Description
1. Commercial Introduction
The SP-8800 syringe pump (Ref. GM35205) is a high-performance infusion and perfusion device designed for hospital and veterinary clinical settings. It stands out for single-gesture syringe installation, its colour LCD screen and ease of use, making it ideal for intensive care, anaesthesia, oncology, paediatrics, emergency and hospitalisation services. It is supplied with a rechargeable battery and is ready for immediate use, ensuring treatment continuity even during transfers or power cuts.
2. Technical explanation of the equipment
The SP-8800 administers drugs and fluids at a controlled flow rate using syringes of different sizes. Its sliding fixation system allows the syringe to be loaded in approximately five seconds, while the silicone keypad and shortcut keys speed up the configuration of frequently used functions. The device incorporates advanced clinical functions such as K.V.O. (open vein maintenance), purge, bolus and anti-bolus: when the infusion reaches the preset value, the flow rate is automatically reduced to 0.1 ml/h to prevent obstruction of the vascular access due to blood clotting.
Standard compatibility with the main syringe brands (BD Plastipak and Precise, Terumo, Rays, Pentaferte and Kova) allows the manufacturer to be selected from the device itself, with calibration capacity for up to ten different syringes per pump. The simultaneous visual and audible alarm system (LED and buzzer) warns of occlusion, end of infusion, low battery, malfunction or syringe disengagement, and allows the occlusion pressure level (high, medium or low) to be adjusted according to the patient’s age and condition. It also includes keypad lock, configuration memory, history and optional dose calculation based on patient weight, injection mass and dilution.
3. Technical Specifications
Product data
| Manufacturer | GIMA |
|---|---|
| Model | SP-8800 |
| Manufacturer reference | 35205 |
| Sales unit | 1 pc |
| Device type | Medical device (medical device) |
| Device class | II B |
| CND | Z12030302 |
Flow rate and volume ranges
| Parameter | Value |
|---|---|
| Flow rate range – 10 ml syringe | 0.1 – 300 ml/h |
| Flow rate range – 20 ml syringe | 0.1 – 400 ml/h |
| Flow rate range – 30 ml syringe | 0.1 – 500 ml/h |
| Flow rate range – 50 ml syringe | 0.1 – 1,200 ml/h |
| Flow rate error margin | ± 2 % |
| Infusion volume limit | 0.1 – 999.9 ml (0.1 ml interval) |
| Estimated total volume | 0.1 – 999.9 ml (0.1 ml interval) |
| K.V.O. flow rate (vein maintenance) | 0.1 – 10 ml/h adjustable |
| Rapid flow rate (purge) | 0.1 – 1,200 ml/h adjustable |
| Bolus | 0.1 – 1,200 ml/h; 0.1 – 1,000 ml |
Physical and power supply characteristics
| Characteristic | Specification |
|---|---|
| Dimensions | 220 x 130 x 108 mm |
| Weight | 2 kg (battery included) |
| Power supply | AC 230 V, 50/60 Hz and DC 9.6 V; internal DC 12/24 V – 2 A adapter |
| Switchable voltage | 110 V / 60 Hz – with adapter 35204 |
| Maximum power consumption | 45 VA |
| Electrical protection class | Class II B, internal power source device, type CF |
| Watertightness level | IPX3 |
| Battery type | Rechargeable Ni-MH |
| Battery life | 4 hours |
| Recharge time | Minimum 6 hours |
| Battery service life | Approximately 6 months |
Alarms
Occlusion, end of infusion (K.V.O.), standby, syringe almost empty, low battery, device malfunction (CPU error), disengagement of the forceps/lever/slider and syringe disengagement.
Special functions
Keypad lock, shortcut keys, configuration memory, remaining time, alarm repetition, open system, K.V.O., purge, bolus, anti-bolus, occlusion level and history.
4. Equivalences and Cross-References
There is no validated equivalence table with other brands for this syringe pump. Reference GM35205 corresponds to the manufacturer’s model SP-8800 (ref. 35205). The device offers standard compatibility with syringes from five major brands: BD (Plastipak, Precise), Terumo, Rays, Pentaferte and Kova, and can be calibrated for any other syringe brand upon sending 5 samples, with an approximate availability of 60 days. Calibration is prepared for 10, 20, 30 and 50 ml syringes. As a related accessory from the same family, adapter 35204 is referenced, necessary for operation at 110 V / 60 Hz. Manual available in German and Italian in the factory version.
5. Selection tips and variants of interest
Select the syringe size appropriate for the required flow rate: the SP-8800 covers from 0.1 ml/h with 10 ml syringes up to 1,200 ml/h with 50 ml syringes. Verify the compatibility of the syringes in your facility: the pump accepts profiles from the main brands and up to ten different calibrations. For installations with a 110 V / 60 Hz supply, request adapter 35204. Adjust the occlusion pressure level (high, medium or low) according to the patient’s age and condition and activate the K.V.O. function in prolonged infusions to prevent occlusion of the vascular access. The device is suitable for both human and veterinary use. Consult our technical-commercial department about the custom calibration system, especially for tenders, bids and fleet orders.