Description
1. Reliable power for your vital signs monitor
Keep your Nihon Kohden vital signs monitor always operational with this rechargeable nickel-metal hydride (Ni-MH) battery, factory reference X075 / SB-671P (Ref. Cahnos 88888786). It is the replacement that ensures the equipment’s autonomy in emergencies, intra-hospital transfers, operating theatres and any situation where mains power is not available, without compromising continuous patient monitoring.
This is a direct-fit replacement for the monitor’s battery compartment, restoring the equipment to full operation in a few minutes at a very competitive cost compared to the official replacement. Ideal for medical electronics departments, hospitals, clinics and medical supply contracts.
2. Technical explanation: how this battery works
The reference X075 is the nickel-metal hydride (Ni-MH) battery that Nihon Kohden installs as standard in the multiparameter monitors of the BSM-6000 and BSM-3000 series and in the WEP-5200 telemetry systems, also identified by the factory designation SB-671P. Internally it groups Ni-MH cells to deliver a nominal voltage of 9.6 V with 2.9 Ah capacity (27.84 Wh of useful energy), sufficient to maintain monitoring of ECG, SpO2, non-invasive blood pressure and respiratory rate during transfers and equipment movements.
While the monitor is connected to the mains, the battery remains in float charge mode; when disconnected or if the supply is interrupted, it automatically takes over powering the screen, the monitoring electronics and the measurement modules without interruption. Its Ni-MH chemistry requires no electrolyte maintenance, supports cyclic recharging with the equipment’s own integrated charger and should be replaced when loss of autonomy is detected or the low battery warning appears on screen.
3. Technical Specifications
The following specifications correspond to the battery compatible with reference X075 / SB-671P (Ref. Cahnos 88888786), verified against the available technical documentation of the original equipment.
| Parameter | Detail |
|---|---|
| Reference Cahnos (SKU) | 88888786 |
| Application | Nihon Kohden vital signs monitors BSM-6000, BSM-6300, BSM-6500, BSM-6700, BSM-3000, BSM-3562, MU-671RK, MU-651K and WEP-5200 telemetry (per manufacturer’s IFU) |
| Original reference | X075 / SB-671P |
| Technology | Ni-MH (nickel-metal hydride), rechargeable |
| Nominal voltage | 9.6 V |
| Capacity | 2.9 Ah (2900 mAh) |
| Energy | 27.84 Wh |
| Weight | Approximately 0.55 kg (may vary slightly by batch) |
| Dimensions | 76 x 96 x 35 mm |
| Device manufacturer | Nihon Kohden |
| Mounting type | Direct replacement, without adapters or special tools |
| Load | Integrated into the monitor (equipment’s own charger) |
| Presentation | 1 unit |
| Recycling | Do not dispose of as household waste; recycle in accordance with WEEE regulations |
4. Equivalences and Cross-References
Nihon Kohden does not publish an official cross-reference table authorising third-party batteries for reference X075. The alternatives listed below share chemistry (Ni-MH), voltage (9.6 V) and capacity (2.9 Ah), and have been verified against the official datasheets of their own manufacturers or in authorised distributor catalogues; they are classified as comparable by category, not 1:1 interchangeable, and require verification of capacity, dimensions and connector against the specific equipment before replacement. Reference SB-671P is registered in the FDA GUDID database (NIHON KOHDEN CORPORATION and NIHON KOHDEN AMERICA, INC., class II, product code MHX, battery for BSM-6000).
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Nihon Kohden (factory designations) | BSM-6000 / BSM-3000 / BSM-6300 / BSM-6500 / BSM-6700 / WEP-5200 | X075 = SB-671P | Exact reference equivalence (same manufacturer) | Factory codes for the same Ni-MH 9.6 V battery for the BSM-6000/BSM-3000 family, cited in the manufacturer’s official IFU and registered in GUDID/FDA; verify the exact variant printed on the label of the installed cell. |
| R&D Batteries | Medical battery for Nihon Kohden Lifescope TR / BSM-6001 / BSM-6301 / BSM-6501 / BSM-6701 | 6336 | Comparable by category (not interchangeable 1:1) | Ni-MH 9.6 V, 2.9 Ah, rechargeable, OEM reference SB-671P. Different manufacturer (not original Nihon Kohden); same nominal dimensions in inches (2.95 x 1.41 x 3.77 in ≈ 75 x 36 x 96 mm); confirmed in the manufacturer’s official datasheet. |
| Cables & Sensors | Medical battery for Nihon Kohden Life Scope TR / BSM-6001 / BSM-6301 / BSM-6501 / BSM-6701 | 11274 | Comparable by category (not interchangeable 1:1) | Ni-MH 9.6 V, 2.9 Ah, cross-referenced with SB-671P and R&D 6336. Different manufacturer/distributor (not original Nihon Kohden); dimensions 3.07 x 1.41 x 3.84 in and weight 1.45 lb (≈ 0.66 kg); confirmed in the authorised distributor’s datasheet. |
| KingSener (distributed by Batterymall) | Medical battery for Nihon Kohden BMS-3000 / BSM-6000 | SE303140 | Comparable by category (not interchangeable 1:1) | Ni-MH 9.6 V, 2900 mAh (27.84 Wh), weight 0.55 kg; replaces X075. Different manufacturer (not original Nihon Kohden); same electrical specifications and weight matching distribution catalogues; confirmed in the distributor’s datasheet. |
| OSI Batteries | Medical battery for Nihon Kohden BSM-6301 | B12154 | Comparable by category (not interchangeable 1:1) | Ni-MH 9.6 V, 2.9 Ah, OEM reference SB-671P. Different manufacturer/distributor (not original Nihon Kohden); datasheet confirmed in the distributor’s official catalogue. |
There is no cross-reference table validated by the manufacturer (Nihon Kohden) that authorises third-party batteries for reference X075; consult with the technical department before use in tenders.
5. Selection tips and variants of interest
Identify the removed cell: before replacing, verify three parameters on the battery label: voltage (9.6 V), technology (Ni-MH) and factory reference (X075 or SB-671P). If your equipment is a BSM-6000, BSM-3000, BSM-6300, BSM-6500, BSM-6700, BSM-3562, MU-671RK, MU-651K monitor or a WEP-5200 telemetry system, this reference is the correct one.
Do not confuse it with other batteries in the family: Nihon Kohden uses different X0xx codes depending on the equipment: X060 corresponds to Lifescope A (Ref. Cahnos 88888774), X062 to Lifescope-i BSM-2300 / BSM-2351 (Ref. 88888428), X064 to Lifescope N (Ref. 88888779) and X076 to PVM27xx monitors (Ref. 88888815). Only one variant fits each equipment; always check the installed cell before purchasing.
Preventive maintenance: Ni-MH cells self-discharge slowly when idle; if the equipment is to remain switched off for long periods, recharge it periodically and avoid repeated deep discharges to not shorten its service life. After replacement, fully charge the battery before first use and allow the monitor to complete its power-on self-test.
Tenders and equipment fleets: contact our technical department indicating the model, serial number and reference of the installed battery; we will confirm compatibility before placing the order and offer special volume conditions.





Reviews