Description
1. Commercial Introduction
The MONOLUME (Ref. 770312) needle is a precision instrument designed for oocyte aspiration in in vitro fertilisation cycles and for aspiration of ovarian cysts under ultrasound guidance. Its single-lumen construction, echogenic tip with atraumatic sharpening and universal connector ensure smooth penetration, stable ultrasound visualisation and safe recovery of follicular material, minimising tissue trauma. Presented in a box of 10 units, it is the professional supply solution for fertility clinics, assisted reproduction units and gynaecology practices seeking a sterile, single-use product with CE marking.
2. Technical explanation of the equipment and procedure
The MONOLUME needle is a single-lumen, single-channel needle intended for atraumatic egg retrieval in in vitro fertilisation procedures, as well as for ultrasound-guided drainage of ovarian cysts. The echogenic tip with atraumatic sharpening allows extraction with smooth and easy tissue penetration, offering good visibility on the ultrasound scanner monitor. It incorporates a universal connector for multiple connections and a female luer lock port that acts as a handle and allows a secure connection to the syringe or suction line. Two lateral wings aligned with the tip bevel help the operator to orient themselves during the penetration phase. The gauge is identified by the luer colour code: light blue for 16G (this model) and transparent for 17G.
In clinical practice, after preparing the vaginal area and the sterile field, the needle is attached to the ultrasound probe and inserted into the vagina following the trajectory on the monitor to the follicle or cyst, proceeding with gentle aspiration or drainage. It is a single-use product, sterilised by ethylene oxide (EtO) with CE marking 0051, and may contain traces of nickel.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference (SKU) | 770312 |
| Brand / family | MONOLUME |
| Product type | Oocyte and ovarian cyst aspiration needle |
| Configuration | Single channel, single lumen (single-conduit simple needle) |
| Gauge | 16G (standard outer diameter ≈ 1.65 mm; light blue luer) |
| Length | 330 mm |
| Tip | Echogenic with atraumatic sharpening |
| Connection | Female luer lock port (handle) + universal connector for pump/syringe |
| Orientation | Two lateral wings aligned with the tip bevel |
| Gauge identification | Luer colour code: light blue (16G) / transparent (17G) |
| Sterilisation | Ethylene oxide (EtO) — CE 0051 marking |
| Use | Single-use (non-reusable) |
| Presentation | Box of 10 units, individually packaged |
| Indication | Oocyte aspiration (IVF) and aspiration of ovarian cysts |
| Warning | May contain traces of nickel; do not use if the packaging is damaged |
4. Equivalences and Cross-References
Below are listed references from other manufacturers verified against official sources (manufacturer catalogue, technical data sheets and regulatory databases) for the same indication: ultrasound-guided oocyte aspiration with a single-lumen needle in 16G gauge and length close to 330 mm. None of them are interchangeable 1:1 without technical validation of each centre’s protocol.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Cook Medical | Ovum Aspiration Needle Single Lumen (OSN), bevel B, regular wall | K-OSN-1633-R-B-90 (G55480) | Comparable by category (not interchangeable 1:1) | 16G × 33 cm (330 mm) with 90 cm suction line; EchoTip® echogenic tip and handle with bevel orientation notch. |
| CooperSurgical | Wallace® Single Lumen Oocyte Needle | ON1633 | Comparable by category (not interchangeable 1:1) | Single-lumen 16G needle, non-coring tri-bevel tip with echogenic marking, presentation in box of 10 units (GUDID); nominal length 33 cm according to model nomenclature. Versions with tube: ONS1633-500. |
| Rocket Medical | Embryon® SX™ Single Lumen Aspiration Needle | R57601-SX-90 | Comparable by category (not interchangeable 1:1) | 16G × 33 cm (330 mm) with 90 cm suction tube; high-definition ultrasound spiral bevel and handle with orientation marker. Dimensions equivalent to those of the MONOLUME. |
| Kitazato | Single Lumen OPU Needle with Luer lock connector | 316001 | Comparable by category (not interchangeable 1:1) | 16G × 32.5 cm (325 mm), single lumen with luer lock connection for manual suction; MDR CE marking (2797). Length 5 mm shorter than the MONOLUME. |
There is no equivalence table validated by the manufacturer for 1:1 cross-reference with other brands; consult the technical department before using in tenders.
5. Selection tips and variants of interest
- Gauge selection: the 16G model (Ref. 770312) offers a larger internal diameter and therefore greater suction flow; if your protocol requires a slightly finer needle with less trauma, consider the MONOLUME 17G version (Ref. 770311).
- Single lumen vs. follicular flushing: if in addition to aspirating you need to irrigate the follicle, consider the BIMONOCYTE family (Refs. 770304/770302 and versions with tube /TL) or the DOUBLECYTE double-lumen (Refs. 770301/770308 and /TL versions).
- Pump connection: for integration into pump suction systems with 80 cm extension and luer lock connector, consider the MONOCYTE-N series (Refs. 770300/TL and 770307/TL).
- Good practices: single-use product; use immediately after opening the package, verify the integrity of the packaging and dispose of it according to current legislation. For high volumes or framework agreements, consult our technical-commercial team.





Reviews