended to consult the Cahnos technical department before using this data sheet in a tendering process or stock replacement.
5. Selection tips and variants of interest
Diameter: the 7 mm variant (TP02-17) is ideal for smaller adults or more limited anatomies; the 8 mm (TP02-18, this kit) is the usual size for medium-built adults; the 9 mm (TP02-19) is reserved for wide airways or patients with higher ventilatory flow demand.
Reinforced tube versus standard: the reinforced helical winding of this kit offers greater resistance to compression and kinking, recommended in patients with short neck, obesity, frequent mobilisation or prolonged tracheostomy. If structural reinforcement is not required, the standard variant of the same size is TP02-08.
With versus without forceps: this kit includes dilatation forceps, ideal for departments that apply the complete Griggs sequence with instrumental dilatation under endoscopic control; if single-dilator dilatation with 14 Fr is preferred, the same reinforced 8 mm configuration without forceps is available (TP02-38).
Adjustable flange: allows the fixation to be adapted to cervical oedema or variable neck thicknesses. If a fixed and simplified fixation is preferred, refer to the Griggs series with fixed flange; and if the patient requires suction of subglottic secretions to prevent ventilator-associated pneumonia, consider the series with subglottic suction lumen (TP02-64 to TP02-76). All kits in the family are single-use: verify the integrity of the sterile packaging and the expiry date before use, and dispose of them after the procedure as clinical waste in accordance with current local regulations.