Description
1. Introduction
The reinforced oral endotracheal tube with cuff and Murphy eye of 5.5 mm (Ref. ETA055C02A2) is the airway solution that anaesthesia, emergency and intensive care services need when face-mask ventilation is not viable. Manufactured in medical-grade polyvinyl chloride (PVC), transparent and latex-free, it combines the strength of the helical anti-kink reinforcement with the airtight sealing of the cuff and the safety of the Murphy eye, providing a patent and stable airway during critical procedures. It is supplied in a box of 10 individually packaged units, designed for the intubation trolley and high-turnover hospital consumables.
2. Technical explanation of the equipment and procedure
In orotracheal intubation, the tube is inserted orally under laryngoscopic vision, passes through the glottis and sits in the trachea below the vocal cords, conducting gases from the ventilator, the anaesthesia circuit or the resuscitation bag to the lungs. The helical reinforcement integrated into the PVC wall prevents tube collapse under external compression, biting or kinking, making it especially suitable for head and neck surgery, ENT, maxillofacial, thyroid, prone positioning and complex surgical positions.
The sealing cuff, inflated via the pilot balloon with self-sealing valve, seals the tracheal lumen, prevents gas leaks, reduces the risk of aspiration of secretions and stabilises the device during positive-pressure ventilation. The Murphy eye, a distal lateral opening, acts as a safety route: if the tip of the tube becomes obstructed by secretions or rests against the tracheal wall, respiratory flow continues through that opening and occlusion is never complete. The universal 15 mm connector, radiopaque line and depth graduation allow connection to the circuit and radiographic verification of position. It is a single-use device: after each patient it must be discarded to avoid any risk of cross-infection.
3. Technical Specifications
The specifications set out below correspond to the technical data published for this reference in the Cahnos catalogue:
| Parameter | Specification |
|---|---|
| Reference (SKU) | ETA055C02A2 (Cahnos distributor’s own code; ETA series reinforced with cuff, without guide) |
| Type | Oral endotracheal tube (orotracheal), Magill-type curvature, reinforced |
| Access route | Oral (orotracheal) |
| Internal diameter (ID) | 5.5 mm |
| Reinforcement | Anti-kink helical spiral integrated into the tube wall |
| Material | Medical-grade polyvinyl chloride (PVC), transparent, latex-free |
| Sleeve | Sealing (cuff inflation), with pilot balloon and self-sealing valve |
| Murphy eye | Present (distal lateral safety opening) |
| Tip | Atraumatic, rounded and gently bevelled |
| Connector | Universal 15 mm, compatible with ISO circuits (ISO 5356-1) |
| Marking | Radiopaque contrast line and depth graduation |
| Sterilisation | Ethylene oxide (EO), sterile, individually packaged |
| Use | Single-use, non-reusable |
| Indication | Positive pressure ventilation and anaesthesia administration when face mask is not viable |
| Indicative population | Older child, adolescent or adult with a reduced-calibre airway |
| Presentation | Box of 10 units |
| Regulatory standard | ISO 5361 (tracheal tubes); connector ISO 5356-1; verify CE/MDR marking of each batch |
Range of sizes documented in Cahnos (ETA…C02A2 series, reinforced with cuff and Murphy eye, without guide):
| Reference | Internal diameter (ID) | Indicative patient |
|---|---|---|
| ETA035C02A2 | 3.5 mm | Newborn |
| ETA045C02A2 | 4.5 mm | Child |
| ETA055C02A2 (this product) | 5.5 mm | Older child / adolescent / adult with reduced airway |
| ETA065C02A2 | 6.5 mm | Small adult |
| ETA075C02A2 | 7.5 mm | Adult |
| ETA080C02A2 | 8.0 mm | Adult |
| ETA085C02A2 | 8.5 mm | Adult / large adult |
| ETA090C02A2 | 9.0 mm | Large adult |
| ETA095C02A2 | 9.5 mm | Large adult |
| ETA010C02A2 | 10.0 mm | Large adult |
Note: this category of tracheal tubes is manufactured in accordance with the requirements of standard ISO 5361 (respiratory and anaesthesia equipment: tracheal tubes and connectors) and the biocompatibility standards applicable to airway devices. Verify the declaration of conformity, CE/MDR marking and the manufacturer’s classification on each supplied batch before use in procedures.
4. Equivalences and Cross-References
Reference ETA055C02A2 is Cahnos’s own catalogue code (ETA series: reinforced oral tracheal tubes with cuff, without guide) and must not be confused with the official reference number of any specific manufacturer. To date, there is no manufacturer-validated equivalence table for this reference, so the following options are classified as comparable at category level (not interchangeable 1:1), pending technical verification against each manufacturer’s official datasheet:
| Manufacturer / Line | Reference | Description | Key difference |
|---|---|---|---|
| Endosid® (Asid Bonz) | 849055 | Spiral reinforced endotracheal tube with Murphy eye and cuff, ID 5.5 mm | Comparable by category; verify inflation system, outer diameter and guide |
| P. J. Dahlhausen | 44.000.20.055 | Reinforced endotracheal tube with cuff (Woodbridge type), ID 5.5 mm, PVC, sterile | Comparable by category; verify presence of Murphy eye and connector |
| P. J. Dahlhausen | 44.000.10.055 | Magill-type endotracheal tube with cuff and Murphy eye, ID 5.5 mm | Comparable by category; differs in reinforcement (without helical armouring) |
| Teleflex (Rüsch® / Sheridan®), Medtronic (Shiley™), ICU Medical (Portex®), Intersurgical | — | Reinforced cuffed endotracheal tube lines on the market | Comparable by category; not interchangeable 1:1 without verification of datasheet and IFU |
Comparables within the Cahnos catalogue (same ID 5.5 mm size, not interchangeable 1:1):
| Reference | Description | Key difference |
|---|---|---|
| ETA055002A2 | Reinforced oral tube without cuff, with Murphy eye, ID 5.5 mm | Without cuff |
| ETAM55C02A1 | Reinforced oral tube with cuff, insertion guide and Murphy eye, ID 5.5 mm | Adds included insertion stylet |
| ETAN55C02 | Oral tube with cuff and Murphy eye, ID 5.5 mm | Without helical reinforcement |
| ETAN55002 | Oral tube without cuff and without reinforcement, with Murphy eye, ID 5.5 mm | Without cuff or reinforcement |
Fields to be checked before comparing: internal diameter (5.5 mm), material (medical-grade PVC), cuff type (high volume / low pressure), presence of Murphy eye, 15 mm connector and helical reinforcement. There is no manufacturer-validated equivalence table for this reference; consult the Cahnos technical department before using this reference as an equivalence in tenders or bids or as a direct replacement for stock from other brands.
5. Selection tips and variants of interest
Size selection: the 5.5 mm ID is intended for older paediatric patients, adolescents or adults with a reduced-calibre airway. As a general clinical practice guideline (not a manufacturer specification), the largest size compatible with the patient’s anatomy and the centre’s protocol is always selected; in this paediatric-adolescent range, the internal diameter directly affects flow and resistance to gas passage.
Cuff pressure: check the integrity of the cuff, pilot balloon and lumen patency before intubation; inflate with a syringe and maintain a sealing pressure of 20–30 cmH₂O (usual range in clinical practice) to minimise the risk of pressure injury to the tracheal mucosa, monitoring with a manometer if one is available.
Pre-intubation verification: check the pre-inflation of the cuff, patency of the Murphy eye, correct attachment of the 15 mm connector and integrity of the helical reinforcement along the tube body.
Variants of interest: if insertion assistance is required, consider Ref. ETAM55C02A1 (same size, with guide); if the procedure is short and the balloon is to be avoided, Ref. ETA055002A2 (reinforced without cuff). Complete the intubation cart stock with the adjacent sizes in the range (ETA045C02A2 and ETA065C02A2) and, in case of risk of ventilator-associated pneumonia (VAP) in prolonged intubations, consult the lines with subglottic suction lumen. Single-use product: do not re-sterilise or reuse. Contact the technical-commercial department of Cahnos for samples, volume conditions and approval of references for your tender.



Reviews