Description
1. Introduction
The RIGIDINJECTOR 6 Fr (Ref. 720225) is the reference rigid uterine injector for hysterosalpingography (HSG) in diagnostic gynaecology, radiology departments and assisted reproduction units. Its rigid radiopaque catheter allows controlled, stable and reproducible introduction through the cervical canal, ensuring precise injection of the contrast medium in each examination. It is supplied in a box of 10 units with individual sterile packaging, ready for single use: the practical and reliable solution required by centres with high turnover of diagnostic material.
2. Technical explanation of the equipment and procedure
The RIGIDINJECTOR is designed for the injection of radiopaque contrast media or methylene blue during hysterosalpingography, and also accepts physiological saline solution in sonohysterography studies. It belongs to the manufacturer’s RIGIDINJECTOR range (refs. 720225, 720239 and 720251 in 6, 5 and 7 Fr), whose specifications are verified in the official technical data sheet covering the three references. The set consists of a rigid introducer with mobile positioner (sliding stop) that is adjusted approximately 1 cm below the uterine depth measured by prior hysterometry, and a rigid radiopaque double-lumen catheter (outer Ø 2 mm in the 6 Fr model): one lumen is used for contrast injection and the other for inflation of the distal oval balloon made of thermoplastic rubber, 2.5 cc.
When inflated, the balloon occludes the internal cervical os, fixes the catheter in position and minimises reflux into the vagina, favouring uterine distension and correct visualisation of the fallopian tubes. The stopcock regulates balloon inflation (2 cc of air, according to the manufacturer’s instructions), the anti-reflux clamp prevents fluid leakage during injection, and the female Luer-Lock connector ensures a secure connection with the included 2.5 cc syringe. Single-use device: use only after hysterometry, under strict medical supervision and only if the packaging is intact.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Device type | Disposable rigid uterine injector for hysterosalpingography |
| Reference | 720225 (RIGIDINJECTOR family: refs. 720225, 720239, 720251) |
| Catheter calibre | 6 Fr (French); outer Ø ≈ 2 mm |
| Catheter configuration | Rigid, radiopaque, double-lumen (2 channels) |
| Distal balloon | Oval, thermoplastic rubber, 2.5 cc; recommended inflation: 2 cc of air |
| Introducer | Rigid with mobile positioner (adjustable sliding stop) |
| Valve / stopcock | Regulates balloon inflation |
| Anti-reflux system | Clamp that squeezes the tube |
| Connection | Female Luer-Lock connector for injection of the contrast medium |
| Syringe included | 2.5 cc, for balloon inflation |
| Compatible media | Radiopaque contrast, methylene blue; physiological saline for sonohysterography |
| Application | Hysterosalpingography (HSG) and sonohysterography |
| Sterilisation | Ethylene oxide (EtO); single-use; sterile individual packaging |
| Marking | CE 0051 |
| Presentation | Box of 10 units |
Set components
| Component | Description |
|---|---|
| Catheter (1) | Rigid radiopaque double-lumen, outer Ø 6 Fr (2 mm), with distal oval balloon of thermoplastic rubber, 2.5 cc |
| Rigid introducer (2) | With movable positioner / adjustable sliding stop |
| Syringe (3) | 2.5 cc capacity for balloon inflation |
| Valve / stopcock (4) | Regulates balloon inflation |
| Anti-reflux clamp (5) | Prevents reflux of the contrast liquid |
| Female Luer-Lock connector (6) | For injection of the contrast medium |
4. Equivalences and Cross-References
Reference 720225 corresponds to the RIGIDINJECTOR range, verified against the manufacturer’s official technical data sheet (rigid radiopaque double-lumen catheter, 2.5 cc thermoplastic rubber balloon, CE marking 0051). There is no official equivalence table validated by the manufacturer that cross-references this reference with those of other brands: the cited references belong to the same functional segment (intrauterine contrast injection with balloon for hysterosalpingography), but should be considered comparable by category and not interchangeable 1:1, since they differ in material, catheter configuration (most are flexible) and set components. Within the RIGIDINJECTOR family itself, the 5 Fr (Ref. 720239) and 7 Fr (Ref. 720251) models share design, materials and fixation system: exact reference equivalence within the same manufacturer, varying only in calibre.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| CooperSurgical | H/S Catheter Set (hysterosonography and hysterosalpingography), 7 Fr | 61-5007 | Comparable by category (not interchangeable 1:1) | Latex-free Kraton balloon catheter with integrated stylet; flexible configuration as opposed to the rigid double-lumen, without integrated stopcock or anti-reflux clamp. Verified on CooperSurgical’s official website. |
| MedGyn Products | HSG Catheter with wide balloon and rigid insertion sheath, 7 Fr (box of 10) | 022728 | Comparable by category (not interchangeable 1:1) | Latex-free wide balloon with rigid insertion sheath and inflation syringe; without double-lumen system with stopcock and anti-reflux clamp as an integrated set. Verified on MedGyn’s official website. |
| Marina Medical Instruments | HSG-Sono Catheter SureFlex, 7 Fr, 5 cc balloon with 6 cc syringe (box of 10) | MM1-607 | Comparable by category (not interchangeable 1:1) | Latex-free flexible catheter with 5 cc balloon and 6 cc syringe; different balloon volume, components and configuration. Verified on Marina Medical’s official website. |
| Cook Medical | Cook® Silicone Balloon HSG Catheter, 7 Fr × 30 cm, open tip | J-CHSG-703000 | Comparable by category (not interchangeable 1:1) | Silicone balloon with included 3 cc syringe; without rigid introducer with mobile positioner or adjustable sliding stop. Verified on Cook Medical’s official website. |
| Argon Medical Devices | HSG Catheter with latex-free concentric balloon, open distal tip that accepts 0.035-inch guidewire, 7 Fr × 40 cm (box of 10) | 660007040 | Comparable by category (not interchangeable 1:1) | Latex-free soft concentric balloon and open tip; 7 Fr calibre versus 6 Fr and without adjustable sliding stop system. Verified on Argon Medical’s official website and in GUDID. |
Bexen Medical markets a flexible-configuration HSG catheter (radiopaque polyurethane sheath, soft latex-free balloon and syringe included) with no public reference code, which must be requested directly from the manufacturer; other brands in the same functional segment are Prince Medical (PM-CARE family, sealed by balloon or vacuum cup, e.g. PM-CARE 2V-CUP ref. PS3460845) and Bioteque, whose gauge and fixation-system equivalents must be verified against their technical data sheets.
There is no equivalence table validated by the manufacturer; consult the technical department before using in tenders or as a product substitute. Exact gauge, length and fixation-system equivalents must always be verified against each brand’s technical data sheets and instructions for use.
5. Selection tips and variants of interest
RIGIDINJECTOR family (rigid catheter): the range is completed with the 5 Fr RIGIDINJECTOR (Ref. 720239, Ø ≈ 1.67 mm), indicated for a narrow cervical canal or patients with greater sensitivity, and the 7 Fr version (Ref. 720251, Ø ≈ 2.33 mm), which provides greater insertion rigidity and stability in cervical fixation. This 6 Fr model (Ø 2 mm) represents the intermediate balance of the family.
Variants according to clinical need: if a flexible catheter with greater anatomical adaptation and less trauma is preferred, the SOFTINJECTOR family (Refs. 720226, 720240 and 720250 in 6, 5 and 7 Fr) offers a radiopaque double-lumen catheter with a 2.5 cc balloon. For a semi-rigid system with steel stylet and double lumen with balloon, consider the Injector Plus (Ref. 730226). For exclusively ultrasound studies of the cavity (sonohysterography) with a smaller gauge, the 5 Fr MININJECTOR (Refs. 770314 and 880314) offer a finer profile, and if active manipulation of the uterus is required in addition to injection, the HANDLIN injector-manipulator (Ref. 720199).
Recommendations for use: always perform prior hysterometry to avoid perforation of the uterine fundus, place the catheter approximately 1 cm below the maximum measured depth, inflate the balloon with only 2 cc of air, confirm the contrast medium to be used (water-soluble iodinated contrast for HSG, methylene blue or saline solution according to protocol) and verify the compatibility of the Luer-Lock connector with the injection route of the equipment. Single-use device: do not reuse, use only if the packaging is intact and destroy after use in accordance with current regulations.