Description
1. Product Presentation
The disposable electrosurgery electrode Ref. F4624 is a single-use consumable with a round loop tip of 15 × 10 mm, made of tungsten wire and mounted on a universal shaft of Ø 2.38 mm (3/32\”). It is sold in a box of 24 units, a format designed to ensure stock rotation and consumption planning in operating theatres, gynaecology units, colposcopy clinics and minor surgery rooms. It is the ideal reference for professionals seeking a reliable, sterile replacement electrode with optimised cost per procedure.
2. Technical explanation of the product and procedure
In monopolar electrosurgery, the active electrode concentrates the high-frequency current emitted by the electrosurgical generator; the energy is applied at the point of contact with the tissue and returns through the neutral plate or dispersive electrode. The round geometry of the loop delimits the contact area and allows controlled cutting and coagulation, extracting a tissue sample with defined margins and facilitating curved cuts on the cervical transformation zone.
Tungsten wire offers two clinical advantages over steel: it maintains the loop shape throughout the procedure even at high power settings and minimises carbonisation and lateral thermal damage, favouring the histological quality of the sample. The 15 × 10 mm profile (width 15 mm, depth 10 mm) provides an intermediate excision, with good width coverage and sufficient depth for the excision of the cervical transformation zone in LEEP/LLETZ techniques. The Ø 2.38 mm shaft corresponds to the compatibility standard of most electrosurgical pencils and handles on the market (equivalent to 3/32\”). Being a single-use product, the risk of cross-contamination and reprocessing costs are eliminated.
3. Technical Specifications
| Characteristic | Value |
|---|---|
| Tip shape | Round loop (circular) |
| Loop dimensions | 15 × 10 mm |
| Loop material | Tungsten |
| Shaft material | AISI stainless steel |
| Shaft diameter | 2.38 mm (3/32″) |
| Total length | 135 mm ± 3% |
| Insulation / protection | High-impact plastic (HIPS) anti-shock protector, latex-free |
| Type of use | Disposable, single-use |
| Sterility | Sterile, individually packaged (ethylene oxide) |
| Latex | Latex-free |
| Presentation | Box of 24 units |
| Application | Cutting and coagulation; cervical excision (LEEP/LLETZ) and soft tissue surgery |
| Compatibility | Electrosurgical pencils and generators with universal connection of Ø 2.38 mm (3/32″) |
| Regulatory classification | Medical device class IIb, in accordance with Directive 93/42/EEC and amendments |
| Expiry | 4 years from production date if stored correctly |
| Storage | Temperature 0 to +50 °C; relative humidity 20–80% |
| Reference | F4624 |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references the F4624 with codes from other brands. The following references are offered as a functional category comparison verified against the cited sources (official manufacturer or authorised distributor catalogue and GUDID regulatory database), never as direct 1:1 substitutes:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| AB Medica Group – LESSA | Disposable sterile tungsten loop electrode (round) | 88.200.74 | Comparable by category (not interchangeable 1:1) | Same dimensions 15 × 10 mm and round geometry, tungsten wire, sterile and individually packaged in a box of 24 units; distribution alternative in Spain with identical presentation in units per box. Verified in the official catalogue of distributor LESSA. |
| Bovie (Aspen Surgical) | Bovie® LLETZ Disposable Loop Electrode | ES11 | Comparable by category (not interchangeable 1:1) | Tungsten loop of 15 × 10 mm with standard 3/32\” (2.3 mm) shaft, sterile and single-use; pack of 5 units. Verified in GUDID (DI 10887482177966) and authorised distribution catalogues. Differs in units per pack. |
| MedGyn | Disposable Loop Electrode | 028068 | Comparable by category (not interchangeable 1:1) | Loop of 15 × 10 mm in tungsten wire on a 2.3 mm stainless steel shaft, sterile and disposable; pack of 5 units. Verified on the manufacturer’s official website. Differs in pack (5 units) and shaft diameter (2.3 mm). |
As comparable homologous families at category level (not 1:1 substitutes), verified against official catalogues of their manufacturers, the following lines of disposable tungsten loops for LEEP/LLETZ are identified: Megadyne (Ethicon group) — LLETZ Loops (refs. 0430 of 15 × 6 mm and 0460 of 15 × 12 mm), CooperSurgical — LEEP Medium-Radius Loop Electrodes (refs. R1505 of 15 × 5 mm, R1507 of 15 × 7 mm and R1515 of 15 × 15 mm) and DTR Medical — Colposcopy LLETZ Loop Electrodes (refs. LEC1131508-GY of 15 × 8 mm and LEC1131512-GN of 15 × 12 mm); in all cases with loop dimensions different from those of the F4624, therefore the exact geometry, shaft diameter and sterility must be checked in the manufacturer’s technical data sheet before any substitution.
There is no equivalence table validated by the manufacturer that allows these references to be treated as direct substitutes for the F4624; consult the technical department before use in tenders. Before substituting with another brand, compare shaft diameter, exact tip geometry (round, square or triangular), sterility and units per pack with the manufacturer’s data sheet.
5. Selection tips and variants of interest
To correctly choose the replacement electrode, first verify the shaft diameter accepted by your electrosurgical pencil: the standard Ø 2.38 mm (3/32\”) is the most common; there are Ø 4 mm versions that require an adapter. Next, select the loop geometry according to the technique: the round 15 × 10 mm profile offers a balance between width coverage and depth, ideal for excision of the cervical transformation zone; smaller loops (10 × 5, 10 × 7, 10 × 10 mm) are reserved for minor lesions, while the 20 × 10 mm format covers larger lesions and square, trapezoidal or triangular loops are used according to contour and access.
This same family also offers interesting variants in boxes of 24 units: F4620 (round 10 × 3 mm), F4621 (round 10 × 5 mm), F4622 (round 10 × 7 mm), F4623 (round 10 × 10 mm), F4625 (round 20 × 10 mm), as well as the square versions F4630 (10 × 4 mm), F4631 (10 × 8 mm) and F4632 (10 × 10 mm), the trapezoidal F4640 (30 × 5 mm), the inverted curved F4641 (23 × 13 mm) and the triangular F4642 (17 × 12 mm). For hospital purchases and tendering processes, always verify the CE marking under Regulation (EU) 2017/745 (MDR) and request special volume conditions from our commercial department.