Description
1. Introduction
The Injector Plus (Ref. 730226) is a disposable uterine injector designed by the Italian manufacturer RI.MOS for diagnostic procedures of hysterosalpingography (HSG) and sonohysterography. Its semi-rigid catheter system with distal balloon allows controlled injection of the contrast medium, obtaining clear images of the uterine cavity and the patency of the fallopian tubes. A reliable, single-use solution for assisted reproduction units, gynaecology clinics and radiology departments, supplied in a box of 10 units.
2. Technical explanation of the equipment
This device is specifically designed for the injection of radiopaque contrast medium or methylene blue during hysterosalpingography. The set consists of a rigid introducer with sliding stop that guides catheter placement, and a semi-rigid radiopaque double-lumen catheter: one lumen is used for contrast injection and the other for inflation of the distal oval balloon of 2.5 cc. When inflated, the balloon occludes the internal cervical os, fixes the catheter in the correct position and minimises reflux of contrast into the vagina, favouring uterine distension and visualisation of the tubes. The steel stylet facilitates passage through the cervix and provides rigidity during insertion; the stopcock regulates balloon inflation, the anti-reflux clamp prevents fluid leakage and the female Luer-Lock connector ensures a secure connection with the included 2.5 cc syringe. The entire set is supplied sterile, individually packaged and ready for single-use.
3. Technical Specifications
The system integrates all the elements needed to perform the procedure quickly, safely and reproducibly.
| Component | Technical description |
|---|---|
| Catheter (1) | Semi-rigid, radiopaque, double-lumen (2 channels); outer Ø 2 mm (6 Fr); distal oval balloon of thermoplastic rubber, 2.5 cc |
| Rigid introducer (2) | With adjustable sliding stop to guide placement |
| Syringe (3) | 2.5 cc capacity for balloon inflation |
| Valve / stopcock (4) | Regulates inflation of the distal balloon |
| Anti-reflux clamp (5) | Retaining collar that prevents fluid reflux |
| Connector (6) | Female Luer-Lock for injection of the contrast medium |
| Stylet (7) | Steel, facilitates insertion through the cervix |
Technical data
| Parameter | Value |
|---|---|
| Catheter outer diameter | 2 mm (6 Fr) |
| Configuration | Semi-rigid radiopaque double-lumen |
| Distal balloon | Oval, thermoplastic rubber, 2.5 cc |
| Fixation system | Balloon inflation + anti-reflux clamp |
| Connection | Female Luer-Lock |
| Compatible media | Radiopaque contrast or methylene blue |
| Application | Hysterosalpingography (HSG) and sonohysterography |
| Sterilisation | Ethylene oxide (EtO), single-use, individually packaged |
| Marking | CE 0051 |
| Presentation | Box of 10 units |
4. Equivalences and Cross-References
This reference corresponds to the Injector Plus from manufacturer RI.MOS (Italy), whose factory code is the same 730226, packaged in boxes of 10 units. Due to its double-lumen distal balloon design, it occupies the same functional segment as catheter sets for HSG/hysterosonography with balloon from other manufacturers such as CooperSurgical, Argon Medical or Bexen Medical. Within the Cahnos catalogue, it is part of the family of uterine injectors for gynaecological imaging diagnostics alongside other references in the same range.
5. Selection tips and variants of interest
Gauge selection: the Injector Plus of 6 Fr (2.0 mm) offers a balance between insertion rigidity and patient comfort, being especially useful when a guided cervical passage with the stylet is required. If a fully flexible catheter is preferred (greater anatomical adaptation, less trauma), the SOFTINJECTOR of 7 Fr should be considered. For exclusively ultrasound studies of the cavity (sonohysterography) with a smaller gauge, the MININJECTOR of 5 Fr is an alternative, and for completely rigid placement the RIGIDINJECTOR of 6 Fr. In all cases, it is recommended to confirm the contrast medium to be used (water-soluble iodinated for HSG or saline / ultrasound gel for sonohysterography) and verify the compatibility of the Luer-Lock connector with the injection route of the equipment.