1. Introduction
The Dilapan-S 4×65 is the reference hygroscopic cervical dilator for obstetrics and gynaecology departments seeking progressive, predictable and tissue-friendly cervical dilation. Thanks to its patented AQUACRYL® hydrogel, it offers non-pharmacological cervical ripening with excellent patient tolerance and a safety profile proven in clinical practice.
This presentation in a box of 25 units is designed for maternity units, family planning clinics, and hospitals with high patient volume: each dilator is supplied in a sterile individual package, ready to use, latex-free and single-use. Cahnos reference: 0929008 (manufacturer’s factory code: DS465).
2. Technical explanation of the procedure
Dilapan-S is a rigid rod of hydrophilic polymer which, when inserted into the cervical canal, acts through a triple mechanism of action:
- Mechanical: the controlled radial pressure it exerts on the cervical wall dilates the cervix progressively and safely.
- Biophysical: partial reversible osmotic dehydration softens the cervical tissue, facilitating dilation.
- Physiological: it promotes the release of endogenous prostaglandins, with collagen degradation and tissue remodelling.
It is indicated for cervical ripening prior to induction of labour, cervical preparation before termination of pregnancy, and other intrauterine procedures requiring prior dilation. The device must be removed before exceeding 24 hours in situ, and its marker thread allows correct placement to be verified. It should not be used in the presence of apparent clinical infection of the genital tract.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Brand / Model | DILAPAN-S 4×65 (factory reference: DS465) |
| Reference Cahnos (SKU) | 0929008 |
| Product type | Rapid-acting hygroscopic cervical dilator |
| Material | Patented AQUACRYL® hydrogel (active part), polypropylene handle, polyamide marker thread; latex-free |
| Initial diameter | 4 mm |
| Length | 65 mm |
| Final diameter | Up to 9 mm of progressive osmotic expansion (distributor data sheet); the manufacturer reports 10.0–11.2 mm at 4 h in an in vitro test (K143447) |
| Mechanism of action | Hygroscopic: mechanical, biophysical (osmotic) and physiological |
| Sterilisation | By irradiation (individually sterile packaged) |
| Use | Disposable, single-use. Do not re-sterilise or reprocess |
| Maximum in situ time | 24 hours (according to the manufacturer’s instructions for use) |
| Presentation | Box of 25 units |
| Expiry | 36 months in sealed packaging |
| Storage | +5 °C to +30 °C, in a dry place, protected from atmospheric agents and direct sunlight; do not freeze |
| Regulatory classification | Medical device class IIa, CE marking (Directive 93/42/EEC) |
| Factory certification | ISO 9001 |
| Biocompatibility | Non-cytotoxic, non-irritant, non-sensitising (tests according to ISO 10993) |
| Application | Obstetrics and gynaecology |
Range and references of the family (Cahnos catalogue):
| Reference | Model | Presentation |
|---|---|---|
| 0929006 | DILAPAN-S 4×55 | Box of 10 units |
| 0929011 | DILAPAN-S 4×55 | Box of 25 units |
| 0929007 | DILAPAN-S 4×65 | Box of 10 units |
| 0929008 | DILAPAN-S 4×65 | Box of 25 units (product of this data sheet) |
Expected dilation (one 4 mm unit, manufacturer’s in vitro data, IFU DILAPAN-S):
| In situ time | Expected diameter |
|---|---|
| 2 hours | 7.8 – 10.0 mm |
| 4 hours | 10.0 – 11.2 mm |
| 6 hours | 10.1 – 12.5 mm |
| 24 hours | 12.7 – 14.6 mm |
4. Equivalences and Cross-References
The Dilapan-S 4×65 belongs to the segment of osmotic/hygroscopic cervical dilators: devices that expand by absorbing fluids from the cervical tissue. The Cahnos reference 0929008 corresponds to the official factory code DS465 (Dilapan-S 4×65, box of 25 units) and, in the US market, to the code 23170314810T (25 units/box), according to the manufacturer’s official ordering guide.
There is no equivalence table validated by the manufacturer for this reference; the cross-references listed below come from each brand’s official catalogues and are functional comparisons, not interchangeable 1:1. Consult the technical department before using in tenders.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| MedGyn Products (USA) | Natural laminaria (tentacles of the alga Laminaria digitata) | 021004 (4 mm × 60 mm, box of 25 units) | Comparable by category (not interchangeable 1:1) | Same indication for osmotic/hygroscopic cervical dilation, but natural material (seaweed) versus synthetic AQUACRYL hydrogel; length 60 mm versus 65 mm; slower expansion and smaller diameter; being of biological origin, it presents a different handling and risk profile. Reference verified in MedGyn’s official catalogue. |
Other products in the same osmotic segment: Lamicel (Medtronic Xomed), a synthetic osmotic dilator with a historical reference in the US market (PMA P820075), with no active order code verifiable in public sources; pending confirmation with the manufacturer before any substitution. In the Cahnos catalogue, the single-use mechanical dilators ComfiDilator (ref. D52230-S 3-4 mm and ref. D52231-S 5-6 mm) and the double Hegar dilator in stainless steel (reusable mechanical) cover the same clinical indication with an immediate dilation mechanism different from the osmotic one, and therefore should not be considered 1:1 interchangeable.
5. Selection tips and variants of interest
- Choose 4×65 mm (refs. 0929007 / 0929008) when a longer working length is required, such as in cervical preparation prior to pregnancy termination; the box of 25 units (0929008) is the most cost-effective option for services with high patient volume.
- Choose 4×55 mm (refs. 0929006 / 0929011) for cervical ripening prior to term labour induction and standard cervical preparation in gynaecology.
- The 3×55 mm model is indicated when the 4 mm one cannot be inserted (early pregnancy) or when removal must be performed in less than 4 hours, according to the manufacturer’s instructions for use.
- Check the packaging before each use: do not use the product if the individual sterile packaging is damaged. It is single-use; discard it after the procedure. Placement and removal must follow the manufacturer’s instructions (do not rotate during extraction; do not exceed 24 hours in situ).
- Main contraindication: apparent clinical infection of the genital tract. Always consult the complete instructions for use before application.
- Support material: complete the procedure with specula, Pozzi forceps and sterile aqueous lubricant or saline for optimal and safe patient management.