Description
1. Product Presentation
The disposable electrosurgery pencil FIAB Ref. F4390/5WB is the handle (electrosurgical unit) that every open surgery operating theatre needs when looking for hands-free control: foot pedal activation allows selection of cut and coagulation without releasing the instrument or interrupting the surgical procedure. It is supplied in a box of 50 sterile single-use units, ready to open for each procedure, and is supplied without electrode included, so that each service can incorporate its own family of replaceable tips (blades, needles or spheres) with a standard 2.38 mm (3/32″) shaft. With its 500 cm cable and 3.2 mm monopolar plug connector, it is the ideal option for hospitals, private clinics, ambulatory surgery centres and purchasing centres looking to standardise the handle, stock rotation and optimised cost per procedure.
2. Technical explanation of the equipment and procedure
The foot pedal control pencil is the active electrode of a monopolar electrosurgery system: the radiofrequency current generated by the electrosurgical generator (ESU) travels through the cable to the active electrode that the surgeon mounts in the handle housing, concentrates at the point of contact with the tissue producing the desired effect (cut, coagulation or mixed) and returns through the neutral (dispersive) plate placed on the patient.
Foot pedal activation: cut and coagulation control is performed with the foot, without manual buttons on the handle, ideal for prolonged procedures and precision dissections. Without electrode: as no mounted tip is included, the user chooses the electrode geometry (blade, needle or sphere) most suitable for each intervention from the replaceable electrodes with 2.38 mm (3/32″) shaft. 500 cm cable: maximum reach around the operating table, without tension or need to reposition the equipment. 3.2 mm plug connector: quick and safe monopolar connection to the standard generator socket; for other standards (Valleylab, 4 mm, Erbe) the reusable adapters from the FIAB range are used. Sterile single-use eliminates the risk of cross-contamination and the costs of cleaning and re-sterilisation.
3. Technical Specifications
Available references:
| Characteristic | Value |
|---|---|
| Reference FIAB | F4390/5WB |
| Product type | Disposable electrosurgery pencil (electrosurgical unit), single-use, foot pedal control |
| System | Monopolar, cut and coagulation |
| Activation | Foot pedal control (hands-free) |
| Electrode | Without electrode included; accepts replaceable electrodes with 2.38 mm (3/32″) shaft |
| Connector | 3.2 mm monopolar plug |
| Cable length | 500 cm |
| Sterility | Sterile, single-use disposable |
| Packaging | Box with 50 units (50 units) |
Compatibility: the 3.2 mm connector fits the standard monopolar socket of most electrosurgical generators on the market. For generators with another connection standard (Valleylab, 4 mm, Erbe), consult the reusable adapters from the FIAB range F4835–F4838.
4. Equivalences and Cross-References
The reference F4390/5WB belongs to the FIAB F4390 family of disposable foot pedal control pencils with housing for a 2.38 mm shaft, in its configuration with 500 cm cable and without electrode. No equivalence table validated by the manufacturer covering this specific reference has been found; the following are functional comparisons verified at category level, not interchangeable 1:1:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences |
|---|---|---|---|---|
| Aspen Surgical (Bovie) | Electrosurgical Foot Control Pencil (disposable foot pedal control pencil with blade electrode) | ESP7 | Comparable by category (not interchangeable 1:1) | Same type of product (disposable foot pedal control pencil, monopolar, sterile, single-use, box of 50 units). Differs: includes mounted stainless steel blade, 1-pin connector instead of the 3.2 mm plug and shorter cable length. |
| Medline Industries | Pencil, Cautery, Foot Switch (cauterisation pencil with foot switch) | ESFS3002 | Comparable by category (not interchangeable 1:1) | Disposable cauterisation pencil with foot pedal activation and mounted electrode. Differs: includes PTFE-coated blade compared to the version without electrode and different connector standard. |
For a field-by-field replacement, check: connector type of your generator (3.2 mm vs 4 mm, 1 pin or Valleylab), cable length (500 cm), electrode shaft diameter accepted (2.38 mm / 3/32″), sterility and presentation.
5. Selection tips and variants of interest
Check your generator connector: this pencil mounts a 3.2 mm monopolar plug connector; if your ESU is of another standard (Valleylab, 4 mm, Erbe), use the reusable adapters from the FIAB range F4835–F4838 or consider the variants with specific connector. Within the F4390 family, the /5 suffix indicates 500 cm cable and the complementary letters identify the factory-mounted electrode: F4390 (blade F4046, 320 cm), F4390/5 (blade F4046, 500 cm), F4390/A (needle F4048), F4390/P (sphere F4044), F4390/L (blade F4050), F4390/AL (needle F4052) and F4390/PL (sphere F4067), all with 2.38 mm shaft.
As this is the version without electrode (F4390/5WB), combine this handle with the family of disposable electrodes with 2.38 mm shaft —blades F4046/F4050, needles F4048/F4052 and spheres F4044/F4067— to adapt the tip to each surgical technique. As complementary accessories, the sterile sheath F4545 (to fix the pencil to the field) and the tip cleaner F7520, which removes the eschar and maintains electrode performance during the procedure, are recommended. Single-use product: do not re-sterilise or reuse. For institutional volumes or tenders, consult our technical-commercial department.