Description
1. Product Presentation
This disposable electrosurgical unit kit from FIAB is the turnkey solution for the open surgery operating theatre: it includes the foot-control pencil Ref. F4390/5 with its removable blade electrode F4046 and the sterile sheath (holster) Ref. F4545, so that the surgical team has everything needed in a single sterile package, ready to use in each procedure. Foot pedal activation frees the surgeon’s hands, who maintains full control of cutting and coagulation with the foot without letting go of the handle or diverting attention from the operative field.
Presented in a box of 25 sterile units, it is the ideal option for hospitals, private clinics, ambulatory surgery centres and purchasing groups seeking a constant, safe supply with an optimised cost per procedure. Single-use product: do not re-sterilise or reuse.
2. Technical explanation of the equipment and procedure
The foot-control electrosurgical unit is a monopolar electrosurgery pencil: the radiofrequency current generated by the generator (ESU) travels through the cable to the active electrode —the removable blade F4046— and concentrates at the point of contact with the tissue, producing the desired surgical effect (cut, coagulation or mixed) and returning through the patient’s neutral (dispersive) plate.
Foot pedal activation: foot control of cutting and coagulation (hands-free), without manual buttons on the handle, ideal for prolonged manoeuvres and precision dissections. Removable blade F4046: stainless steel blade electrode with standard 2.38 mm (3/32″) shaft, compatible with the F404x replaceable electrode family. 500 cm cable: wide reach over the surgical field without tension, designed to distribute the table and move the pencil with total freedom. 3.2 mm connector: standard monopolar plug for quick and secure connection to the generator socket (F4835–F4838 adapters for other standards). Sterile sheath F4545: allows the pencil to be secured to the surgical field and kept always accessible during the procedure.
3. Technical Specifications
| Reference FIAB | F4390/5H |
|---|---|
| Packaging | Box of 25 individual units |
| Units per box (web) | 25 |
Technical data for the kit
| Characteristic | Specification |
|---|---|
| Product type | Disposable electrosurgery pencil (electrosurgical unit), single-use, in kit with sheath |
| System | Monopolar, cut and coagulation |
| Activation | Foot pedal control (hands-free) |
| Electrode | Removable stainless steel blade F4046 (blade) |
| Shaft diameter | 2.38 mm (3/32″) |
| Cable length | 500 cm |
| Connector | 3.2 mm monopolar plug |
| Included accessory | Sterile sheath (holster) F4545 |
| Sterility | Sterile, single-use disposable |
| Presentation | Box of 25 units |
F4390 family · disposable foot-control pencils (2.38 mm shaft)
| Reference | Cable | Connector | Electrode |
|---|---|---|---|
| F4390 | 320 cm | 3.2 mm | Blade F4046 |
| F4390/5 | 500 cm | 3.2 mm | Blade F4046 |
| F4390/A | 320 cm | 3.2 mm | Needle F4048 |
| F4390/P | 320 cm | 3.2 mm | F4044 ball electrode (Ø 4 mm) |
| F4390/L | 320 cm | 3.2 mm | Blade F4050 |
| F4390/AL | 320 cm | 3.2 mm | F4052 needle electrode |
| F4390/PL | 320 cm | 3.2 mm | Sphere F4067 |
| F4390/4M | 320 cm | 4 mm plug | Blade F4046 |
4. Equivalences and Cross-References
The reference F4390/5H corresponds to the official factory construction of the kit formed by the disposable foot-control pencil F4390/5 (500 cm cable, 3.2 mm connector and F4046 blade) with the sterile sheath F4545 (holster). No equivalence table validated by the manufacturer covering this reference has been found in public sources; the following are functional comparisons verified against official catalogues and regulatory databases (GUDID/FDA), not 1:1 equivalences.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Observations |
|---|---|---|---|---|
| Aspen Surgical (Bovie) | Electrosurgical Foot Control Pencil (with sheath) | ESP7 / ESP7H, ESP7HS | Comparable by category (not interchangeable 1:1) | Disposable foot-control pencil, monopolar, sterile, with blade electrode and presented with holster-type accessories; the connector standard, cable length and presentation differ. |
| Medline Industries | Pencil, Cautery, Foot Switch, PTFE Blade | ESFS3002 | Comparable by category (not interchangeable 1:1) | Disposable cauterisation pencil with foot pedal activation and mounted electrode; the electrode coating (PTFE), cable length and connector standard differ. |
For a 1:1 replacement, always verify field by field: electrode type (blade), shaft diameter (2.38 mm / 3/32″), cable length (500 cm), connector type of the installed generator (3.2 mm vs 4 mm or Valleylab), inclusion of the F4545 sheath, sterility and presentation. Consult the technical department before use in tenders.
5. Selection tips and variants of interest
Check your generator connector: this pencil mounts a 3.2 mm monopolar connector; if your ESU is of another standard (Valleylab, 4 mm, Erbe), choose the variant with the specific connector (e.g. F4390/4M with 4 mm plug) or use the reusable adapters from the F4835–F4838 range.
This kit provides the maximum reach of the family with a 500 cm cable and incorporates the sterile sheath F4545 to secure the pencil to the field. If you prefer the kit without the long cable or with other electrode geometries, the same family offers the F4390/H variants (with F4545 sheath and F4046 blade), as well as the kits with F7520 tip cleaner (F4390/75 and F4390/5-75). For other electrode geometries, the family is completed with the simple pencil F4390 (F4046 blade, 320 cm), F4390/A (F4048 needle), F4390/P (F4044 sphere), F4390/L (F4050 blade), F4390/AL (F4052 needle) and F4390/PL (F4067 sphere).
As complementary accessories from the electrosurgery catalogue, the sterile tip cleaner F7520 is recommended to maintain electrode performance by removing eschar during the procedure. Single-use product: do not re-sterilise or reuse. For institutional volumes or tenders, consult our technical-commercial department.




Reviews