Description
1. Commercial Introduction
The BIMONOCYTE needle (Ref. 770304) is a precision instrument for dual-channel oocyte aspiration and follicular flushing in 16G × 330 mm, designed for oocyte retrieval in in vitro fertilisation cycles with maximum safety, precision and respect for the gamete. Its configuration allows aspiration of the cumulus-oocyte complex and flushing of the follicle with culture medium without changing instruments, reducing handling, procedure times and stress on the oocyte.
It is supplied in a box of 10 units with individual packaging, making it the professional supply solution for fertility clinics, assisted reproduction units and gynaecology practices requiring a sterile, single-use device with CE 0051 marking. This no-extension version connects directly by Luer lock to a syringe or suction equipment line.
2. Technical explanation of the equipment and procedure
The BIMONOCYTE needle is a dual-channel device with a single-lumen needle: the first channel is used for oocyte aspiration through the needle lumen, and the second for follicular flushing, which runs through an external tube in transparent material connected by a Y-connector that also acts as a handle or grip. It consists of:
- Echogenic tip with atraumatic bevel: smooth penetration and excellent visibility on the ultrasound scanner monitor, minimising follicular and tissue trauma.
- Blue-tinted tube: protects the oocyte from light and UV rays during aspiration.
- Medical thermoplastic rubber cap: connects securely to the collection tube, sealing the follicular fluid inside.
- Y-connector (handle): connects the needle to the aspiration and flushing tubes, equipped with 2 universal accessories.
- Transparent follicular flushing line with stopcock: prevents reflux of the flushing fluid.
- Tube clamp: regulates follicular aspiration during the procedure.
- Direct Luer lock connection (no-extension version) completed with an additional universal adapter for different connection types.
In clinical practice, after preparing the vaginal area and the sterile field, the needle is attached to the ultrasound probe guide and introduced following the trajectory on the monitor to the follicle, proceeding with gentle aspiration of the cumulus-oocyte complex and follicular flushing. It is a single-use product, sterilised by ethylene oxide (EtO) with CE 0051 marking.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference (SKU) | 770304 |
| Brand / family | BIMONOCYTE (no-extension version, direct Luer lock connection) |
| Product type | Dual-channel oocyte aspiration and follicular flushing needle (single-lumen needle) |
| Gauge | 16G (azzurro / blue Luer lock connector) |
| Length | 330 mm |
| Tip | Echogenic with atraumatic sharpening |
| Aspiration channel | Needle lumen + direct Luer lock connection (no extension) |
| Follicular flushing channel | Transparent external tube with anti-reflux stopcock and Y-connector with 2 universal accessories |
| Collection cap | Medical thermoplastic rubber for collection tube |
| Oocyte protection | Blue-tinted tube against UV rays |
| Orientation | Two lateral wings aligned with the tip bevel |
| Gauge identification | Connector colour code: azzurro (16G) / transparent (17G) |
| Sterilisation | Ethylene oxide (EtO) — CE 0051 marking |
| Use | Single-use (non-reusable) |
| Presentation | Box of 10 units, individually packaged |
| Indication | Oocyte aspiration and follicular flushing in assisted reproduction (IVF / OPU) |
BIMONOCYTE / BIMONOCYTE-N family variants:
| Reference (without tube) | Reference (with 80 cm extension) | Gauge | Luer lock connector colour | Presentation |
|---|---|---|---|---|
| 770304 (this product) | 770304/TL | 16G × 330 mm | Azzurro (blue) | Box of 10 units |
| 770302 | 770302/TL | 17G × 330 mm | Transparent | Box of 10 units |
4. Equivalences and Cross-References
The following are references from other manufacturers cross-checked against official sources (manufacturer catalogue and technical data sheets and the GUDID/FDA regulatory database) for the same indication: ultrasound-guided transvaginal oocyte aspiration with follicular flushing (dual-channel / double-lumen) in 16G calibre and length close to 330 mm. None of them is interchangeable 1:1 without technical validation of each centre’s protocol.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Cook Medical | EchoTip® Double-Lumen Ovum Aspiration Needle (K-OPSD series) | K-OPSD-1633-A-S-US (G55789) | Comparable by category (not interchangeable 1:1) | 16G × 33 cm (330 mm) with true double lumen in the needle and 75 cm aspiration line; EchoTip® echogenic tip and handle with bevel orientation marker (official Cook catalogue / GUDID). |
| CooperSurgical | Wallace® Dual Lumen Oocyte Recovery Set | DNS1633-750 | Comparable by category (not interchangeable 1:1) | 16G × 33 cm (330 mm) with 750 mm aspiration tube and integrated flushing line to the tip; non-coring three-bevel tip with echogenic marking; box of 10 units (verified in GUDID 00888937020908). |
| Rocket Medical | Embryon® Double Lumen Oocyte Aspiration Needle | R57610 | Comparable by category (not interchangeable 1:1) | 16G × 32 cm (320 mm) with 70 cm tube, Oxford-style tube set and three-way stopcock as standard; engraved echogenic band; 10 units/box (official Rocket catalogue). Length 10 mm shorter. |
| Kitazato | Double Lumen OPU Needle (with aspiration line) | 376350 | Comparable by category (not interchangeable 1:1) | 16G × 35 cm (350 mm), 80 cm aspiration line and 100 cm flushing line; thin-wall technology, triple cut and tip with echogenic marking; tested with MEA assay (official Kitazato catalogue). Length 20 mm longer. |
Other reference brands in the category of oocyte needles with follicular flushing (for example, Vitrolife, the FAS line from Gynetics or Rocket Medical in SX™ double-lumen versions, R57611) have different lengths, gauges or connection designs and must be compared field by field (gauge, length, tip type, connection, sterility) before any substitution. There is no equivalence table validated by the manufacturer for a 1:1 cross-reference with other brands; consult the technical department before use in tenders.
5. Selection tips and variants of interest
- Gauge selection: the 16G model (Ref. 770304) offers a larger internal diameter and therefore greater suction flow, recommended when a high flow rate is required or follicles with dense fluid; if your protocol requires less follicular trauma, consider the BIMONOCYTE 17G version (Ref. 770302), with transparent connector.
- Without tubing vs. with tubing (-N version): this reference (770304) connects directly via Luer lock to a syringe or equipment line; if it needs to be integrated into pump suction systems with an 80 cm extension and Luer lock connector, consider the BIMONOCYTE-N version (Ref. 770304/TL in 16G and 770302/TL in 17G).
- Dual-path vs. true double lumen in the needle: the BIMONOCYTE performs follicular flushing via the external path (single-lumen needle); if you need the flushing to run through a second inner lumen of the needle, consider the DOUBLECYTE-N double-lumen family (Refs. 770308/TL in 17G and 770301/TL in 16G).
- Without follicular flushing: if you only require suction, the MONOCYTE single-lumen series (Ref. 770300 in 16G) or the MONOLUME range (Refs. 770311 in 17G and 770312 in 16G) cover this indication more economically.
- Good practices: single-use product sterilised by EtO; use it immediately after opening the individual packaging, check the integrity of the packaging and dispose of it in accordance with current legislation. For high volumes or framework agreements, consult our technical-commercial team.





Reviews