Description
Paediatric foam return plate with TAB connection
The paediatric foam electrosurgical plate (Ref. F7805P) is the essential neutral return electrode for completing the monopolar circuit of any high-frequency electrosurgery equipment when operating on paediatric patients. With a 148 x 90 mm support and 78 cm² of active surface, it is a model without REM system and without cable, with tab (TAB) connection: it works with the generator’s own reusable plate cable, a configuration that reduces the cost per procedure and facilitates warehouse management.
It is presented in a box of 25 units with individual heat-sealed packaging that protects the conductive hydrogel until the moment of use: the option chosen by hospitals, clinics and ambulatory surgery centres that need reliable, homogeneous consumption with permanent stock for their paediatric surgical scheduling.
How the return plate works and why it is important
In any monopolar electrosurgical unit procedure, the high-frequency current leaves through the active electrode (handle or pencil), passes through the patient’s tissue and must return to the generator via a low current density path. The neutral plate collects this current in a controlled manner: if the contact is not correct or the surface is insufficient, the current density at the return point increases and, with it, the risk of thermal burn to the patient.
This plate is constructed with a polyethylene foam backing (PE foam) and conductive hydrogel adhesive, which favours adaptation to the paediatric patient’s body contour and uniform contact with the skin. It belongs to the without REM system series: it does not activate the generator’s return monitoring control (REM), so it must be used with conventional equipment without the requirement of a split plate, always selecting the lowest possible energy level. As it is a cable-free plate, the professional must use a reusable plate cable (F7902 universal 6.3 mm, F7903 Valleylab type or F7904 for Erbe) that connects to the plate’s tab.
Technical specifications
| Characteristic | Value |
|---|---|
| Reference / SKU | F7805P |
| Product type | Disposable return plate (electrode) for monopolar electrosurgery |
| Population | Paediatric |
| Measurements | 148 x 90 mm |
| Active surface | 78 cm² |
| Support | Polyethylene foam |
| Conductive adhesive | Hydrogel |
| Safety system | Without REM (does not activate the generator’s REM control) |
| Cable | Not included (tab connection) |
| Connection | TAB (tabs) |
| Use | Single-use (disposable, non-sterile) |
| Latex-free | Yes |
| Presentation | Box of 25 units with individual packaging |
| Regulatory classification | Medical device with CE marking |
Paediatric range of the foam family (148 x 90 mm): all references share the same foam backing and hydrogel adhesive; the difference lies in the REM system and the mode of connection to the generator.
| Reference | Measurements | Active surface | REM | Cable | Connection |
|---|---|---|---|---|---|
| F7805P | 148 x 90 mm | 78 cm² | No | No | TAB (tabs) |
| F7805PW/V | 148 x 90 mm | 72-78 cm² | No | 300 cm | Valleylab |
| F7820P | 148 x 90 mm | 72 cm² | Yes | No | TAB (tabs) |
| F7820PW/V | 148 x 90 mm | 72 cm² | Yes | 300 cm | Valleylab |
Equivalences and cross-references
There is no official equivalence table validated by the manufacturer that cross-references the F7805P reference with codes from other brands. The comparisons shown below are functional comparisons verified against official documentation from each cited manufacturer (catalogue, technical data sheet and GUDID/FDA regulatory databases) and should not be treated as 1:1 interchangeable without validation from the technical department.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Nissha Medical Technologies (ex Graphic Controls) | SWAROPLATE NON-Split Diathermy Pad, Children, non-REM connector (Valleylab-NONREM) | 1262H / 6000840N (1262H-SWARO-V1-8-96) | Comparable by category (not interchangeable 1:1) | Structurally very similar option: non-split plate (without REM) for children (> 5 kg and < 15 kg), polyethylene foam backing (PE foam) and conductive hydrogel adhesive. The dimensions differ (144 x 90 mm vs 148 x 90 mm), the adhesive surface (72 cm² vs 78 cm² active surface) and the connection mode (Valleylab-NONREM connector with 3 m cable vs TAB without cable). Code and attributes verified in the manufacturer’s official catalogue. |
| Medtronic / Covidien (Valleylab) | Valleylab™ PolyHesive™ II Infant Patient Return Electrode with cable (REM) | E7510-25 (UDI 10884524001043) | Comparable by category (not interchangeable 1:1) | Return plate for paediatric/infant patient (< 13.6 kg / 30 lb) with integrated 2.7 m (9 ft) cable and Valleylab-type connector. The monitoring system differs (REM/split plate vs without REM), the substrate (PolyHesive™ II adhesive hydrogel vs polyethylene foam) and the connection (integrated cable vs TAB without cable). Code verified in GUDID/FDA and on the manufacturer’s official website. |
| Leonhard Lang (Skintact® Cool Contact) | Paediatric non-split (unsplit) neutral electrode, with 3 m cable and 6.3 mm jack connector | RS06B30 | Comparable by category (not interchangeable 1:1) | Non-split plate (without contact monitoring) for paediatric patients of 5-15 kg, low-impedance hydrogel on flexible foam, with a power limit of 120 W. The conductive surface differs (76 cm² vs 78 cm²) and the connection mode (3 m cable with 6.3 mm jack connector vs TAB without cable). Code and attributes verified in the manufacturer’s official technical documentation (RS datasheet, Leonhard Lang GmbH). |
The cited references have been cross-checked against official catalogues, technical data sheets and regulatory databases of each cited manufacturer. Although they coincide in function (return without REM control or paediatric single-use), they differ in substrate, exact measurements, active surface or connection system, so they are not 1:1 interchangeable. Homologous brands and families that must be evaluated with their technical data sheet in hand before any substitution: AESCULAP (B. Braun) Patient Plates; PolyHesive series from Medtronic-Covidien / Valleylab; Nissha Medical Technologies (SWAROPLATE); Leonhard Lang (Skintact); BOWA Medical return plates; EMED neutral electrodes; GIMA disposable plates for electrosurgical units; Taktvoll Patient Return Electrode. In all cases, it must be checked field by field: exact dimensions, active surface in cm², backing and type of hydrogel, REM system, physical TAB connector and compatibility with the installed electrosurgical generator. Do not assume equivalence based only on the generic product name.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
Selection tips and variants of interest
- Check your generator connection: the F7805P plate without cable with TAB connection requires a reusable plate cable (F7902 universal 6.3 mm, F7903 Valleylab type or F7904 for Erbe). Choose the cable according to the plug of your electrosurgical unit.
- Prefer a plate with integrated cable? The F7805PW/V reference offers the same plate body (148 x 90 mm) with a 300 cm cable and Valleylab-type connection, for direct use without an additional cable.
- Does your generator have REM control? The equivalent alternative in the same size (148 x 90 mm) with REM system and TAB connection is the F7820P; with a 300 cm cable and Valleylab connection, the F7820PW/V.
- Other versions of the same family without REM: adult F7805 (202 x 101 mm, 132 cm², TAB) and neonatal F7805N (88 x 74 mm, 36 cm², TAB).
- Extended compatibility: adapters F4828 (from Valleylab-type connection to 4 mm plug) and F4829 (to 6.3 mm plug) for other generators.
- Application: always place on dry, intact, hair-free skin, on a well-perfused muscle area away from bony prominences. Check the integrity of the individual packaging and the expiry date before each use: it is a single-use, non-sterile product. Do not reuse or cut the plate.