Description
1. Introduction
The preformed oral endotracheal tube with cuff and Murphy eye, ID 7.0 mm (reference ETPO70C02), is the airway solution that anaesthesia, emergency and intensive care services need to ensure ventilation of adult patients of medium or small build. Its preformed RAE-type geometry keeps the ventilation circuit away from the surgical field, which is especially valuable in oral, maxillofacial, ophthalmological and otorhinolaryngological surgery. It is supplied in a box of 10 sterile units to maintain a hospital stock that is always available.
2. What it is and how it works
In orotracheal intubation, the tube is inserted through the mouth, passes through the glottis and sits in the trachea to conduct gases from the ventilator or anaesthesia system to the lungs. The preformed curve follows the anatomy of the upper airway: it reduces the risk of kinking, minimises local trauma and facilitates fixation of the tube at chin level. The high-volume, low-pressure cuff (pneumatic seal) seals the trachea after inflation via the pilot balloon with self-sealing valve, allowing positive pressure ventilation without significant leaks. The Murphy eye, a distal lateral opening, guarantees an alternative gas passage if the tube tip becomes obstructed (by secretions or by the bevel resting against the tracheal wall). A continuous radiopaque line and depth graduation allow correct placement to be verified radiographically. It is a single-use device: it must be discarded after each patient to avoid any risk of cross-infection.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Material | Medical-grade polyvinyl chloride (PVC), transparent, latex-free |
| Configuration | Oral preformed (RAE/AGT type) with inflatable cuff and Murphy eye |
| Internal diameter (ID) | 7.0 mm |
| Cuff type | High volume / low pressure, with pilot balloon and self-sealing valve |
| Murphy eye | Yes, distal lateral eye |
| Connector | 15 mm standard, compatible with ISO circuits (ISO 5356-1) |
| Marking | Continuous radiopaque line and depth scale (cm) |
| Use | Single-use (sterile) |
| Presentation | Box of 10 units |
| Indication | Positive pressure ventilation and anaesthesia administration when face mask is not viable |
Other sizes in the same family (ETPO series, same design and presentation):
| Inner diameter | SKU (box of 10 units) |
|---|---|
| 3.5 mm | ETPO35C02 |
| 4.0 mm | ETPO40C02 |
| 4.5 mm | ETPO45C02 |
| 5.0 mm | ETPO50C02 |
| 5.5 mm | ETPO55C02 |
| 6.5 mm | ETPO65C02 |
| 7.0 mm | ETPO70C02 |
| 8.0 mm | ETPO80C02 |
| 8.5 mm | ETPO85C02 |
| 9.0 mm | ETPO90C02 |
Note: endotracheal tubes in this category are manufactured in accordance with the requirements of standard ISO 5361 (tracheal tubes and connectors) and the biocompatibility standards applicable to airway devices; verify the manufacturer’s declaration of conformity and CE marking on each supplied batch.
4. Equivalences and Cross-References
The SKU ETPO70C02 is Cahnos’s own code for this generic PVC reference; it must not be confused with the official part number of any specific manufacturer. To date, there is no manufacturer-validated equivalence table for this reference; we therefore classify the following options as comparable at category level (not interchangeable 1:1), pending technical verification against the manufacturer’s data sheet:
- Oral RAE/AGT type with cuff: preformed oral tubes with cuff and Murphy eye from lines such as the Rüsch AGT/RAE from Teleflex (indicative reference 111780070 for 7.0 mm, pending confirmation against primary source) or AirLife Preformed Cuffed Oral correspond to the same clinical indication and general geometry; they differ in material (PVC Ruschelit® in the case of Rüsch), length and cuff design, so they must be checked against the official technical data sheet before use in tenders.
- Generic imported preformed tubes: same category and materials (PVC with low-pressure cuff, Murphy eye, 15 mm connector and radiopaque line), with slight differences in geometry, length or cuff inflation pressure.
- Confirmed in Cahnos’s own catalogue: the diameters 6.5 mm (ETPO65C02), 8.0 mm (ETPO80C02), 8.5 mm (ETPO85C02), 9.0 mm (ETPO90C02) and the rest of the ETPO series share design and presentation; only the internal diameter changes.
There is no manufacturer-validated equivalence table for this reference; consult Cahnos’s technical department before using this reference as an equivalence in specifications or tenders.
5. Selection tips and variants of interest
- Selection by patient (usual practice in clinical use, not manufacturer specification): the 7.0 mm size is commonly used in adult patients of medium or small build and in adult women; as a general guide, tubes of ID 7.0–8.5 mm are usually used in women and 8.0–9.0 mm in men. The final choice must be based on the department’s protocol and anaesthetic assessment.
- Cuff pressure: inflate the cuff with a syringe through the valve and maintain a sealing pressure of 20–30 cmH2O (usual clinical practice range) to avoid tracheal damage; monitor with a manometer if the department has one.
- Preferred indications: oral and maxillofacial surgery, ENT, ophthalmology and head and neck procedures where it is important to keep the circuit out of the operative field, as well as mechanical ventilation in intensive care.
- Variants of interest in the catalogue: if you need another size, the ETPO series covers from 3.5 mm to 9.0 mm (ETPO35C02 to ETPO90C02). For rapid intubation in emergencies or prehospital care, consider the emergency kits with stylet and syringe (Endosid 845 series); for difficult airway or risk of kinking, the metal spiral-reinforced tubes from the same intubation section.