Description
1. Introduction
The MONOCYTE-N (Ref. 770307/TL) is a single-lumen, single-channel oocyte aspiration needle designed for atraumatic oocyte retrieval in in vitro fertilisation (IVF) cycles. Its version with pump connection tubing makes it the ideal solution for assisted reproduction units that integrate vacuum aspiration systems, ensuring stable flow, clear ultrasound visualisation and safe follicular retrieval. Supplied in a box of 10 units, it is a sterile, single-use product with CE marking, intended for professional procurement by fertility clinics.
2. Technical explanation of the equipment and procedure
The MONOCYTE-N needle is a single-channel device for ultrasound-guided oocyte aspiration. It incorporates an echogenic tip with atraumatic bevel that facilitates tissue penetration with excellent visibility on the ultrasound scanner monitor, and a bluish tube that protects the oocyte from UV rays during aspiration. The end of the circuit terminates in a medical thermoplastic rubber stopper that connects to the collection tube, sealing the follicular fluid inside.
The handle incorporates two lateral wings aligned with the tip bevel, which guide the operator during the penetration phase, and the gauge is identified by the handle colour code: white for 17G and blue for 16G. Connection to the suction pump is made via a 80–100 cm tube with female luer lock connector, completed with an additional universal adapter allowing different types of connection. In clinical practice, after preparing the vaginal area and the sterile field, the needle is attached to the ultrasound probe and introduced following the trajectory on the monitor to the follicle, proceeding with gentle aspiration. The product includes MEA embryotoxicity test and LAL bacterial endotoxin testing.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference (SKU) | 770307/TL |
| Brand / family | MONOCYTE-N (version with pump connection tube) |
| Product type | Single-lumen oocyte aspiration needle, single channel |
| Gauge | 17G (external diameter ≈ 1.5 mm; white handle) |
| Length | 330 mm |
| Configuration | Single channel, single lumen (monolumen) |
| Tip | Echogenic with atraumatic sharpening |
| Light protection | Bluish tube that protects the oocyte from UV rays |
| Pump connection | 80–100 cm tube with female luer lock connector + additional universal adapter |
| Collection cap | Medical thermoplastic rubber for connection to collection tube |
| Orientation | Two lateral wings aligned with the tip sharpening |
| Gauge identification | Handle colour code: white (17G) / blue (16G) |
| Sterilisation | Radiation (beta rays) according to the manufacturer’s technical data sheet; CE marking 0051 |
| Use | Single-use, individually packaged |
| Quality controls | MEA embryotoxicity test and LAL endotoxin control |
| Presentation | Box of 10 units |
| Application | Oocyte aspiration in IVF procedures; atraumatic collection of oocytes |
4. Equivalences and Cross-References
The following are references from other manufacturers cross-checked against official sources (manufacturer catalogue and technical data sheets, GUDID/FDA regulatory databases) for the same indication: ultrasound-guided oocyte aspiration with a single-lumen needle in 17G gauge and length close to 330 mm. None of them is interchangeable 1:1 without technical validation of each centre’s protocol.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Cook Medical | Ovum Aspiration Needle Single Lumen (OPS), bevel A, thin wall | K-OPS-1733-T-A-60 (G55491) | Comparable by category (not interchangeable 1:1) | 17G × 33 cm (330 mm) with 60 cm suction line; EchoTip® echogenic tip and handle with bevel orientation notch. |
| CooperSurgical | Wallace® Single Lumen Oocyte Recovery | ON1733 | Comparable by category (not interchangeable 1:1) | 17G × 33 cm (330 mm) needle, non-coring three-facet tip with echogenic marking, presentation in box of 10 units (GUDID 00888937021141); version with tube: ONS1733. |
| Rocket Medical | Embryon® SX™ Single Lumen Aspiration Needle (Oxford) | R57602-00-SX | Comparable by category (not interchangeable 1:1) | 17G × 33 cm (330 mm) with 70 cm tube; OD 1.40 mm / ID 1.00 mm; 40 mm high-definition echogenic spiral and handle with orientation marker. |
| Gynemed | Oocyte Aspiration Needle, single lumen without flushing | GY-711724LF | Comparable by category (not interchangeable 1:1) | 17G × 32 cm (320 mm), single lumen without flushing, sterile by gamma irradiation, CE class IIa. Length 10 mm shorter than the MONOCYTE-N. |
| Vitrolife | Conventional Single Lumen (oocyte retrieval) | 17189 | Comparable by category (not interchangeable 1:1) | 17G (OD 1.5 / ID 1.2 mm) × 350 mm with 900 mm aspiration tube; presentation of 10 units. Length 20 mm longer. |
Other reference brands in the single-lumen oocyte needle category (e.g. Gynetics, with its Premium FAS #4551 series in 17G NS-CS1 or NS-DS1 versions, or Kitazato) have different lengths, wall thicknesses, connector designs and tube types and must be compared field by field (gauge, length, internal diameter, tip type, connection, sterility) before any substitution. There is no equivalence table validated by the manufacturer for 1:1 cross-referencing with other brands; consult the technical department before use in tenders.
5. Selection tips and variants of interest
Gauge selection: the 17G model offers an excellent balance between rigidity and flow with minimal follicular trauma. If your protocol requires a larger internal diameter and higher suction flow rate, consider the MONOCYTE-N 16G × 330 mm version (Ref. 770300/TL).
Connection versions: the MONOCYTE family is available without tubing (Refs. 770307 in 17G and 770300 in 16G, with direct luer lock connection) and with 80–100 cm pump tubing (/TL series: 770307/TL and 770300/TL), as in this reference.
Single lumen vs. double lumen: if in addition to aspirating you need to irrigate or flush the follicle, consider the dual-channel BIMONOCYTE/BIMONOCYTE-N family (Refs. 770302/770304 and /TL versions in 17G/16G) or the double-lumen DOUBLECYTE (Refs. 770301/770308 and /TL versions).
Good practice: single-use product; use it immediately after opening the individual package, check the integrity of the packaging and dispose of it in accordance with current legislation on healthcare waste. For high volumes or framework agreements, contact our technical-commercial team.