Description
1. Introduction
The oral endotracheal tube with cuff and Murphy eye of 6.5 mm (Ref. ETAN65C02) is the airway solution that anaesthesia, emergency and intensive care departments need to ensure ventilation of adolescent and adult patients of small build or with a reduced-calibre airway. Made of medical-grade polyvinyl chloride (PVC), sterile and single-use, this tube guarantees an effective seal and a patent, safe airway during mechanical ventilation, surgical procedures and resuscitation. It is supplied in a box of 10 units, designed to keep hospital stock always available.
2. What it is and how it works
In orotracheal intubation, the tube is inserted through the mouth, passes through the glottis and sits in the trachea to conduct gases from the ventilator, the anaesthesia circuit or the resuscitation bag to the lungs. The inflatable cuff (pneumotamponade) seals the trachea after being inflated via the valve, allowing positive-pressure ventilation without significant leaks and reducing the risk of aspiration. The Murphy eye, a distal lateral opening, acts as an alternative ventilation route: if the main lumen were obstructed by secretions or by the bevel resting against the tracheal wall, airflow continues through this opening. The radiopaque line and depth graduation allow radiographic verification of correct tube placement. It is a single-use device: it must be discarded after each patient to avoid any risk of cross-infection.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference | ETAN65C02 |
| Access route | Oral (orotracheal) |
| Internal diameter (ID) | 6.5 mm |
| External diameter (OD) | 8.7 mm |
| Length | 290 mm |
| Cuff diameter | 20 mm |
| Type | With inflatable cuff and Murphy eye |
| Material | Medical-grade polyvinyl chloride (PVC), transparent, latex-free |
| Connector | Universal 15 mm (ISO 5356-1) |
| Marking | Radiopaque contrast line and depth graduation |
| Sterilisation | Ethylene oxide (EO), sterile |
| Use | Single-use, disposable |
| Presentation | Box of 10 units |
| Indication | Positive pressure ventilation and anaesthesia administration when face mask is not viable |
The ETAN series (the family to which this product belongs) covers the following sizes in the same configuration and presentation, in boxes of 10 units:
| Inner diameter | Reference (box of 10 units) |
|---|---|
| 2.5 mm | ETAN25C02 |
| 3.0 mm | ETAN30C02 |
| 3.5 mm | ETAN35C02 |
| 4.0 mm | ETAN40C02 |
| 4.5 mm | ETAN45C02 |
| 5.0 mm | ETAN50C02 |
| 5.5 mm | ETAN55C02 |
| 6.0 mm | ETAN60C02 |
| 6.5 mm | ETAN65C02 (this product) |
| 7.0 mm | ETAN70C02 |
| 7.5 mm | ETAN75C02 |
| 8.0 mm | ETAN80C02 |
| 8.5 mm | ETAN85C02 |
| 9.0 mm | ETAN90C02 |
| 9.5 mm | ETAN95C02 |
| 10.0 mm | ETAN10C02 |
Note: endotracheal tubes in this category are manufactured in accordance with the requirements of standard ISO 5361 (tracheal tubes and connectors) and the biocompatibility standards applicable to airway devices; verify the manufacturer’s declaration of conformity and CE/MDR marking on each supplied batch.
4. Equivalences and Cross-References
The SKU ETAN65C02 is Cahnos’s own code (DEAFLUX family) for this generic PVC tube; it should not be confused with the official part number of any specific manufacturer. To date, there is no manufacturer-validated equivalence table for this reference; we therefore classify the following options as comparable at category level (not 1:1 interchangeable), pending technical verification against each manufacturer’s official datasheet:
- Comparable by category (same 6.5 mm calibre, with cuff and Murphy eye): ENDOSID tube with low-pressure cuff and Murphy eye Ref. 841065 (Cahnos’s own catalogue); Magill-type tube with Murphy eye and cuff from P. J. Dahlhausen Ref. 44.000.10.065; PRO-BREATHE tube from ISSO Ref. PBET41265. They may differ in PVC compound, length, cuff inflation pressure and bevel design.
- Confirmed in Cahnos’s own catalogue: the remaining diameters of the ETAN series (see previous table) share the same design and presentation; only the calibre varies. In the same 6.5 mm calibre, the catalogue also offers the ENDOSID Magill tube without cuff (Ref. 840065), the version without cuff with Murphy eye (Ref. 843065), the emergency intubation kit with stylet and syringe (Ref. 845065), the preformed RAE-type with cuff (Ref. ETPO65C02) and the spiral-reinforced (Ref. ETA065C02A2).
There is no manufacturer-validated equivalence table for this reference; consult Cahnos’s technical department before using this reference as an equivalence in tenders or bids.
5. Selection tips and variants of interest
Selection by patient (usual practice in clinical use, not manufacturer specification): the 6.5 mm calibre is used as a transition size in adolescents and adults of small build or with a narrow airway; as a general guide, internal diameters of 7.0–8.5 mm are used in women and 8.0–9.0 mm in men. The final choice must be based on the department’s protocol and the patient’s anaesthetic assessment.
Cuff pressure: check the integrity of the cuff and valve before intubation, inflate with a syringe and maintain a sealing pressure of 20–30 cmH2O (usual clinical practice range) to avoid pressure injuries on the tracheal mucosa, monitoring with a manometer if the department has one.
Variants of interest in the catalogue: if you need another calibre, the ETAN series covers from 2.5 mm to 10.0 mm (ETAN25C02 to ETAN10C02). For long-duration procedures or airtight sealing, consider the ENDOSID with low-pressure cuff (Ref. 841065); for oral, maxillofacial or head and neck surgery, the preformed ETPO65C02; in case of kinking risk, the reinforced ETA065C02A2; and for prolonged ventilation with subglottic drainage, the ETSL series (Ref. ETSL65C02). Consult our technical-commercial team at Cahnos for reference approval and configuration of the most suitable size stock for your department.