Description
1. Product Presentation
The Cardioline ECG100L FULL electrocardiograph (ref. 54202) is the resting electrocardiography solution that your cardiology practice, diagnostic centre or primary care service needs to obtain accurate, fast and fully documentable 12-lead traces. Designed by Elettronica Trentina and distributed in its Cardioline line, this device stands out for its ease of use: a bright 5″ colour touchscreen guides the professional step by step from electrode placement to quality control, acquisition, printing and storage of the record. This FULL version incorporates the Glasgow interpretation algorithm and the ECG EasyApp application, making it a complete device for cardiological diagnosis in the small medical practice environment.
2. Technical explanation of the equipment
The ECG100L FULL simultaneously acquires the 12 standard electrocardiographic leads (I, II, III, aVR, aVL, aVF and V1-V6) via a standard 15-conductor patient cable (10 wires), with a sampling rate of 32,000 samples/second/channel, resolution of 5 µV/LSB and a common-mode rejection ratio (CMRR) greater than 100 dB that guarantees clean and stable traces even in environments with electrical interference. It offers automatic (12-lead), manual (3/6), Stat (12) and rhythm (1/3) acquisition modes, with standard or Cabrera configuration, and supports 3, 3+1, 3+3, 6 and 12-channel print formats.
The rhythm recording allows up to 3 minutes of ECG to be analysed, including heart rate trend and its variability. The Glasgow interpretation programme diagnoses the anomalies usually recognisable in an ECG: myocardial infarction (including AMI), ventricular hypertrophy, ST-T segment and rhythm anomalies, as well as conduction disturbances and QT interval prolongation. Files are stored on the device itself (up to 50 ECGs) and can be exported via USB to a PC or USB memory stick; the ECG EasyApp application, included in this version, receives recordings in SCP format, allows them to be printed, patient data and automatic interpretations to be modified, and the layout of rhythm strips to be customised (speed, gain, filters, etc.).
3. Technical Specifications
ECG acquisition
- ECG channels: 12 leads (I, II, III, aVR-L-F, V1-6)
- Patient cable: standard 15D, 10 wires
- CMRR: >100 dB
- Input impedance: 100 MOhm
- Sampling rate: 32,000 samples/second/channel
- ECG resolution: 5 µV/LSB; 500 s/s
- Dynamic range: ±325 mV
- Bandwidth: performance equivalent to 0.05-150 Hz
- Pacemaker detection: hardware detection with digital convolution filtering
- Filters: linear-phase digital high-pass (per 60601-2-25 2nd ed.), adaptive digital mains interference filter 50/60 Hz and digital low-pass filters at 25/40 Hz (display and print only)
- Defibrillation protection: AAMI/IEC standards
- Front-end performance: ANSI/AAMI IEC 60601-2-25:2011
- Acquisition modes: automatic (12 leads), manual (3/6), Stat (12), rhythm (1/3)
- Configuration: standard or Cabrera
- Disconnected electrode detection: independent on all leads
- Heart rate meter: 30-300 bpm
- ECG measurements: all leads, average, corrected; HR, mean RR, PR interval, QRS duration, QT and QTc intervals (Hodges, Bazett and Fridericia formulas), maximum R [V5] or [V6] and S [V1], Sokolow-Lyon index; P, R, T axes
- Interpretation: Glasgow analysis programme (adult, paediatric, STEMI)
- ECG storage: 50 ECG
- Export format: SCP-PDF
- PC software: ECG EasyApp, ECG management application for PC (included)
Display
- Type: 5″ backlit colour TFT LCD with capacitive touch panel
- Resolution: 800×480
- Data: 3/6/12 real-time leads
Printer
- Thermal printhead: 108 mm
- Resolution: 8 dots/mm
Physical characteristics
- Device dimensions/weight: 270×190×60 mm · 1.48 kg
Power supply
- Power supply: 100-240 V, 50/60 Hz
- Power supply protection rating: IP20
- Battery: NiMH
- Battery life: more than 500 ECGs, more than 6 hours
- Charging time: 4 hours to 85% capacity
Standards and safety
- MDD 93/42/EEC, class IIa
- IEC 60601-1-2 (electromagnetic compatibility): group 1, class B
- IEC 60601-1 (electrical safety)
- Protection against electric shock: IP (internally powered ME equipment), class I (with external AC/DC power supply)
- Protection against water: IXP 0
Product data
| Brand | CARDIOLINE |
|---|---|
| Model | ECG100L FULL |
| Manufacturer reference | 54202 |
| Sales unit | 1 pc |
| Device type | Medical device (medical device) |
| Device class | II A |
| CND | Z12050302 |
| GMDN | 16231 |
| EAN | 8056732654983 |
Included accessories
Patient cable, 2 rolls of paper, battery, power cable, adapter for 4 mm snap connector (box of 10), Glasgow interpretation, disposable adhesive electrodes (box of 100) and manual CD (GB, FR, IT, ES, PT, DE, NL, CZ, HR, HU, PL, RO, RU).
4. Equivalences and Cross-References
There is no official and validated equivalence table published by the manufacturer that cross-references this item with other models or brands on a 1:1 interchangeable basis. The following references are classified as comparable by category (resting 12-lead ECG) and must be compared field by field (channels, display, interpretation, connectivity and regulatory classification) before any substitution:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Observations |
|---|---|---|---|---|
| Cardioline (Elettronica Trentina) | ECG100L Basic – ECG100L with interpretation (5″) | 54200 / 54201 | Comparable by category (not interchangeable 1:1) | Same 12-lead platform and 5″ touchscreen; version 54201 adds Glasgow and 54200 does not; this reference 54202 is the FULL (Glasgow + EasyApp). |
| Cardioline (Elettronica Trentina) | ECG200L (7″ touchscreen) | 54204 / 54205 | Comparable by category (not interchangeable 1:1) | Versions with 7″ display and 12-channel printing (12 only in 54204-5); compatible with Glasgow and EasyApp. Verify dimensions and consumables before substituting. |
| Cardioline S.p.A. | ECG100S / ECG200S | ECG100S / ECG200S | Comparable by category (not interchangeable 1:1) | 12-lead ECG with USB connectivity to PC, SCP/PDF export and optional Glasgow interpretation; non-touchscreen in the basic series. Confirm specifications before use. |
5. Selection tips and variants of interest
This reference 54202 (ECG100L FULL) corresponds to the 5″ version with Glasgow interpretation and ECG EasyApp included, the most complete in the small-screen series. If your centre prefers a larger screen (7″) or needs to print 12 simultaneous channels, consider the ECG200L series (ref. 54204/54205); if you do not require automatic interpretation, the basic version 54200 can cover resting recording. The ECG EasyApp software, included in this version (and in 54205), is the ideal complement for digital archiving and customising reports on PC.
Before placing the order, confirm the printing consumable: this device uses thermal paper roll for a 108 mm print head; in the recording paper section of Cahnos you will find formats compatible with Cardioline electrocardiographs. For institutional orders, tenders or ongoing supply contracts, contact our technical-commercial department at Cahnos to confirm the exact version (54200/54201/54202/54204/54205) and obtain a formalised quote under B2B conditions.









Reviews