Description
1. Introduction
Expand the diagnostic capacity of your portable tympanometer Timpani (code 53560) with the REFLEX PLUS Licence (ref. 53563). This is the original software activation code that enables the ipsilateral acoustic reflex test, an essential test in audiological examination of paediatric patients and in the functional diagnosis of the middle ear. A simple solution, without additional hardware, to professionalise your ear, nose and throat (ENT) or audiology practice with a single software accessory.
2. Technical explanation
The acoustic reflex (or stapedial reflex) is the involuntary contraction of the stapedius muscle in response to a sound stimulus of sufficient intensity. Its recording allows objective assessment of the integrity of the auditory pathway and the middle ear, and is particularly useful in detecting conditions such as otosclerosis, hearing loss or brainstem dysfunctions. With this licence, the tympanometer Timpani performs ipsilateral stimulation at 226 Hz, i.e. the stimulus and measurement occur in the same ear, presenting stimuli at frequencies of 0.5, 1, 2 and 4 kHz. You can run the autothreshold (automatic threshold search) and fixed intensity modes, with a maximum intensity of 100 dB HL (ipsi) and a stimulus duration of 1 second.
3. Technical Specifications
Specifications of the acoustic reflex test (PLUS licence):
| Parameter | Specification |
|---|---|
| Enabled function | Acoustic reflex (acoustic reflex) |
| Stimulation | Ipsilateral at 226 Hz |
| Available tests | Autothreshold, fixed intensity |
| Stimuli | 0.5, 1, 2 and 4 kHz (with reflex Plus licence) |
| Maximum intensity | 100 dB HL (ipsi) |
| Stimulus duration | 1 second |
Product data:
| Value | Detail |
|---|---|
| Manufacturer | GIMA |
| Manufacturer reference | 53563 (licence) / 53560 (Timpani base unit) |
| Sales unit | 1 pc |
| Device type | No medical device |
4. Equivalences and Cross-References
This product is an original software licence. and has no equivalents from other manufacturers: its activation is exclusive to the portable tympanometer Timpani (code 53560). It should not be confused with standard ipsilateral reflex licences (which enable only the basic 226 Hz frequency); the PLUS variant extends the number of stimuli to 4 frequencies (0.5, 1, 2 and 4 kHz). As this is a proprietary activation code, its replacement by third-party solutions is not recommended without prior confirmation from the technical department.
5. Selection tips and variants of interest
Purchase this REFLEX PLUS licence (ref. 53563) if your centre already has the handheld tympanometer Timpani (ref. 53560) and wishes to incorporate the study of the ipsilateral acoustic reflex into its battery of audiological tests, especially in paediatric and preoperative assessment. Always verify the reference of the base unit before purchase, as activation is specific to this model. To complete the examination, the Timpani supports the usual tympanometry and screening audiometry tests already included as standard; for conditions of use, software configuration or compatibility queries, contact the technical-commercial department of Cahnos.