Description
The IMAMI-20 from the Unicare IMAMI® series is a latest-generation digital fetal Doppler monitor that combines Bluetooth connectivity and a mobile app to provide precise, comfortable and safe monitoring of fetal well-being, both at home and in the clinic. Its compact and lightweight design, together with wireless synchronisation of readings with the mobile phone, allows access to real-time graphs, data history and automatic evaluation reports, providing peace of mind for expectant mothers and follow-up by healthcare professionals.
Technical explanation of the equipment
The device operates using continuous wave Doppler technology: the 2.5 MHz ultrasonic transducer emits waves that, when reflected by the movement of the fetal heart, generate an audible signal and a digital reading of the fetal heart rate (FHR). The mobile app, synchronised via Bluetooth, displays the heart rate curve in real time, automatically records fetal movements, generates alerts for abnormal heartbeats and produces automatic fetal evaluation reports. Data is stored with privacy protection and can be shared quickly and easily with family members or professionals. The integrated rechargeable lithium battery allows more than 120 minutes of continuous measurement, with battery and signal quality indicators for uninterrupted use.
Technical specifications
| Characteristic | Detail |
| Fetal heart rate range | 50 – 210 bpm |
| Resolution | 1 bpm |
| Accuracy | ±1 bpm |
| Nominal transducer frequency | 2.5 MHz (2.5 ±10% MHz) |
| Emission mode | Continuous wave Doppler ultrasound |
| Peak negative acoustic pressure | < 0.1 MPa |
| Output beam intensity (Iob) | < 20 mW/cm² |
| Spatial peak temporal average intensity (Ispta) | < 100 mW/cm² |
| Ultrasonic output power (Isata) | < 50 mW |
| Effective ultrasonic emission area | < 2.45 ±0.5 cm² |
| Connectivity | Bluetooth with included mobile app |
| App functions | Real-time FHR curve, trends, automatic fetal movement count, abnormal heartbeat alerts, automatic evaluation report, signal quality indicator |
| Power supply | Integrated rechargeable lithium battery |
| Battery life | More than 120 minutes of continuous measurement |
| Included accessories | App, headphones, Doppler base (charging and synchronisation), rechargeable lithium battery |
| Data security | Privacy and data security guaranteed |
The model supports two functional configurations: with a Doppler base (Dock) for wireless synchronisation of readings and battery charging, or with a mobile app for viewing and managing data on the phone.
Equivalences and cross-references
The IMAMI-20 belongs to the IMAMI® series from the manufacturer Shenzhen Unicare (family models with app and Bluetooth: IMAMI-10, IMAMI-20 and IMAMI-40), so it shares technology, app and usage criteria with the rest of the range. In our catalogue it can be cross-referenced with the IMAMI10 (same family with Bluetooth and mobile app) and with the IMAMI40 (2″ LED screen and 2.5 MHz transducer). As a professional-range alternative without connectivity, the Ecosound fetal Doppler with waterproof catheter (Ref. 35.71-2600) is recommended. For complete cardiotocographic monitoring in the delivery room, the EcoTwin LCD fetal monitors (Ref. 35.71-3480), Smart 1 (Ref. 35.11-7000) and Monitor Fetal 21A (Ref. MCF21A) are compatible in application. In the international market, devices with comparable characteristics are pocket fetal Dopplers such as Sonicaid/Huntleigh and portable ranges with app from other European and Asian manufacturers.
Selection tips and variants of interest
For home use and daily monitoring of pregnancy, the IMAMI-20 is the most complete option in the series due to its app, automatic reports and alerts. If you prefer direct on-screen viewing and greater robustness for the clinic, consider the IMAMI40, which adds a 2″ LED screen and the option of AA batteries. For centres requiring continuous CTG recording with ultrasound transducers and TOCO, a bedside fetal monitor (EcoTwin or 21A series) is recommended. Always remember to use specific ultrasound transmission gel and follow the indications of the healthcare professional; the device is designed for complementary self-checking and does not replace medical assessment or regulated prenatal tests.





Reviews
There are no reviews yet.