Description
1. Commercial Introduction
The Philips Laerdal HeartStart XL+ MFR 989803167281 battery (Ref. Cahnos 88881012) is the standard lithium-ion accumulator that powers the HeartStart XL+ defibrillator-monitor. It is a high-capacity factory spare part that ensures the device passes its self-tests, supports continuous monitoring and provides reserve power for shocks when there is no nearby mains socket.
Designed for emergency services, out-of-hospital emergencies, mobile intensive care units, medical transport and medical electronics services, this battery prevents periods of device downtime: it installs in seconds, charges in the defibrillator itself and communicates its status to the monitor via the BMS (Battery Management System) protocol.
2. Technical explanation: the role of the battery in the HeartStart XL+
The HeartStart XL+ is a professional defibrillator-monitor from Philips (a family also marketed under Laerdal) that operates on mains power or its internal battery when moving to the patient, during monitoring in an ambulance or during procedures without an electrical supply. The 989803167281 battery supplies the energy needed for defibrillation, ECG monitoring, external pacing and event recording.
It is a rechargeable lithium-ion accumulator with a nominal 14.8 V and 6.6 Ah capacity (approximately 97.7 Wh), with BMS electronics that verify the pack status at each start-up. When replacing the battery, the HeartStart XL+ performs a power-on self-test that includes verifying communication with the battery module; therefore, the start-up must be allowed to complete without interruption and a first full charge cycle should be performed before putting the device into clinical rotation.
3. Technical Specifications
| Characteristic | Value |
|---|---|
| Battery type | Rechargeable lithium-ion (Li-ion), with electronic BMS management |
| Nominal voltage | 14.8 V (14.4 V min. according to verified spare parts references) |
| Nominal capacity | 6.6 Ah (6600 mAh); third-party spare parts up to 6.75 Ah |
| Nominal energy | ≈97.7 Wh |
| Original manufacturer reference | 989803167281 (Philips HeartStart XL+ Battery) |
| Alternative OEM references cited in authorised distribution | M6479, M6479-O, SE02211, SE-02211 |
| Reference Cahnos | 88881012 |
| Compatible equipment | Defibrillator-monitor Philips HeartStart XL+ and its monitor/defibrillator variants. Not compatible with the original HeartStart XL, which uses the M3516A lead battery |
| Rechargeable | Yes |
| Installation | Direct fit in the rear compartment, guided connector, no tools required |
| Approximate dimensions | 150.2 x 116.8 x 23.5 mm (verified spare parts reference values) |
| Approximate weight | 0.67 kg |
| Maintenance | Charges in the device itself; allow the self-test to complete after replacement |
| Presentation | Unit |
| Recycling | Management in accordance with WEEE regulations |
Technical note: the electrical characteristics (Li-ion, 14.8 V, 6.6 Ah) match the manufacturer’s technical documentation and specialised biomedical distribution catalogues. Reference 989803167281 corresponds to the HeartStart XL+ battery; that of the original HeartStart XL (M3516A, sealed lead 12 V / 2.3 Ah) is a different product and is not interchangeable.
4. Equivalences and Cross-References
Note on nomenclature: the manufacturer’s official reference is 989803167281, identified in authorised distribution as Philips HeartStart XL+ Battery (rechargeable lithium-ion battery). In catalogues and compatibility tables from specialised distributors, the alternative part numbers M6479, M6479-O, SE02211 and SE-02211 are consistently cited for this same pack. The Cahnos reference for this product is 88881012. The third-party spare parts listed below belong to the same product category (biomedical Li-ion battery, 14.8 V / 6.6 Ah for the HeartStart XL+ family) and have been verified against replacement battery manufacturers’ data sheets; they are classified as comparable by category, not 1:1 interchangeable, and require verification of connector, seating and BMS self-test result before replacement.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| OSI Batteries (Unipower) | Biomedical battery for Philips HeartStart XL+ | B12104 | Comparable by category (not interchangeable 1:1) | Rechargeable Li-ion, 14.8 V / 6600 mAh; replaces 989803167281 according to the replacement manufacturer’s data sheet. Verify BMS handshake, connector and device self-test. |
| Cameron Sino (CS) | Spare part for Philips HeartStart XL+ (monitor/defibrillator) | CS-PHX281MD | Comparable by category (not interchangeable 1:1) | Li-ion 14.8 V / 6600 mAh (97.68 Wh); replaces 989803167281, M6479, M6479-O, SE02211 and SE-02211 according to the replacement manufacturer’s data sheet; dimensions 150.2 x 116.8 x 23.5 mm. Confirm first full charge cycle and BMS self-test. |
There is no equivalence table validated by the manufacturer (Philips) that cross-references this reference with third-party batteries; the codes B12104 and CS-PHX281MD come from replacement battery manufacturers and must be checked with the technical department before use in tenders or before replacing a unit in service. The reference 989803167281 was not found in the public search of the GUDID/FDA database at the time of writing this datasheet; its definitive verification must be made against the manufacturer’s batch label.
5. Selection tips and variants of interest
- Do not confuse XL with XL+: the original HeartStart XL uses the sealed lead battery M3516A (Ref. Cahnos 88888255); the 989803167281 corresponds to the HeartStart XL+ with lithium-ion. Always check the model printed on the equipment label before ordering.
- Replace the battery when the monitor warns of low battery (Low Battery), when it reports a battery failure in the self-test, or when it detects a failed communication with the BMS.
- After installing a new battery, allow the equipment to complete the start-up self-test without interruptions and perform a first full charge-discharge cycle so that the indicator recalibrates.
- If the equipment is to remain unused for a long time, store the battery with a partial charge (20–40 %) in a cool, dry place, and check its level periodically.
- Variants of interest for other Philips/Laerdal defibrillators available at Cahnos: 88888255 (HeartStart XL, M3516A), 88888363 (HeartStart MRx, M3538A), 88888328 (HeartStart FR2, M3863A), 88888783 (HeartStart FR3, 989803150161), 88888342 (HeartStart HS1/OnSite/FRx, M5070A), 88889059 (Efficia DFM100, 989803190371) and 88889102 (HeartStart FR1, BT1).
- If you have any doubts about compatibility with your batch of equipment or need terms for a tender, contact our technical department before placing the order.