Description
1. Introduction
The oral endotracheal tube with cuff, suction lumen and Murphy eye (ref. ETSL65C02) is an airway management device designed for orotracheal intubation in small-built adult patients. It is the reference solution in intensive care units, anaesthesia operating theatres and emergency departments where a sealed, stable airway with continuous suction capability is required during mechanical ventilation or general anaesthesia. It is supplied in a box of 10 units, individually packaged and sterile, ready for immediate use.
2. Technical explanation of the device
The tube is inserted orally through the glottis into the trachea and is connected at its proximal end to a ventilator, an anaesthesia circuit or a resuscitation bag via the standard 15 mm connector (in accordance with ISO 5356-1). Once inflated via the pilot balloon with Luer connection, the cuff creates an airtight seal against the tracheal wall that allows positive pressure ventilation and minimises the risk of leakage and aspiration of gastric or pharyngeal contents.
Its integrated suction lumen runs along the wall of the tube and opens into a proximal port that can be connected to a vacuum system, allowing subglottic secretions accumulated above the cuff to be aspirated without the need to extubate the patient. This function reduces bacterial colonisation of the lower respiratory tract and is a recognised measure for preventing ventilator-associated pneumonia (VAP). The Murphy eye, a lateral opening at the distal end, ensures airflow even if the tip of the tube becomes occluded by contact with the tracheal wall.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Sleeve | Yes, sealing (inflated via pilot balloon with Luer connection) |
| Suction lumen | Yes, integrated for continuous or intermittent subglottic suction |
| Murphy eye | Yes, distal lateral safety opening |
| Material | Medical-grade polyvinyl chloride (PVC), transparent |
| Internal diameter (ID) | 6.5 mm |
| Use | Orotracheal intubation, single-use, non-reusable |
| Sterility | Sterile, individually packaged (ethylene oxide) |
| Latex | Natural rubber latex-free |
| Connector | Universal 15 mm (ISO 5356-1) |
| Presentation | Box of 10 units |
ETSL range – oral endotracheal tubes with cuff and suction lumen (Cahnos references):
| Reference | ID (mm) | Indicative patient |
|---|---|---|
| ETSL60C02 | 6.0 | Paediatric / small adult |
| ETSL65C02 (this product) | 6.5 | Small adult |
| ETSL70C02 | 7.0 | Adult |
| ETSL75C02 | 7.5 | Adult |
| ETSL80C02 | 8.0 | Adult |
| ETSL85C02 | 8.5 | Adult / large adult |
| ETSL90C02 | 9.0 | Large adult |
4. Equivalences and Cross-References
This tube belongs to the family of endotracheal tubes with subglottic suction lumen, a category represented by manufacturers such as Teleflex (Rüsch® Guardian™ with evacuation lumen), Medtronic (Shiley™ Evac with TaperGuard™) and ICU Medical (Portex SACETT™ / Blue Line™). All share the same general clinical indication (drainage of secretions above the cuff to reduce VAP), but they differ in cuff material (ultra-thin polyurethane versus PVC), geometry and connection system, so this is a functional category equivalence, not interchangeable 1:1.
The reference ETSL65C02 is a code specific to the Cahnos range and could not be verified against an official third-party manufacturer code in public sources: there is no equivalence table validated by the manufacturer; consult the technical department before use in tenders or as a stock replacement.
Comparable by category within the Cahnos catalogue in adjacent sizes: ETSL60C02 and ETSL70C02 (same range, same configuration, different ID). In size 7.0 mm and for reference from other lines: ETAN70C02 (with cuff and Murphy eye, without suction lumen), ETPN70C02 (nasal preformed with cuff and Murphy eye), ETA070C02A2 (reinforced with cuff) and ETAM70C02A1 (reinforced with cuff and guide); all comparable by category but not equivalent 1:1.
5. Selection tips and variants of interest
The 6.5 mm ID is reserved for small-built adult patients or adolescents; as a general guide, tubes of 7.0–9.0 mm are used in adults depending on build, always selecting the largest size compatible with the patient’s anatomy. Before intubation, check the integrity of the cuff, the patency of the suction lumen and the pilot balloon, and control the sealing pressure (usual recommended cuff pressure: 20–30 cmH₂O; common market practice, verify the centre’s protocol).
If your centre works with prolonged intubations or patients at high risk of VAP, the ETSL range with suction lumen is the indicated option. For procedures requiring resistance to kinking or frequent repositioning, consider the reinforced tubes from the ETA series. For head and neck surgery or nasal route, consider the preformed tubes from the ETPN series. Contact the Cahnos technical department for samples, volume conditions and hospital supply lead times.