Description
1. Introduction
The ALSATOM SU 50-MPC electrosurgical unit (order reference B700) is a high-frequency electrosurgical unit designed for monopolar and bipolar surgical procedures with exceptional performance and intuitive handling. Manufactured in Italy by ALSA Apparecchi Medicali under the Alsatom brand, this 80 W unit offers continuous and pulsed emission currents that minimise the thermal effect on tissues and reduce surgical smoke, enabling very fine cuts similar to those of radiofrequency units, and very delicate coagulations. It is the reference solution for dermatology, dentistry, ENT and minor surgery practices, as well as for veterinary clinics and general surgery departments looking for a reliable, compact unit with exclusive pedal activation.
2. Technical explanation of the equipment
The SU 50-MPC generates high-frequency currents (450 kHz) controlled by a microprocessor with power-on self-diagnosis, lock-out in the event of technical faults via error codes (MPC System) and automatic self-regulation of the emitted power according to the tissue (ADC System), without the need for periodic calibrations. The front panel has a current selector and power control with micro-adjustment of 1 W between 0 and 30 W and increments of 2 W beyond that value, with independent cut and coagulation settings stored and restored at power-on.
It incorporates standard and pulsed currents (PURE, P PULSED, BLEND, B PULSED 1, B PULSED 2, MICRO, M PULSED, FULG, F PULSED and BIPOLAR) that allow from very fine pure cuts to delicate coagulation with adjustable pulses from 0.1 to 1 s and fulguration, also under liquid. The neutral plate safety circuit verifies the connection and contact with tissues (dual-field plates), blocking emission with an alarm in the event of poor contact. Activation is via a simple pneumatic, waterproof and explosion-proof foot switch (STOP/PN, no electrical current in the pedal), making it suitable for operating theatres and environments with a risk of explosive atmospheres. Cooling is by convection, without a fan, and the unit belongs to protection class I with an isolated type CF output circuit protected against defibrillation.
3. Technical Specifications
Data verified according to the manufacturer’s official technical data sheet (T.D.S. ALSATOM SU-MPC models, revision F 02-2019).
| Characteristic | Value |
|---|---|
| Reference (SKU) | SU 50-MPC |
| ALSTATOM order reference | B700 (without accessories) |
| Operating mode | Monopolar and bipolar |
| Maximum output power | 80 W (pure cut PURE, 500 Ω) / 80 W (bipolar, 100 Ω) |
| Working frequency | 450 kHz (±10 %) |
| Power supply | 230/110 V ~ 50/60 Hz, 370 VA |
| Classification and type / output circuit | Class I, type CF, isolated circuit protected against defibrillation (HF leakage < 150 mA) |
| CE classification / standard | Class IIb, CE0051 (No. 187/MDD), in accordance with IEC 601-1 and IEC 60601-2-2 (MDD) |
| Control system | MPC System: microprocessor with self-diagnosis and lockout (Error codes) |
| Output power control | ADC System: automatic with dynamic real-time control and lock-out (Output error) |
| Power regulation | 1 W from 0 to 30 W; then increments of 2 W (settings stored and restored at power-on) |
| Neutral plate safety | Connection and good contact monitoring with power lock-out and alarm indication |
| Cooling | By convection, without fan |
| Dimensions (L × D × H) | 25 × 24 × 12 cm |
| Weight | 4.5 kg |
| Activation | Simple pneumatic foot switch, sealed and explosion-proof (STOP/PN) |
| Applications | Dermatology, dentistry and dental surgery, ENT, general surgery and minor surgery, gynaecology, maxillofacial surgery, ophthalmology, gastroenterology, plastic and aesthetic surgery and veterinary (also laparoscopy) |
| Presentation | Price per unit (equipment without accessories) |
Currents and powers of the SU 50-MPC model (according to official technical data sheet):
| Current | Power | Load | Clinical indication |
|---|---|---|---|
| PURE | 80 W | 500 Ω | Pure cut; also under liquid in small hysteroscopy |
| P PULSED | 40 W | 500 Ω | Pulsed pure cut, minimum thermal effect and less smoke (conisations, blepharoplasty) |
| BLEND | 80 W | 500 Ω | Coagulating cut |
| B PULSED 1 | 40 W | 500 Ω | Pulsed coagulating cut, lower thermal effect |
| B PULSED 2 | 33 W | 500 Ω | Slow pulsed coagulating cut (polypectomies) |
| MICRO | 80 W | 500 Ω | Delicate coagulation with few sparks |
| M PULSED | 40 W | 500 Ω | Microcoagulation with adjustable pulses from 0.1 to 1 s |
| FULG | 80 W | 750 Ω | Macro coagulation (fulguration), also under liquid |
| F PULSED | 40 W | 750 Ω | Fast pulsed fulguration, more delicate and less smoke |
| BIPOLAR | 80 W | 100 Ω | Bipolar coagulation: forceps, scissors, double-needle electrodes and laparoscopy |
4. Equivalences and Cross-References
The SU 50-MPC belongs to the ALSATOM SU-MPC series, whose models share the same construction platform, control system (MPC/ADC) and standard accessories (B700/A, B700/B and B700/D series). Within the same manufacturer, the variants differ in power, activation system and available currents: SU 100-MPC (B710) with 100 W in pure cut, recommended by the manufacturer for gynaecology, veterinary and underwater coagulation in the bladder; SU 140-MPC (B720) with 140 W and additional activation via a CUT/COAG hand switch; SU 140/D-MPC (B730) with 160 W max. and double pneumatic foot pedal; and SU 140/BD-MPC (B740), for exclusively bipolar use with saline vaporisation for arthroscopy.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Micromed (Germany/Italy) | MD100 – Electrosurgery unit | 150-100-001 | Comparable by category (not interchangeable 1:1) | Compact 100 W unit with monopolar cutting and coagulation (two cutting and two coagulation modes) and bipolar cutting and coagulation function with AUTO-START; adjustment from 1 to 25 W in 1 W increments; 500 kHz; class IIb, type CF. Higher power and no pulsed fulguration currents; verify compatibility of neutral plates, handpieces and foot switches. |
| Micromed (Germany/Italy) | RF 130 – 4 MHz (radiofrequency) | 150-130-001 | Comparable by category (not interchangeable 1:1) | 4 MHz, 130 W radiofrequency generator with radio wave technology (focused effect, minimal carbonisation); clean and sharp cutting, gentle/intense coagulation and bipolar function; class IIb, type BF. 4 MHz technology is not directly comparable 1:1 to conventional 450 kHz high-frequency electrosurgery. |
On the European market there are other electrosurgical units of similar power and applications from manufacturers such as KLS Martin (maxium series), Erbe (VIO), BOWA (ARC series), Bovie (Aaron series) and Eschmann. For a direct replacement it is essential to verify in their technical data sheets the power per current, the type of applied part (CF/BF), the compatibility of neutral plates, handpieces and foot switches, and compliance with current regulations.
There is no equivalence table validated by the manufacturer for equipment from other manufacturers; consult the technical department before using in tenders.
5. Selection tips and variants of interest
SU 50-MPC (B700): recommended option for minor surgery, dentistry, dermatology and ENT, with reduced powers of up to 80 W and exclusive control via foot switch. If your main activity is gynaecology, veterinary or underwater bladder coagulation, consider the SU 100-MPC (B710); for major surgery or advanced laparoscopy, the SU 140 series (B720, B730 and B740) offers higher power and additional controls.
Accessories and consumables: standard series B700/A (simple pneumatic foot switch STOP/PN, autoclavable electrode handpiece MPE/F with 2.5 m cable, electrodes E1 (×2), E5, E7, E12 ball Ø 2.5 mm, E14 ball Ø 4 mm, neutral plate EIP/9 in stainless steel with 2.5 m cable and elastic band FFE), B700/B with flexible conductive rubber neutral plate NP/GP and B700/D for dental use with manual neutral plate EIP/S and series of 8 dental electrodes SEL/D. As reusable neutral plates, the EIP/9, NP/GP and EIP/S can be used; for disposable dual-field plates, the connection cable CMS/E with adhesive plates EIP/SA (adults) and EIP/SP (paediatric).
Remember that the pneumatic foot switch does not carry electrical current, making it suitable for operating theatres with a risk of explosive atmospheres, and that the equipment does not require periodic calibrations thanks to its self-diagnosis. For institutional budgets, custom configurations, H10/AB trolleys or CS200 smoke evacuation systems, consult our technical-commercial team.





Reviews