Description
1. Introduction
The MININJECTOR CM (Ref. 880314) is a disposable single-use uterine injector specifically designed for sonohysterography: the instillation of physiological saline solution into the uterine cavity under transvaginal ultrasound guidance. Its 5 Fr catheter with malleable stylet or plunger, rigid introducer and adjustable anti-reflux positioner allow precise, stable and minimally traumatic placement even in patients with a difficult-to-access cervical canal or stenosis. It is supplied in a box of 10 units with individual sterile packaging, ready for use in gynaecology clinics, assisted reproduction units and diagnostic radiology services.
2. Technical explanation of the equipment and procedure
Sonohysterography (also called saline infusion sonohysterography or SIS) combines transvaginal ultrasound with the slow instillation of physiological saline into the uterine cavity: the fluid distends the cavity and acts as an ultrasound contrast agent, allowing identification of polyps, submucosal fibroids, adhesions or endometrial thickening with far greater clarity than conventional ultrasound.
To this end, the MININJECTOR CM incorporates a transparent catheter with a radiopaque line along its entire length and a centimetre scale on the distal portion, which helps the clinician control the depth of placement. The soft tip with a single distal orifice allows atraumatic outflow of the saline. The set includes a rigid introducer that facilitates passage through the cervical canal and an adjustable positioner made of soft plastic material that acts as a stop: as it advances over the external cervical os, it prevents fluid reflux and ensures stable retention throughout the examination. The CM version adds a malleable plunger or mandrel (stylet) that provides temporary rigidity to the catheter during insertion, particularly useful in cases of cervical stenosis; once placed, the stylet is removed, the syringe is connected to the female Luer-Lock connector and the saline solution is injected slowly.
3. Technical Specifications
The specifications of the MININJECTOR CM (Ref. 880314) are summarised in the following table:
| Parameter | Specification |
|---|---|
| Device type | Uterine injector for sonohysterography (saline solution infusion) |
| Reference | 880314 |
| Catheter calibre | 5 Fr (French; outer Ø ≈ 1.67 mm) |
| Catheter | Transparent, with radiopaque line along its entire length and centimetre measurement scale on the distal part |
| Tip | Soft, with a distal orifice for the outflow of the saline solution |
| Introducer | Rigid, facilitates introduction into the cervix |
| Positioner | Adjustable, made of soft plastic material; acts as a fixation stop and prevents reflux |
| Stylet / plunger | Malleable stylet included, to facilitate insertion into the uterine cavity (cervical stenosis) |
| Connection | Female Luer-Lock connector for syringe |
| Compatible medium | Physiological saline solution (saline) |
| Application | Sonohysterography / saline infusion under transvaginal ultrasound |
| Sterilisation | Sterile, single-use, individually packaged |
| Presentation | Box of 10 units |
Set components:
| Component | Technical description |
|---|---|
| Catheter | Transparent 5 Fr, with radiopaque line along its entire length and distal centimetre scale |
| Distal tip | Soft, with a distal orifice for atramatic saline exit |
| Introducer | Rigid, facilitates passage through the cervix |
| Positioner | Adjustable, made of soft plastic; acts as a cervical stop and prevents reflux |
| Stylet / plunger | Malleable stylet included for insertion; removed after placement |
| Connection | Female Luer-Lock connector for the physiological saline syringe |
4. Equivalences and Cross-References
There is no official equivalence table validated by the manufacturer that cross-references 880314 with other brands. The references listed below belong to the same functional segment (5 Fr catheter for sonohysterography/hysterosonography, with stylet or cervical fixation system), but they are not interchangeable 1:1: they differ mainly in the fixation system (distal balloon versus adjustable positioner) and in the configuration of the set.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| CooperSurgical | H/S Catheter Set (hysterosonography and hysterosalpingography), 5 Fr, with integrated stylet and placement sheath | 61-5005 | Comparable by category (not interchangeable 1:1) | Catheter with latex-free distal balloon and integrated stylet; verified in the official technical data sheet of CooperSurgical and in the GUDID regulatory database (DI 60888937016477). The MININJECTOR CM does not incorporate a balloon: fixation is achieved by means of an adjustable positioner with rigid introducer. |
| MedGyn Products | HSG Catheter with rigid insertion sheath, 5 Fr, latex-free | 022726 | Comparable by category (not interchangeable 1:1) | Catheter with wide balloon and rigid insertion sheath; includes syringe for balloon inflation; code consistently verified in the official MedGyn data sheet and catalogues of authorised distributors. Balloon fixation configuration versus the stop system of the MININJECTOR CM. |
| Cook Medical | Goldstein Sonohysterography Catheter, 5.3 Fr × 26 cm, with adjustable acorn positioner | G17149 | Comparable by category (not interchangeable 1:1) | Balloon-free catheter with adjustable acorn positioner that reduces saline reflux at the external cervical os — a concept similar to the stop of the MININJECTOR CM — and a closed rounded tip; calibre 5.3 Fr and length 26 cm. Verified on the official Cook Medical website and catalogues of authorised distributors. |
Bexen Medical also markets a 5 Fr HSG catheter (flexible radiopaque polyurethane sheath, soft latex-free balloon 3 mm from the tip and syringe included) without a public reference code in its official data sheet; the exact reference must be confirmed with the manufacturer before any substitution. There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders or as a product substitute.
5. Selection tips and variants of interest
Same family without stylet: if the cervical canal does not present access difficulty and a more economical format is preferred, the 5 Fr MININJECTOR (Ref. 770314) offers the same saline infusion function without a malleable stylet.
For hysterosalpingography (HSG) with contrast medium and fixation balloon: the SOFTINJECTOR family (Refs. 720240, 720226 and 720250 in 5, 6 and 7 Fr) offers a flexible double-lumen catheter with a 2.5 cc balloon, and the RIGIDINJECTOR family (Refs. 720239, 720225 and 720251) the rigid catheter version. The Injector Plus (Ref. 730226, 6 Fr semi-rigid with steel mandrel) and the HANDLIN injector manipulator (Ref. 720199) complete the range for specific needs of guided injection with mandrel or active uterine manipulation.
Usage recommendations: always perform prior hysterometry to avoid contact with the uterine fundus, lubricate the tip before insertion, remove the stylet once the catheter is placed and inject the saline solution slowly. Verify the compatibility of the Luer-Lock connector with the service syringe. Single-use device: do not reuse, use only if the packaging is intact and dispose of after use in accordance with current regulations.