Description
1. Reusable electrosurgical pencil for the operating theatre
We offer you the reusable electrosurgical pencil (handpiece) Ref. F4244, a professional reference from the European high-frequency monopolar electrosurgery line. Designed for continuous use in operating theatres, major ambulatory surgery, minor surgery clinics and general surgery departments, it combines an ergonomic high-resistance plastic body (latex-free), activation push-buttons, a universal Valleylab-type connector and a 300 cm cable.
It is supplied in a box of 5 handpieces together with the reusable blade electrode F4060 (also in a box of 5 units), ready to be integrated into the department’s reusable instrument inventory. Its warranty of up to 30 sterilisation cycles offers a very competitive cost per procedure compared to disposable pencils, with consistent cutting and coagulation performance in every cycle.
2. Technical explanation of the equipment and procedure
In a monopolar electrosurgery system, the electrosurgical generator (ESU) sends high-frequency current through the pencil to the active electrode —the F4060 blade mounted in the handpiece—, which concentrates the energy at the point of tissue contact; the circuit is completed through the neutral (dispersive) plate placed on the patient. Depending on the waveform selected on the generator, the effect is cutting, coagulation or blended.
The handpiece incorporates manual control via cutting and coagulation push-buttons, whose actuation click provides immediate recognition of activation. The universal Valleylab-type connector allows use with the vast majority of electrosurgical generators on the market without the need for adapters (Covidien/Medtronic and other ESUs with Valleylab socket). The electrode shaft diameter of 2.38 mm (3/32″) is the industry standard for electrode holders, making it easy to replace with other tips from the reusable family, and the 300 cm cable provides ample movement over the surgical field.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Reference / SKU | F4244 |
| Product type | Reusable electrosurgical pencil (handle) with push buttons |
| System | Monopolar, cut and coagulation |
| Activation | Manual push buttons (cut / coagulation) |
| Cable length | 300 cm |
| Electrode shaft diameter | 2.38 mm (3/32″) – universal fit |
| Connector type | Universal tripolar, Valleylab type |
| Included electrode | Reusable blade F4060 (box of 5 units) |
| Body material | Sterilisable high-resistance plastic, latex-free |
| Number of sterilisations | Up to 30 cycles (handle) |
| Presentation | Box of 5 pieces, individually packed in heat-sealed polyethylene bag – non-sterile |
| Compatibility | Electrosurgical generators with Valleylab-type connector |
Technical note: the value 2.38 mm corresponds to the electrode shaft diameter (3/32″), according to the manufacturer’s official data sheet; the original data sheet expressed this measurement as «2.38 cm» referring to the electrode shaft.
Sterilisation and regulatory compliance
| Parameter | Specification |
|---|---|
| Handle sterilisation | Ethylene oxide or autoclave: up to 30 cycles |
| Autoclave – pre-vacuum | 134 °C / 2.05 bar, 12 min exposure (+ pre-vacuum and drying) |
| Autoclave – alternative | 121 °C / 1.05 bar, 20 min exposure |
| F4060 electrode sterilisation | Up to 20 cycles |
| Electrical safety | EN 60601-1 · EN 60601-2-2 · EN 60601-1-2 |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC and amendments) |
| Storage | 0 to +50 °C · relative humidity 20–80 % |
Variants of the reusable family with Valleylab-type connection
| Reference | Cable | Electrode | Sterilisations |
|---|---|---|---|
| F4244 (this product) | 300 cm | F4060 (box of 5 units) | Up to 30 |
| F4242 | 300 cm | F4060 | 100 |
| F4141 | 320 cm | F4060 (box of 5 units) | 30 |
| F4243 | 320 cm | F4060 | 100 |
| F4151 | 500 cm | F4060 | 30 |
| F4254 | 500 cm | F4060 | 100 |
4. Equivalences and Cross-References
Reference F4244 corresponds to the official catalogue of the European reusable electrosurgery line (electrosurgical pencil with push-buttons, Valleylab-type connector, 300 cm cable, F4060 electrode with 2.38 mm shaft and warranty of up to 30 sterilisation cycles), and Cahnos markets it without altering characters. The following equivalents are functional comparisons verified against the official catalogue of the manufacturer cited in each row and against regulatory databases (GUDID/FDA); they must not be treated as 1:1 interchangeable without validation by the technical department.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Gima (Italy) | ORION PLUS – autoclavable handpiece with dual push-button and screw clamp fixing, 3 m silicone cable and blade electrode | 30539 | Comparable by category (not interchangeable 1:1) | Code and attributes verified in the official catalogue of Gima: autoclavable handpiece with 2 push-buttons of the «click» type, 3 m silicone cable, blade electrode included, up to 100 autoclave cycles (electrode up to 20), compatible with electrode shaft Ø 2.4 mm, class II B and EAN 8033003262665. Differences from Ref. F4244: guaranteed sterilisation cycles (100 vs. 30), cable length (3 m vs. 300 cm) and electrode fixing system (screw vs. locking). |
| Bovie / Aspen Surgical (USA) | Reusable pencil with button-type push-button, autoclavable, 3-pin connection (electrode not included) | ESPR2 | Comparable by category (not interchangeable 1:1) | Code and attributes verified in the official catalogue of Aspen Surgical and in GUDID/FDA: reusable monopolar push-button pencil, standard 3-pin Valleylab-type connection, compatible with 3/32″ (2.38 mm) shaft electrodes and validated for up to 100 autoclave cycles; non-sterile, electrode not included. Differences from Ref. F4244: sterilisation cycles (100 vs. 30), electrode not included and primarily North American market. |
There is no equivalence table validated by the manufacturer that cross-references the F4244 reference with codes from other brands. On the market, reusable push-button pencils with Valleylab-type connection and 2.38 mm (3/32″) shaft are sold under brands such as Medtronic/Valleylab (Covidien), Erbe, ConMed and other European manufacturers, whose specific codes must be verified against the technical data sheet or IFU of each manufacturer before any substitution. There is no equivalence table validated by the manufacturer; consult with the technical department of Cahnos before use in a tender.
5. Selection tips and variants of interest
According to cable length: Ref. F4244 (300 cm) is the balanced option for standard operating theatres. If your team works with the generator away from the table, consider the 500 cm variant from the same family (F4151, 30 sterilisation cycles); if your priority is to extend the service life per handpiece, the versions with 100 sterilisation cycles (F4242 of 300 cm, F4243 of 320 cm or F4254 of 500 cm) offer greater durability.
Electrode replacements: the reusable family offers F4060 blades (2.38 mm shaft) and, for 4 mm electrode holders, the F4060/4 with handpieces F4246/F4256. Always verify the shaft diameter against the receptacle of your handpiece before placing the order; the same family includes ball and needle tips for contact coagulation and fine dissection.
Generator compatibility: the Valleylab-type connector of this pencil is compatible with the majority of electrosurgical generators on the market (Covidien/Medtronic and similar). For generators with a 4 mm or 3.2 mm socket, consult the reusable adapters from the range (F4830–F4838).
Maintenance in the operating theatre: use the sterile tip cleaner Tip Cleaner F7520 (box of 100 units) to remove eschar from the blade during the procedure and extend the service life of the electrode, and the sterile sheath F4545 to store the handpiece between uses. Record the number of sterilisation cycles (handpiece up to 30, electrode up to 20) and withdraw the handpiece from service at any sign of deterioration of the body or cable. For B2B volume conditions, hospital supply agreements or validation of equivalents from other manufacturers, contact the technical-commercial team at Cahnos.