Description
SuperPass® Percutaneous Insertion Kit for Vascular Cannulation
Five essential components · Dilation from 8 Fr to 28 Fr · J-tip guides of 0.025″ and 0.038″
1. Commercial Introduction
The SuperPass® percutaneous insertion kit from MicroPort® Surgical is the complete solution for vascular cannulation in cardiovascular surgery and procedures involving an extracorporeal circuit. It brings together in a single sterile tray the five essential components for establishing vascular access in a controlled, safe and reproducible manner: puncture needle, dilator, scalpel, aspiration syringe and metal guide.
With a dilation range from 8 Fr to 28 Fr, single- and two-stage dilators, and J-tip guidewires of 0.025″ and 0.038″, the SuperPass® range covers everything from paediatric to adult patient needs, with seven kit configurations to tailor the contents to the intended cannula size. A CE-marked medical device, for use only on medical prescription, designed for the professional who demands precision and standardisation in the operating theatre.
2. Technical explanation of the equipment and procedure
Percutaneous vascular cannulation is the essential preliminary step for connecting the patient to the extracorporeal circuit (cardiopulmonary bypass, ECMO or other support systems). The SuperPass® kit integrates all the instruments needed to perform access in a controlled manner:
- 18 G puncture needle: allows initial puncture of the vessel with minimal trauma and serves as a reference for the rest of the sequence.
- J-tip metal guidewire (0.025″ and 0.038″): advances safely inside the vessel, reducing the risk of dissection or endothelial damage.
- Single- and two-stage dilators: progressively dilate both the percutaneous tract and the vascular channel up to the intended cannula size (from 8 Fr to 28 Fr).
- 11# scalpel: allows enlarging the cutaneous entry point when a larger-calibre access is required.
- 10 mL aspiration syringe: facilitates verification of correct placement and blood flow before inserting the final cannula.
By grouping all the elements in a single sterile kit, the technique is standardised, the number of manipulations is reduced, surgical times are optimised and the risk of field contamination is minimised.
3. Technical specifications and reference tables
| Description | Percutaneous insertion kit for vascular cannulation |
| Family | SuperPass® |
| Components | Puncture needle, dilator, scalpel, suction syringe and metal guidewire |
| Function | Establish the percutaneous access route and dilate the vascular channel |
| Dilation range | From 8 Fr to 28 Fr |
| Dilators | Single-stage and two-stage |
| Guides | J-tip of 0.025″ and 0.038″ |
| Puncture needle | 18 G |
| Syringe | 10 mL |
| Scalpel | 11# |
References and ordering information
Seven kit configurations to adjust the contents to the intended cannula size:
| Reference | Needle | Syringe | Scalpel | Guide | Dilator |
|---|---|---|---|---|---|
| D25-14 | 18 G | 10 mL | 11# | 0.025″ | Two-stage: 8/10 Fr, 12/14 Fr |
| D38-14 | 18 G | 10 mL | 11# | 0.038″ | Two-stage: 8/10 Fr, 12/14 Fr |
| D38-18 | 18 G | 10 mL | 11# | 0.038″ | Two-stage: 8/10 Fr, 12/14 Fr, 16/18 Fr |
| D38-22 | 18 G | 10 mL | 11# | 0.038″ | Two-stage: 8/10 Fr, 12/14 Fr, 16/18 Fr, 20/22 Fr |
| D38-24 | 18 G | 10 mL | 11# | 0.038″ | Two-stage: 8/10 Fr, 12/14 Fr, 16/18 Fr, 20/22 Fr · Single-stage: 24 Fr |
| D38-26 | 18 G | 10 mL | 11# | 0.038″ | Two-stage: 8/10 Fr, 12/14 Fr, 16/18 Fr, 20/22 Fr · Single-stage: 24 Fr, 26 Fr |
| D38-28 | 18 G | 10 mL | 11# | 0.038″ | Two-stage: 8/10 Fr, 12/14 Fr, 16/18 Fr, 20/22 Fr · Single-stage: 24 Fr, 26 Fr, 28 Fr |
Regulations and documentation
| Product | Medical device — exclusive use under medical prescription |
| European Regulation | Regulation (EU) 2017/745 on medical devices |
| National regulations | Royal Decree 192/2023, regulating medical devices |
| CE marking | Attached as a separate document |
| Documentation | Indications, contraindications, warnings and instructions for use in the instructions for use accompanying the product |
4. Equivalences and Cross-References
The SuperPass® family consists of seven equivalent references (D25-14, D38-14, D38-18, D38-22, D38-24, D38-26 and D38-28), which share the same 18 G puncture needle, 10 mL syringe and 11# scalpel, differing in guidewire size (0.025″ or 0.038″) and in the set of dilators included. This modularity allows the entire cannulation sequence planned in cardiovascular surgery to be covered: the reference D25-14 is associated with a fine 0.025″ guidewire for paediatric access, while the D38 series (0.038″) covers from 14 Fr to 28 Fr depending on the number of dilators included.
In the Cahnos catalogue, the clinician can complement this kit with the rest of the consumables and medical devices for vascular access, drainage and percutaneous kits for critical pathway procedures. For the selection of an alternative reference, attention must be paid to the size of the final cannula, a criterion common to all percutaneous vascular cannulation systems on the market.
5. Selection tips and variants of interest
- Cannula size: select the reference whose dilator range includes the size (Fr) of the cannula to be implanted. For 14 Fr cannulas choose D25-14 or D38-14; for 18 Fr, D38-18; for 22 Fr, D38-22; for 24-28 Fr, D38-24, D38-26 or D38-28.
- Paediatric or adult guidewire: the 0.025″ J-tip guidewire (D25-14) is the reference option in paediatric patients and smaller vessels; the 0.038″ guidewire (D38 series) is the standard for adult patients and large vessels.
- Single- or two-stage dilators: two-stage dilators allow gradual progression through ascending sizes (e.g., 8/10 Fr, 12/14 Fr), while the final references in the series add single-stage dilators for single sizes of 24, 26 and 28 Fr.
- Logistics and standardisation: as this is a complete kit in a sterile tray, it is recommended to keep several sizes available in the cardiovascular surgery department to adapt to the surgical plan without delays.
- Compatibility: always verify the cannula size and the connector of the extracorporeal circuit with which the kit will be used before the procedure.