Description
1. Safe ventilation in anaesthesia and paediatric care
The inflatable laryngeal mask size 1.5 is the supraglottic device indicated to secure the airway quickly, atraumatically and reliably in paediatric patients of small build (approximately 5 to 10 kg). Its anatomical and translucent structure facilitates visual monitoring of the airway during the procedure and adapts comfortably to the child’s anatomy.
It is the reference solution for anaesthesia services, paediatric intensive care units, emergency departments and medical transport that need to guarantee spontaneous or controlled ventilation with a disposable, sterile and ready-to-use product. It is supplied in boxes of 10 units for the continuous supply of the service.
2. What a laryngeal mask is and how it works
Unlike endotracheal intubation, the laryngeal mask is an extraglottic device that rests on the hypopharynx and seals around the laryngeal inlet by means of an inflatable cuff, without passing through the glottis. This makes it quicker to place, well tolerated at lighter anaesthetic planes and especially useful in difficult airways, as it does not require laryngoscopy.
The epiglottis elevating bars integrated into the distal opening prevent the epiglottis from obstructing the ventilation lumen, improving patency and seal security. The cuff, with a large volume and elongated shape, expands symmetrically to adapt to the paediatric anatomy; the inflation line incorporates a pilot balloon with a self-sealing valve and the ventilation tube is marked with the size, recommended patient weight and cuff inflation volume. It is a single-use, latex-free and DEHP-free product, sterilised with ethylene oxide (EO).
3. Technical Specifications
| Parameter | Specification |
| Product type | Disposable laryngeal mask (supraglottic airway device) |
| Size | 1.5 (Paediatric S) |
| Recommended patient | Children from 5 to 10 kg |
| Maximum cuff inflation volume | 7 ml of air |
| Inner diameter | 6.0 mm (±0.5 mm) |
| Outer diameter | 10.0 mm (±0.5 mm) |
| Radius of curvature | 120 mm (±30 mm) |
| Connector | Standard 15 mm (15M) |
| Material | PVC, polycarbonate (PC) and polyester; latex-free and DEHP-free |
| Design | Epiglottis elevating bars, high-volume cuff, anatomical and translucent body, colour-coded by size |
| Sterilisation | Ethylene oxide (EO), sterile, in accordance with ISO 11135 |
| Use | Single-use (disposable) |
| Presentation | Box of 10 sterile units |
| Reference Cahnos | LM0015101 |
| Manufacturer reference | 833315 (EAN box of 10 units: 8432229227364) |
Size dimension table for the family (according to the manufacturer’s official technical data sheet):
| Size | Recommended patient | Max. inflation vol. | Internal diameter (±0.5 mm) | External diameter (±0.5 mm) | Radius of curvature (±30 mm) | Manufacturer ref. |
| 1.0 | Neonate under 5 kg | 4 ml | 5.5 mm | 8.3 mm | 100 mm | 833310 |
| 1.5 | Children 5-10 kg | 7 ml | 6.0 mm | 10.0 mm | 120 mm | 833315 |
| 2.0 | Children 10-20 kg | 10 ml | 7.0 mm | 11.3 mm | 120 mm | 833320 |
| 2.5 | Children 20-30 kg | 14 ml | 9.0 mm | 13.3 mm | 140 mm | 833325 |
| 3.0 | Children/Adults 30-50 kg | 20 ml | 10.0 mm | 15.3 mm | 165 mm | 833330 |
| 4.0 | Adults 50-70 kg | 30 ml | 10.0 mm | 15.3 mm | 165 mm | 833340 |
| 5.0 | Adults 70-100 kg | 40 ml | 11.5 mm | 16.7 mm | 185 mm | 833350 |
4. Equivalences and Cross-References
Exact reference equivalence confirmed: size 1.5 of this line corresponds to manufacturer reference 833315 (same size, same recommended patient weight of 5-10 kg, same cuff volume of 7 ml, same internal diameter of 6.0 mm and same packaging of 10 units), according to the manufacturer’s official technical data sheet (Grupo Unidix) verified in the technical documentation of Cahnos. EAN of box of 10 units: 8432229227364.
Comparable at category level (not 1:1 interchangeable): reference 57581 (GIMA data sheet), same size No. 1.5, same weight range (5-10 kg) and same cuff volume (7 ml), but with a box packaging of 50 units and a tube length of 100 mm; it differs in presentation and tube dimensions, so it must be verified before substituting it in a tender.
Not verified: there is no equivalence table validated by the manufacturer against brands such as Ambu, Teleflex (LMA) or Intersurgical (i-gel); consult the technical department of Cahnos before using this reference as a direct substitute in specifications or tenders.
5. Selection tips and variants of interest
The main selection criterion should be the patient’s weight, not age: size 1.5 is indicated for children from 5 to 10 kg (designation Paediatric S). During use, it is recommended to inflate the cuff to a pressure of 18-20 cmH2O, check the integrity of the pilot balloon before insertion and confirm the seal by monitoring the expired volume.
At Cahnos you have the complete range of the same family in boxes of 10 units: sizes 1.0 neonatal (LM0010101), 2.0 (LM0020101), 2.5 (LM0025101), 3.0 (LM0030101), 4.0 (LM0040101) and 5.0 adult (LM0050101). If the procedure requires gastric access, the double-lumen version with drainage tube in size 1.5 (LM1115015, box of 5 units, with pre-inserted stylet) reduces the risk of aspiration. Always check the integrity of the sterile packaging and the expiry date before use; the device is single-use and must be disposed of after the procedure as medical waste according to current local regulations.





Reviews