Description
1. Introduction
The preformed oral endotracheal tube with cuff and Murphy eye of 4.5 mm (reference ETPO45C02) is the airway solution that anaesthesia, paediatrics, emergency and intensive care departments need to ensure ventilation of paediatric patients or adults with a narrow airway. Its RAE/AGT preformed geometry keeps the ventilation circuit away from the surgical field, which is especially valuable in oral, maxillofacial, ENT and ophthalmological surgery. It is supplied in a box of 10 sterile units, designed to maintain an always-available hospital stock and ensure traceability of a single-use device.
2. What it is and how it works
In orotracheal intubation, the tube is inserted through the mouth, passes through the glottis and is placed in the trachea to conduct gases from the ventilator or anaesthesia system to the lungs. The preformed curve follows the anatomy of the upper airway: it reduces the risk of kinking, minimises local trauma and facilitates fixation of the tube at chin level. The high-volume, low-pressure cuff seals the trachea after inflation via the pilot balloon with self-sealing valve, allowing positive pressure ventilation without significant leaks. The Murphy eye, a distal lateral opening, ensures an alternative gas passage if the tube tip becomes obstructed, for example by secretions or by the bevel resting against the tracheal wall. The continuous radiopaque line and depth scale allow radiographic confirmation of correct placement. It is a single-use device: it must be discarded after each patient to avoid any risk of cross-infection.
3. Technical Specifications
| Parameter | Specification |
| Reference | ETPO45C02 |
| Type | Preformed oral endotracheal tube with cuff and Murphy eye |
| Configuration | Oral preformed (RAE/AGT type) with inflatable cuff and Murphy eye |
| Internal diameter (ID) | 4.5 mm |
| Material | Medical-grade polyvinyl chloride (PVC), transparent, latex-free |
| Cuff type | High volume / low pressure, with pilot balloon and self-sealing valve |
| Murphy eye | Yes, distal lateral safety opening |
| Connector | 15 mm standard, compatible with ISO circuits (ISO 5356-1) |
| Marking | Continuous radiopaque line and depth scale (cm) |
| Use | Single-use (sterile) |
| Clinical indication | Positive pressure ventilation and anaesthesia administration when a ventilation face mask is not viable |
| Presentation | Box of 10 units |
Note: this category of tracheal tubes is manufactured in accordance with the requirements of standard ISO 5361 (tracheal tubes and connectors) and the biocompatibility standards applicable to airway devices; verify the manufacturer’s declaration of conformity and CE marking on each supplied batch.
Other sizes in the same family (ETPO series, same design and presentation):
| Inner diameter | SKU (box of 10 units) |
| 3.5 mm | ETPO35C02 |
| 4.0 mm | ETPO40C02 |
| 4.5 mm | ETPO45C02 |
| 5.0 mm | ETPO50C02 |
| 5.5 mm | ETPO55C02 |
| 6.5 mm | ETPO65C02 |
| 7.0 mm | ETPO70C02 |
| 8.0 mm | ETPO80C02 |
| 8.5 mm | ETPO85C02 |
| 9.0 mm | ETPO90C02 |
4. Equivalences and Cross-References
SKU ETPO45C02 is Cahnos’s own code for this generic PVC reference; it must not be confused with the official part number of any specific manufacturer, so no exact reference equivalence can be declared and there is no manufacturer-validated equivalence table.
- Comparable at category level (not 1:1 interchangeable): the Rüsch® AGT Oral Preformed Cuffed ETT line from Teleflex, reference 111780045 (size 4.5, box of 10), has been verified in Teleflex’s official catalogue and catalogues of authorised distributors, matching the same indication and general geometry (oral preformed, with cuff and Murphy eye). It differs in material (thermo-sensitive PVC Ruschelit®), outer diameter (approx. 6.3 mm), length (approx. 145 mm) and cuff design, so it must be checked against the manufacturer’s technical data sheet before any use in tenders.
- Comparable by category, without 1:1 verification: generic imported oral preformed tubes (BOENMED, Greetmed, Pinmed, etc.), with PVC low-pressure cuff, Murphy eye, 15 mm connector and radiopaque line, but with slight differences in geometry, length or cuff inflation pressure.
- Confirmed in Cahnos’s own catalogue: the remaining diameters of the ETPO series (see table) share the same design and presentation; only the internal diameter changes.
There is no manufacturer-validated equivalence table for this reference; consult Cahnos’s technical department before using this reference as an equivalence in specifications or tenders.
5. Selection tips and variants of interest
- Selection by patient (usual use in clinical practice, not manufacturer specification): the 4.5 mm ID cuffed size is usually used in paediatric patients of approximately 6–8 years or in small airways; the final choice must always be based on the department’s protocol and anaesthetic assessment (age, weight and estimated airway size).
- Cuff pressure: inflate the cuff with a syringe through the valve and maintain a sealing pressure of 20–30 cmH2O (usual clinical practice range) to avoid tracheal damage; monitor with a manometer if the department has one.
- Preferred indications: oral and maxillofacial surgery, ENT, ophthalmology and head and neck procedures where it is important to keep the circuit out of the operative field, as well as mechanical ventilation in intensive care.
- Variants of interest in the catalogue: if you need other sizes, the ETPO series covers from 3.5 mm to 9.0 mm (ETPO35C02 to ETPO90C02). For nasal approach of the same diameter, the ETPN series is available (preformed nasal tracheal tube, e.g. ETPN45C02); for patients with a difficult airway or risk of kinking, consider the metal spiral-reinforced tubes from the same intubation section.





Reviews