Description
1. Introduction
The preformed nasal tracheal tube with cuff and Murphy eye of internal diameter 4.0 mm (ref. ETPN40C02) is the reference solution for airway management via the nasal route in anaesthesia operating theatres, intensive care units, emergency departments and resuscitation. Its preformed geometry reproduces the nasopharyngeal anatomical curvature, which facilitates insertion, reduces tissue trauma and keeps the surgical field clear in head and neck, oral, maxillofacial, ENT and ophthalmic procedures. Manufactured by DEAS S.r.l. (Deaflux Respiratory Products, Italy) and distributed in Spain by Cahnos, it is supplied in a box of 10 individually packaged, sterile units, ready for immediate use.
2. Technical explanation of the device and procedure
In nasotracheal intubation, the tube is inserted through a nostril, advances through the nasopharynx, passes through the glottis and sits in the trachea, where it conducts gases from the ventilator or anaesthesia circuit to the lungs. The preformed curvature with south orientation (curved towards the chin) follows the anatomy of the upper airway: it facilitates tube fixation, minimises the risk of kinking and prevents the circuit from interfering with the surgical field.
The low-pressure, high-volume cuff, inflated via the pilot balloon with self-sealing valve, seals the trachea and allows positive pressure ventilation without significant leaks, with lower pressure on the mucosa than high-pressure cuffs. The Murphy eye, a distal lateral hole smoothly formed and without sharp edges, guarantees an alternative gas passage if the tube tip becomes obstructed by secretions or by resting against the tracheal wall. The longitudinal radiopaque line, the double depth marker and the graduation every centimetre allow radiological and clinical verification of correct placement. The atraumatic, rounded tip minimises damage during intubation and extubation. It is a single-use device: it must be discarded after each patient to avoid the risk of cross-infection.
3. Technical Specifications
Technical data verified against the official technical data sheet of the ETPN range from manufacturer DEAS S.r.l. (Deaflux Respiratory Products, Italy), document ref. 04906-3e.doc Rev.3, and the ETPN range catalogue from Cahnos.
| Parameter | Specification |
|---|---|
| Reference | ETPN40C02 (own code of Cahnos) |
| Type | Preformed nasal tracheal tube with cuff and Murphy eye |
| Access route | Nasal (nasopharyngeal), preformed south-facing design |
| Internal diameter (ID) | 4.0 mm |
| External diameter (OD) | 5.3 mm |
| Total length | 210 mm |
| Inflated cuff diameter | 14 mm |
| Sleeve | Sealing, low pressure and high volume, with pilot balloon and self-sealing valve |
| Murphy eye | Yes, smoothly formed, distal lateral safety orifice |
| Material | Transparent, thermosensitive medical-grade polyvinyl chloride (PVC) |
| Connector | 15 mm universal connector conforming to ISO 5356-1 |
| Marking | Longitudinal radiopaque line, double depth marker and graduation every cm |
| Latex | Natural rubber latex-free |
| Phthalates | DEHP/DOP-free |
| Sterility | Sterile, individually packaged (ethylene oxide per EN 11135-1; residues per ISO 10993-7) |
| Use | Single-use, non-reusable |
| CE classification | Medical device class IIa (Directive 93/42/EEC, notified body TÜV SÜD) |
| Indication | Positive pressure ventilation and anaesthesia administration when a ventilation face mask is not viable |
| Presentation | Box of 10 individually packaged units |
ETPN range – preformed nasal tracheal tubes with cuff (same design, presentation and manufacturer, Cahnos references):
| Reference | ID (mm) | OD (mm) | Length (mm) | Cuff Ø (mm) |
|---|---|---|---|---|
| ETPN40C02 (this product) | 4.0 | 5.3 | 210 | 14 |
| ETPN45C02 | 4.5 | 6.0 | 225 | 14 |
| ETPN50C02 | 5.0 | 6.7 | 245 | 17 |
| ETPN55C02 | 5.5 | 7.3 | 275 | 17 |
| ETPN60C02 | 6.0 | 8.0 | 285 | 20 |
| ETPN65C02 | 6.5 | 8.7 | 295 | 20 |
| ETPN70C02 | 7.0 | 9.3 | 310 | 25 |
| ETPN75C02 | 7.5 | 10.0 | 315 | 25 |
| ETPN80C02 | 8.0 | 10.7 | 330 | 26 |
Note: the range of diameters, lengths and cuff diameters in the table comes from the manufacturer’s official technical data sheet (DEAS S.r.l. / Deaflux) and the ETPN range catalogue from Cahnos; do not extend these values to other sizes without technical confirmation. This category of tracheal tubes is manufactured in accordance with the requirements of standard ISO 5361 (tracheal tubes and connectors) and the biocompatibility standards applicable to airway devices; verify the manufacturer’s declaration of conformity and CE marking on each supplied batch.
4. Equivalences and Cross-References
The reference ETPN40C02 is an own code of the Cahnos range (manufacturer DEAS S.r.l. / Deaflux Respiratory Products, Italy) and must not be confused with the official part number of any third-party manufacturer. There is no exact reference equivalence table validated by the manufacturer for this code; consult the technical department of Cahnos before using this reference in specifications or tenders.
Functional category equivalence (not 1:1 interchangeable): preformed nasal tubes with cuff and Murphy eye from lines such as Rüsch® AGT/RAE nasal and Sheridan® (Teleflex), Shiley™ RAE nasal with TaperGuard™ cuff (Medtronic) or BOENMED nasal preformed tubes respond to the same general clinical indication and RAE nasal-type geometry. Any specific reference must be individually checked against the manufacturer’s official technical data sheet, as cuff technology, curve geometry, balloon material and connection system may differ; the ID 4.0 mm calibre of each brand must be verified before proposing stock substitution.
Comparables by category within the Cahnos catalogue (same ID 4.0 mm, other configurations):
| Reference | Description |
|---|---|
| ETAN40C02 | Oral endotracheal tube (Magill type) with cuff and Murphy eye |
| ETPO40C02 | Preformed oral endotracheal tube with cuff and Murphy eye |
The verification of this reference has been carried out against the manufacturer’s official technical data sheet (DEAS S.r.l. / Deaflux, ref. 04906-3e.doc Rev.3) and the Cahnos catalogue; there is no equivalence table validated by the manufacturer; consult the technical department before using in tenders.
5. Selection tips and variants of interest
Patient selection: the ID 4.0 mm calibre with cuff is usually used in paediatric patients or adults with narrow airways and nasal passages; the final choice must always be based on the department’s protocol and anaesthetic assessment (age, weight and estimated airway and nasal passage calibre).
Cuff pressure: inflate through the valve with a syringe and maintain a sealing pressure of 20–30 cmH₂O (usual clinical practice range, not a manufacturer specification) to minimise tracheal mucosal damage; monitor with a manometer if available in the department.
Position verification: check by bilateral auscultation and, if possible, radiography using the radiopaque line and the tube’s double depth marker.
Preferred indications: oral, maxillofacial, ENT and ophthalmic surgery, where it is important to keep the circuit out of the operative field, as well as mechanical ventilation in intensive care.
Catalogue variants of interest: if you need another size, the ETPN series covers from 4.0 mm to 8.0 mm (ETPN40C02 to ETPN80C02). For oral intubation, consider the ETAN (Magill type) and ETPO (oral preformed) ranges. Contact the Cahnos technical department for samples, volume conditions and hospital supply lead times.