Description
1. Reliable power for your Welch Allyn VSM6000 monitor
Keep your Welch Allyn VSM6000 / Connex VSM 6000 vital signs monitor always operational with this replacement battery (Ref. Cahnos 88888849), designed for the Connex VSM 6000 platform and associated with the original reference BATT99. It provides a stable, rechargeable power source that allows continued measurement of non-invasive blood pressure, oxygen saturation, pulse rate and temperature even when the device is moved between consultations, between floors or during a power supply interruption.
A replacement solution designed for institutional purchasing: reliable, easy to install and available so the maintenance department always has a spare on hand and the monitor never goes out of service. In fleets with several VSM6000 units, having this backup battery avoids monitoring downtime and extends the useful life of the monitor fleet.
2. How this battery works in your monitor
The Welch Allyn VSM6000 series monitors (Connex VSM 6000) are compact bedside and transport devices that integrate a screen, parameter modules and signal processing electronics for continuous monitoring of vital signs. When operating without mains power —in emergency, triage, consultations or internal transfers—, the internal battery ensures continuity of monitoring and prevents restarts or loss of patient data.
This reference uses rechargeable Ni-MH (nickel-metal hydride) chemistry, with a nominal voltage of 10.8 V and a capacity of 6.75 Ah (6,750 mAh), which translates to 72.9 Wh of available energy. Once installed in the device compartment, the monitor automatically recharges it when connected to its power supply, without the need for external chargers or specialised tools. When battery life noticeably decreases or the device warns of low battery after a few hours of charging, it is time to replace the accumulator.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Application | Vital signs monitor Welch Allyn VSM6000 / Connex VSM 6000 (platform associated with original ref. BATT99) |
| Original reference (MFR) | 4500-84-1 (not verifiable in public sources for the VSM6000; see note in block 4) |
| Reference Cahnos (SKU) | 88888849 |
| Technology | Ni-MH (nickel-metal hydride) rechargeable |
| Nominal voltage | 10.8 V |
| Capacity | 6.75 Ah (6,750 mAh) |
| Energy | 72.9 Wh |
| Weight | Approximately 0.8 kg |
| Dimensions | 100 x 85 x 45 mm |
| Load | Automatic in the monitor (rechargeable) |
| Device manufacturer | Welch Allyn |
| Presentation | 1 unit |
| Recycling | WEEE |
Technical note: the original BATT99 reference from Welch Allyn for the Connex VSM 6000 platform is documented in regulatory databases (GUDID/FDA) and in specialised distribution as a lithium-ion (Li-ion) pack of 10.8 V 6.75 Ah (with variants up to 7.5 Ah / 81 Wh). This datasheet corresponds to the Ni-MH variant of 10.8 V 6.75 Ah; in addition, the internal Cahnos record documents reference 88888849 as a lithium-ion cell of 10.8 V 6.75 Ah, so before replacing a battery in service, the specific chemistry supplied and the monitor’s charging circuit compatibility must be confirmed with the technical department.
4. Equivalences and Cross-References
The documented OEM reference for the internal battery of the Welch Allyn VSM6000 / Connex VSM 6000 monitor is BATT99, also cited on the market as 11476/11477 (specialised distribution) and 6408-A (R&D Batteries). In the replacement market, the same application is covered by replacement battery manufacturers that claim to replace the BATT99. The references listed below are classified as comparable by category (not 1:1 interchangeable) and must be verified one by one —voltage, capacity, chemistry, connector and dimensions— against the specific unit before replacement. Baxter, current brand owner, recommends using exclusively the authorised BATT99 battery in accordance with its safety and performance tests.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Welch Allyn (Hillrom / Baxter) | Original battery Connex VSM 6000 / Connex Vital Signs Monitor 6000 | BATT99 (11476 / 11477 and 6408-A in distribution) | Exact reference equivalence (same manufacturer) | OEM reference for the VSM6000 battery; documented in GUDID and distribution as Li-ion 10.8 V 6.75 Ah (81 Wh variants); this datasheet describes the Ni-MH variant; verify charging circuit chemistry before installing. |
| OSI Batteries (Unipower) | Medical series Welch Allyn Connex | B11694 | Comparable by category (not interchangeable 1:1) | Li-ion 10.8 V 7,500 mAh (81 Wh); declared as a replacement for BATT99; verified in the pack manufacturer’s own catalogue and on BatteryMart; capacity and chemistry superior to the Ni-MH variant in this datasheet; check connector and charging circuit. |
| Cameron Sino | Medical series Welch Allyn Connex | CS-WB150MD | Comparable by category (not interchangeable 1:1) | Li-ion 11.1 V 7,800 mAh (≈86.58 Wh); declared for BATT69/BATT99 on Connex VSM6000, Connex Spot and CP150; verified on BatteryMart, BatteryWorld and BatteryBob (dimensions ≈203.6 x 71.2 x 23.7 mm); different voltage and dimensions; check fit and connector. |
| Artisan Power | Replacement for VSM6000 / Connex 6000 | MB-BATT99-L | Comparable by category (not interchangeable 1:1) | Li-ion; registered in GUDID/FDA (primary UDI 00819595016750) as a replacement for BATT99/BATT69 for VSM6000 and Connex Vital Signs Monitor 6000; chemistry different from the Ni-MH variant in this datasheet; check before fitting. |
| R&D Batteries | BATT99 replacement for Connex Spot / Connex / CP150 | 6408-A | Comparable by category (not interchangeable 1:1) | Li-ion 11.1 V 6.6 Ah (66 Wh); aftermarket replacement for BATT99 verified on the manufacturer’s website, which lists as cross references Anybattery 8535, Cables and Sensors 11476, NCE BATT99, CS-WB150MD, MED0266 and B11694; different voltage and capacity; check connector and compartment. |
| Interstate All Battery | Connex series medical battery | MED0266 | Comparable by category (not interchangeable 1:1) | Cited as a cross reference for BATT99 at R&D Batteries; stated in distribution datasheets as Li-ion 10.8 V 6.75 Ah; detailed specification and UDI not verified in public sources; check chemistry, connector and dimensions before replacing. |
Note on the manufacturer’s reference: the code 4500-84-1 shown in the specification table cannot be verified as such in public sources for the VSM6000 monitor; the documented references for the battery of the Connex VSM 6000 platform are BATT99 (OEM) and, in distribution, 11476/11477 and 6408-A. Furthermore, on the market the code 4500-84 / 4500-84-1 is associated with the 6 V battery of the Vital Sign Spot LXI monitor, with different voltage and chemistry: it must not be confused. Confirm the reference on the label of the removed unit before replacing.
There is no equivalence table validated by the manufacturer (Welch Allyn / Hillrom / Baxter) that authorises direct cross-referencing between this reference and the third-party codes cited; they are listed as documentary support and must be verified against the equipment in service. Consult the technical department before using this reference in a tender or replacing the battery of a monitor in use.
5. Selection tips and variants of interest
- Verify the exact monitor model: this battery corresponds to the Welch Allyn VSM6000 / Connex VSM 6000 platform (original reference BATT99). Also confirm the label of the installed battery. Do not install it in other families such as VSM300, Spot LXI, Connex Spot, Propaq or PIC50 without confirming compatibility: each device accepts a specific voltage, capacity and geometry.
- Confirm the charger chemistry: this datasheet describes the Ni-MH 10.8 V 6.75 Ah variant and the internal Cahnos record documents the reference as lithium-ion; if your monitor worked with the original cell documented in GUDID as Li-ion, check with the technical department the supplied chemistry and that the charging circuit supports it before installing.
- Spare part rotation: in fleets with several VSM6000 monitors, keep a spare battery charged and rotate its use to extend the service life of the set. Ni-MH batteries self-discharge at rest; if the device is to remain off for long periods, carry out periodic recharging to avoid deep discharges.
- Other Welch Allyn variants at Cahnos: 6.4 V 6 Ah battery for Spot LXI (MFR 105632, refs. 88881196 / 88889726), 6 V 6 Ah battery for Vital Sign Spot LXI (MFR 4500-84-1, ref. 88888799), 6 V 4.5 Ah battery for VSM300 (ref. 88888655), 7.2 V 3.2 Ah battery for Connex Spot (MFR BATT22, ref. 88889217), 6.4 V 6 Ah battery for Spot 420 (ref. 88889216) and 8.4 V 1.9 Ah battery for Propaq-LT (ref. 88888181). Do not interchange these references between models: each device accepts a specific voltage, capacity and geometry.
- Institutional purchases: for tenders, continuous supply contracts or large volumes, contact our technical-commercial department to confirm compatibility, chemistry, lead times and a formalised quotation.





Reviews