Description
1. Commercial Introduction
The Monocyte-N (Ref. 770300/TL) is a single-lumen oocyte aspiration needle in gauge 16G × 330 mm, designed for oocyte retrieval in in vitro fertilisation cycles with maximum safety and precision. Its -N version incorporates a suction extension of 80 cm with luer lock connector, allowing a clean and stable connection to the operating theatre suction pump. It is supplied in a box of 10 units with individual packaging, being the professional supply solution for fertility clinics, assisted reproduction units and gynaecology practices that require a sterile, single-use device with CE marking.
2. Technical explanation of the equipment and procedure
The Monocyte-N needle is a single-lumen needle intended for ultrasound-guided transvaginal follicular aspiration (OPU technique / oocyte puncture). Its design includes an echogenic tip with atraumatic sharpening that facilitates smooth tissue penetration and good visualisation on the ultrasound scanner monitor, minimising follicular and tissue trauma during oocyte collection.
- Echogenic tip with atraumatic sharpening: controlled penetration and excellent ultrasound visibility.
- Blue-tinted tube: protects the oocyte from light and UV rays during suction.
- Medical thermoplastic rubber cap: connects securely to the collection tube.
- Two lateral wings aligned with the bevel: help the operator to orientate during the penetration phase.
- 80 cm extension with luer lock connector: connection to the suction pump without the need for intermediate couplings.
- Universal adapter included: compatible with a wide variety of connections.
In clinical practice, after preparing the vaginal area and the sterile field, the needle is attached to the ultrasound probe guide and introduced following the trajectory on the monitor to the follicle, proceeding to gently aspirate the cumulus-oocyte complex. It is a single-use product, sterilised by ethylene oxide (EtO) with CE marking 0051.
3. Technical Specifications
| Parameter | Specification |
| Reference (SKU) | 770300/TL |
| Family / series | Monocyte-N (version with pump extension) |
| Product type | Single-lumen oocyte aspiration needle |
| Configuration | Single lumen, single channel (monolumen) |
| Gauge | 16G (standard outer diameter ≈ 1.65 mm; blue luer) |
| Length | 330 mm |
| Tip | Echogenic with atraumatic sharpening |
| Lumen | Single lumen (single suction channel) |
| Suction extension | 80 cm with Luer lock connector |
| Connection | Luer lock + universal adapter included |
| Collection cap | Medical thermoplastic rubber for collection tube |
| Oocyte protection | Blue-tinted tube against UV rays |
| Orientation | Two lateral wings aligned with the tip bevel |
| Sterilisation | Ethylene oxide (EtO) — CE 0051 marking |
| Use | Single-use (non-reusable) |
| Presentation | Box of 10 units, individually packaged |
| Indication | Oocyte aspiration in assisted reproduction (IVF) |
4. Equivalences and Cross-References
Below are listed references from other manufacturers cross-checked against official sources (manufacturer catalogue and website, technical data sheets and GUDID regulatory databases) for the same indication: ultrasound-guided oocyte aspiration with a single-lumen needle in gauge 16G and length close to 330 mm. None of them is interchangeable 1:1 without technical validation of each centre’s protocol.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Cook Medical | Ovum Aspiration Needle Single Lumen (OSN), bevel B, regular wall | K-OSN-1633-R-B-90 (G55480) | Comparable by category (not interchangeable 1:1) | 16G × 33 cm (330 mm) with a 90 cm suction line; EchoTip® echogenic tip and handle with bevel orientation notch. Verified in the official Cook Medical catalogue and GUDID. |
| CooperSurgical | Wallace® Single Lumen Oocyte Needle (needle only) / Set with tube | ON1633 / ONS1633 | Comparable by category (not interchangeable 1:1) | 16G × 33 cm needle, non-coring three-facet tip with echogenic marking to the tip, box of 10 units (GUDID 00888937021134); version with tube ONS1633 (950 mm). Blue tube sleeve (16G identification). |
| Rocket Medical | Embryon® SX™ Single Lumen Aspiration Needle | R57601-SX-90 | Comparable by category (not interchangeable 1:1) | 16G × 33 cm (330 mm) with a 90 cm suction tube; high-definition ultrasound spiral bevel and handle with orientation marker (official Rocket datasheet: OD 1.60 mm / ID 1.20 mm). |
| Kitazato | Single Lumen OPU Needle with Luer lock connector | 316001 | Comparable by category (not interchangeable 1:1) | 16G × 32.5 cm (325 mm), single lumen with luer lock connection for manual suction, box of 10 units; Thin Wall triple-cut design with echogenic marking (official Kitazato website). Length 5 mm shorter. |
Other reference brands in the single-lumen oocyte needle category (e.g. Vitrolife, with its Conventional single lumen range in gauges 16.5–17G and lengths of 300–350 mm, or Gynétics FAS Premium, usually in 17G) have different gauges, lengths or connector designs and must be compared field by field (gauge, length, tip type, connection, sterility) before any substitution.
There is no equivalence table validated by the manufacturer for the 1:1 cross-reference with other brands; consult the technical department before use in tendering.
5. Selection tips and variants of interest
- Gauge selection: the 16G model (Ref. 770300/TL) offers a larger internal diameter and higher suction flow; if your protocol requires a finer needle with less trauma, consider the Monocyte-N 17G version (Ref. 770307/TL).
- With tube (-N version) vs. without tube: the Monocyte-N series includes an 80 cm extension with luer lock for integration into pump suction systems; the Monocyte version without extension (Refs. 770300 and 770307) connects directly via luer lock to a syringe or equipment line.
- Single lumen vs. follicular flushing: if in addition to aspirating you need to irrigate the follicle, consider the Bimonocyte-N family (Refs. 770302/TL and 770304/TL) or the Doublecyte double-lumen (Refs. 770301/TL, 770308/TL and versions without tube).
- Economical alternative without tube: the Monolume range (Refs. 770312 in 16G and 770311 in 17G) offers a single lumen with universal fitting and female luer lock port for direct connection to a syringe, as well as allowing aspiration of ovarian cysts.
- Good practices: single-use product, sterile by EtO; use it immediately after opening the packaging, verify the integrity of the packaging and dispose of it in accordance with current legislation. For high volumes or framework agreements, consult our technical-commercial team.





Reviews
There are no reviews yet.