Description
1. Product Presentation
The ELASPUMP® elastomeric infusion pump of 275 ml (Ref. 093005/41) is the ideal solution for the continuous and controlled administration of medications in hospitals, clinics, home hospitalisation units and outpatient care centres. It is a single-use, sterile device manufactured free of DEHP, PVC and latex, which allows drugs to be infused at a constant and calibrated flow rate of 2.5 ml/hour without relying on electronic pumps, batteries or maintenance, with a very competitive cost per treatment and no risk of cross-contamination between patients. Its ease of handling makes it the practical choice for both healthcare professionals and patients themselves, who retain their autonomy throughout the infusion.
2. Technical description and operating principle
The ELASPUMP® pump consists of a silicone elastomer contained within a hermetic, rigid and transparent housing that prevents accidental leaks and allows visual inspection of the drug level from the outside without having to open the device. Its operating principle is based on the elastic recovery pressure of the elastomer, which drives the medication through the tubular system at a constant flow rate calibrated in the factory at 2.5 ml/hour, resulting in an approximate autonomy of about 4 days and 14 hours (110 hours) with the full volume of 275 ml. Its technical elements include:
- Silicone elastomer: encapsulated in a hermetic, rigid and transparent housing, with visual level inspection and no accidental leaks.
- Closed-circuit one-way filling valve: with a flat bottom, compatible with a standard syringe, which facilitates filling of the elastomer.
- Hydrophobic filter on male luer-lock cap: allows priming of the tubular device once the elastomer has been filled without removing the cap, ensuring greater internal sterility until actual use.
- In-line particle filter: retains impurities and ensures the purity of the administered drug.
- Removable sliding clamp: easy to remove when necessary.
- Colour-coded label: with the flow rate printed for immediate and simple identification of the flow to the patient.
- Patient label: designed for easy personalisation of the treatment.
- Full traceability: variable data and product code printed on the external housing and on the pump body.
3. Technical Specifications
| Characteristic | Specification |
| Brand | ELASPUMP® |
| Reference | 093005/41 |
| Capacity | 275 ml |
| Flow type | Fixed continuous (factory-calibrated) |
| Infusion flow rate | 2.5 ml/hora |
| Approximate duration | ≈ 4 days + 14 hours (approx. 110 hours at full volume) |
| Drive mechanism | Silicone elastomer in rigid and transparent casing |
| Filling system | Closed-circuit one-way valve, compatible with standard syringe |
| Connection | Male Luer-lock with cap and hydrophobic filter |
| Filtration | Hydrophobic filter + in-line particle filter |
| Routes of administration | Intravenous, arterial, epidural and subcutaneous |
| Clinical indications | Chemotherapy, pain therapy, antiviral and antibiotic treatments |
| Power supply | Does not require battery or electrical power source |
| Use | Single-use, sterile |
| Composition | DEHP-free, PVC-free, latex-free |
Indicative values calculated at full volume; the actual duration may vary slightly depending on the conditions of use and the characteristics of the drug.
Indicative infusion duration by flow rate (volume of 275 ml):
| Flow rate | Approximate duration |
| 1 ml/hora | 11 days + 11 hours |
| 2 ml/hora | 5 days + 18 hours |
| 2.5 ml/hour (current product) | 4 days + 14 hours |
| 5 ml/hora | 2 days + 7 hours |
| 7 ml/hora | 1 day + 15 hours |
| 8 ml/hora | 1 day + 10 hours |
| 10 ml/hora | 1 day + 3 hours |
| 125 ml/hora | 0 days + 2 hours |
| 250 ml/hora | 0 days + 1 hour |
4. Equivalences and Cross-References
The ELASPUMP® 275 ml continuous flow (Ref. 093005/41) belongs to the range of elastomeric pumps of the ELASPUMP® family and is complementary to the following references available at Cahnos:
| Reference | Description | Equivalence level |
| 093005/8/B | ELASPUMP® 275 ml with UV protection, continuous flow 10 ml/hour (1 day + 3 hours) | Comparable by category (not interchangeable 1:1) |
| 093005/2/B | ELASPUMP® 275 ml with UV protection, continuous flow 1 ml/hour (11 days + 11 hours) | Comparable by category (not interchangeable 1:1) |
| 093005/250 | ELASPUMP® 275 ml high continuous flow, 250 ml/hour (0 days + 1 hour) | Comparable by category (not interchangeable 1:1) |
| 093005/M1 · M2 · M3 | ELASPUMP® 275 ml variable flow (4 and 8 speeds), DEHP-free | Comparable by category (not interchangeable 1:1) |
| 093004/41/B | ELASPUMP® 200 ml with UV protection, continuous flow 2.5 ml/hour (3 days + 8 hours) | Comparable by category (same flow rate, smaller volume) |
| 093002/5/B | ELASPUMP® 100 ml with UV protection, continuous flow 4 ml/hour (1 day + 1 hour) | Comparable by category (lower flow rate and volume) |
| 093001/4 | ELASPUMP® 60 ml continuous flow, 2 ml/hour (1 day + 6 hours) | Comparable by category (lower flow rate and volume) |
Note on external cross-references: due to its 275 ml capacity, its pressurised elastomer system with luer-lock connector, hydrophobic filter and fixed calibrated flow rate, this device is comparable at category level with disposable continuous-flow elastomeric pumps from other manufacturers used in ambulatory infusion, patient-controlled analgesia and home intravenous therapy, such as the Easypump® ST/LT families from B. Braun, Baxter’s Intermate-type disposable elastomeric infusers and other continuous infusion elastomer systems of 250-275 ml available on the market. However, there is no equivalence table validated by the manufacturer with other brands; 1:1 interchangeability must be confirmed by comparing dimensions, material, flow rate and accessories in the official technical data sheet before use in tenders. Please consult our technical-commercial department to verify compatibility with your usual reference supplier.
5. Selection tips and variants of interest
- Volume and duration: the 275 ml version at 2.5 ml/hour is ideal for long-duration therapies of approximately 4 days and 14 hours at full volume. For smaller volumes, versions of 60 ml (Ref. 093001/4), 100 ml (Ref. 093002/5/B) and 200 ml (Ref. 093004/41/B, same flow rate of 2.5 ml/h) are available.
- Required flow rate: if your protocol is fixed, this continuous flow version at 2.5 ml/hour offers the best cost-benefit ratio. If you need to adjust the speed during treatment or protocol different regimens, consider the variable flow versions of 275 ml (Refs. 093005/M1, M2 and M3).
- Photosensitive drugs: for light-sensitive medications, always choose the ELASPUMP® versions with UV protection (Refs. 093002/5/B and 093005/8/B).
- Rapid infusions: if you need to administer the volume in a short time, consider the high continuous flow versions of 275 ml (Refs. 093005/125 and 093005/250).
- Recommended accessories: combine the pump with 1-way or 3-way infusion extension lines with a stopcock to adapt the system to the patient’s needs.
- Purchase conditions: contact our technical-commercial department to request samples, technical data sheets or special conditions for tenders and hospital purchases.