Description
1. Safe ventilation in anaesthesia and neonatal care
The inflatable laryngeal mask size 1.0 neonatal is the reference supraglottic device for securing the airway quickly, atraumatically and reliably in newborns and neonates up to 5 kg. Its anatomical and translucent body, made of PVC, polycarbonate (PC) and polyester, allows permanent visual monitoring of the airway throughout the procedure, while the epiglottis elevating bars keep the lumen free and ensure an optimal seal around the laryngeal inlet.
It is the usual choice for paediatric anaesthesia services, neonatal intensive care units, delivery rooms, emergency departments and medical transport that need a disposable, sterile and ready-to-use device. It is presented in a box of 10 sterile units for continuous supply to the service, with an excellent balance between patient safety and cost per procedure.
2. What a laryngeal mask is and how it works
Unlike endotracheal intubation, the laryngeal mask is a supraglottic device that rests on the hypopharynx and seals around the laryngeal inlet via an inflatable cuff, without passing through the glottis. This allows faster placement, better tolerance at lighter anaesthetic planes and is particularly useful in difficult airways, as it does not require laryngoscopy.
The epiglottis elevating bars integrated into the distal opening prevent the epiglottis from obstructing the ventilation lumen, improving patency and seal security in a patient with such a small and delicate anatomy as the neonate. The cuff, high-volume and elongated, expands symmetrically to adapt to neonatal anatomy; the inflation line incorporates a pilot balloon with self-sealing valve and the ventilation tube is marked with the size, recommended patient weight and cuff inflation volume. It is a single-use, latex-free and DEHP-free product, sterilised with ethylene oxide (EO) and with a standard 15 mm (15M) connector.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Product type | Disposable laryngeal mask (supraglottic airway device) |
| Size | 1.0 (Neonatal) |
| Recommended patient | Neonates up to 5 kg |
| Maximum cuff inflation volume | 4 ml of air |
| Internal diameter (±0.5 mm) | 5.5 mm |
| External diameter (±0.5 mm) | 8.3 mm |
| Radius of curvature (±30 mm) | 100 mm |
| Connector | Standard 15 mm (15M) |
| Material | PVC, polycarbonate (PC) and polyester; latex-free and DEHP-free |
| Design | Epiglottis elevating bars, high-volume cuff, anatomical and translucent body, colour-coded by size |
| Inflation system | Pilot balloon with self-sealing valve |
| Recommended inflation pressure | 18-20 cmH2O |
| Sterilisation | Ethylene oxide (EO), sterile, in accordance with ISO 11135 |
| Use | Single-use (disposable), not reusable |
| Presentation | Box of 10 sterile units |
| Reference Cahnos | LM0010101 |
| Manufacturer reference | 833310 (Unidix Group) |
Dimension table by size for the family (according to the manufacturer’s official technical data sheet, standard series with epiglottis elevating bars):
| Size | Recommended patient | Max. inflation vol. | Internal diameter (±0.5 mm) | External diameter (±0.5 mm) | Radius of curvature (±30 mm) | Manufacturer ref. |
|---|---|---|---|---|---|---|
| 1.0 (this product) | Neonate < 5 kg | 4 ml | 5.5 mm | 8.3 mm | 100 mm | 833310 |
| 1.5 | Children 5-10 kg | 7 ml | 6.0 mm | 10.0 mm | 120 mm | 833315 |
| 2.0 | Children 10-20 kg | 10 ml | 7.0 mm | 11.3 mm | 120 mm | 833320 |
| 2.5 | Children 20-30 kg | 14 ml | 9.0 mm | 13.3 mm | 140 mm | 833325 |
| 3.0 | Children/Adults 30-50 kg | 20 ml | 10.0 mm | 15.3 mm | 165 mm | 833330 |
| 4.0 | Adults 50-70 kg | 30 ml | 10.0 mm | 15.3 mm | 165 mm | 833340 |
| 5.0 | Adults 70-100 kg | 40 ml | 11.5 mm | 16.7 mm | 185 mm | 833350 |
Technical note: the volume, diameter and radius of curvature values correspond to the manufacturer’s official technical documentation and are engraved on the tube of each unit, together with the size and recommended patient weight.
4. Equivalences and Cross-References
Confirmed reference equivalence: size 1.0 of this line corresponds to the manufacturer’s official reference 833310 (Unidix Group), according to the official technical data sheet verified in Cahnos’s technical documentation: same recommended patient (neonates up to 5 kg), same maximum cuff volume (4 ml), same diameters (internal 5.5 mm / external 8.3 mm) and same packaging pattern for the series. This correspondence is consistent with the rest of the family (833315 for size 1.5, 833320 for 2.0, 833325 for 2.5, 833330 for 3.0, 833340 for 4.0 and 833350 for 5.0).
Comparable at category level (not 1:1 interchangeable): reference 57580 (GIMA data sheet, size No. 1, patients up to 5 kg, 4 ml inflation, 15 mm connector) responds to the same general clinical indication for neonatal supraglottic airway, but is marketed in a box of 50 units and with a different tube length; it differs in presentation and tube dimensions, so it must be verified before substituting it in a tender.
Not verified: there is no equivalence table validated by the manufacturer against brands such as Ambu, Teleflex (LMA®) or Intersurgical (i-gel); consult Cahnos’s technical department before using this reference as a direct substitute in specifications or tenders.
5. Selection tips and variants of interest
The main selection criterion should be the patient’s weight, not age: size 1.0 is indicated for neonates up to 5 kg. During use, it is recommended to inflate the cuff to a pressure of 18-20 cmH2O, check the integrity of the pilot balloon and valve before insertion, and confirm the seal by monitoring the expired volume.
At Cahnos, the complete range of the same family is available in boxes of 10 units: sizes 1.5 (LM0015101), 2.0 (LM0020101), 2.5 (LM0025101), 3.0 (LM0030101), 4.0 (LM0040101) and 5.0 adult (LM0050101). If the neonatal procedure requires gastric access or there is a risk of regurgitation and aspiration, the double-lumen series with drainage tube (LM1115015 in size 1.5, with pre-inserted stylet) offers additional protection. Always check the integrity of the sterile packaging and the expiry date before use; the device is single-use and must be disposed of after the procedure as medical waste in accordance with current local regulations.