We use our own and third-party cookies to analyse the use of the website and show you advertising related to your preferences, based on a profile built from your browsing habits (for example, pages visited). More information

Cahnos.com Compare invoice

Replacement Battery 12V 7.75Ah LiSO2 Non-Rechargeable for Cardiac Science Powerheart G5 Defibrillator · Replaces XBTAED001A (GUDID/FDA 00812394020133) · Ref. Cahnos 88881203

656,00 €

Non-rechargeable primary lithium LiSO2 battery, 12 V and 7.75 Ah, for the Cardiac Science Powerheart G5 automated external defibrillator (G5A-80A/C and G5S-80A/C). Replicates the original reference XBTAED001A, confirmed in GUDID/FDA (00812394020133), with Intellisense® technology: Rescue Ready® self-tests, up to 4 years standby once installed and 5 years service life from manufacture. Direct replacement, no tools or configuration required, to keep your AED always operational. Ref. Cahnos 88881203 (SKU 0159026).

International shipping. Import duties and taxes are payable by the recipient.

Datasheet (opens in a new tab)

Reference
0159026

Description

1. Backup power for your Powerheart G5 defibrillator

Keep your automated external defibrillator Cardiac Science Powerheart G5 always ready to act with this 12 V, 7.75 Ah lithium battery (Ref. Cahnos 88881203, SKU 0159026), a replacement that reproduces the original battery with reference XBTAED001A. This is the power source used by the device itself: reliable, proven and a direct replacement, designed so that the AED passes its automatic self-tests and responds with maximum effectiveness in every cardiac arrest. An ideal solution for emergency services, ambulances, health centres, companies with AEDs on their premises and medical electronics maintenance programmes.

2. How this battery works: technical explanation

The Powerheart G5 is an automated external defibrillator (AED) with STAR® biphasic waveform designed to operate fully autonomously: the internal battery powers the rhythm analysis electronics, capacitor charging and emergency discharge, also enabling monitoring with RescueCoach™ voice prompts. This battery uses lithium sulfur dioxide (LiSO2) chemistry and is a non-rechargeable primary cell: it must be replaced with a new unit when the device warns of low level or after its service life, never recharged. Thanks to its very low self-discharge, it keeps the device operational for long periods without maintenance. Its Intellisense® technology integrates intelligent circuitry that works with the device’s Rescue Ready® function: the AED performs daily, weekly and monthly self-tests (electrical circuit, software, electrodes and battery status), so any degradation is detected before it compromises a real intervention.

3. Technical Specifications

Data verified against the device manufacturer’s technical documentation, internal reference register and the GUDID/FDA regulatory database. The reference XBTAED001A is registered in AccessGUDID as «Intellisense® Battery Powerheart® G5» (primary battery, GMDN Primary battery, class III) under primary DI 00812394020133 (Cardiac Science Corporation) and with updated DI 00812394020904 (Zoll Medical Corporation, DI 00812394020133 as previous).

Parameter Specification
Reference Cahnos (SKU) 88881203 (SKU 0159026)
Original manufacturer reference XBTAED001A · GUDID/FDA 00812394020133 (current DI 00812394020904)
Type / technology Primary lithium-sulfur dioxide (LiSO2) battery, non-rechargeable · Intellisense® technology
Nominal voltage 12 V DC
Capacity 7.75 Ah (7,750 mAh; ≈ 90 Wh; depending on distribution also 7.5 Ah)
Lithium content ≈ 9.2 g (according to the device manufacturer’s official datasheet)
Installed standby life 4 years (full replacement warranty, subject to conditions)
Shelf life 5 years from date of manufacture
Rechargeable No; primary battery, replacement with a new unit
Compatibility Powerheart G5 (models G5A-80A, G5A-80C, G5S-80A, G5S-80C)
Self-tests Rescue Ready®: daily, weekly and monthly (circuit, software, electrodes and battery)
Installation Direct replacement in the original compartment, without tools or configuration
Approximate weight ≈ 0.5 kg (according to specialised distribution; verify batch)
Presentation 1 unit
Recycling Do not dispose of as municipal waste; manage in accordance with WEEE regulations (regulated lithium transport)

4. Equivalences and Cross-References

The original reference stamped on the battery of Powerheart G5 defibrillators is XBTAED001A, consistently confirmed in the internal reference register, the device manufacturer’s official catalogue and the GUDID/FDA regulatory database (DI 00812394020133, updated to 00812394020904). Cardiac Science (now part of Zoll Medical) does not publish an official equivalence table authorising third-party batteries for the Powerheart G5; the references listed below match in device, LiSO2 chemistry, voltage and capacity window, but are classified as comparable by category and require connector and batch verification before replacement.

Manufacturer / Brand Family / Model Code / Reference Equivalence level Technical differences / Remarks
Cardiac Science / Zoll Medical Original Intellisense® Powerheart G5 battery XBTAED001A Exact reference equivalence (same manufacturer) Original reference reproduced by this replacement: LiSO2 lithium, 12 V, 7.75 Ah (≈ 90 Wh), non-rechargeable, 4-year standby warranty. Confirmed in GUDID/FDA (00812394020133 / 00812394020904) and official catalogue.
AED Battery Exchange (aftermarket) Recell G5-ABE for Powerheart G5 G5-ABE Comparable by category (not interchangeable 1:1) LiSO2, 12 V, 7.5 Ah (90 Wh), dimensions ≈ 165 × 72 × 49 mm and ≈ 0.5 kg, for G5A-80A/C and G5S-80A/C; replaces XBTAED001A according to the replacement manufacturer’s datasheet and authorised distribution. Not an original Cardiac Science battery; verify connector and batch.
Physio-Control (Stryker) Primary battery LIFEPAK 500 / 1000 11141-000158 (also 11141-000013, 3005380-026) Comparable by category (not interchangeable 1:1) LiSO2 lithium, 12 V, 7.5 Ah, non-rechargeable, for LIFEPAK 500/1000 AEDs; same nominal electrical performance as this reference but with proprietary connector and management electronics of the Physio-Control device. Does not fit in the Powerheart G5 compartment.

There is no manufacturer-validated equivalence table for this battery; consult the technical department before using any cross-reference in tenders. As support, specialised distribution supplies this same configuration (LiSO2, 12 V, 7.5–7.75 Ah) under the cited codes G5-ABE and 11141-000158. Before accepting an alternative battery, each item must be verified one by one: voltage (12 V), capacity (7.75 Ah), chemistry (primary non-rechargeable LiSO2 lithium), compartment connector and compatibility with the device’s Intellisense®/Rescue Ready® electronics.

5. Selection tips and variants of interest

Identify the exact model of your device: this battery is the one indicated for the Cardiac Science Powerheart G5 in its G5A-80A, G5A-80C, G5S-80A and G5S-80C versions. Check three parameters on the label of the removed cell: voltage (12 V), technology (non-rechargeable lithium) and capacity (7.75 Ah). Do not install it in other models of the Cardiac Science family: the Powerheart 9145 uses battery 9145-301 (Ref. Cahnos 88888683) and the 9146/G3 Plus uses 9146-202/302 (Ref. Cahnos 88888768), both 12 V 7.5 Ah lithium but with different geometry and connector, so they are not interchangeable with the G5.

Do not recharge the battery: as it is a single-use primary LiSO2 lithium cell, recharging must not be attempted. Replace it when the device warns of low level, when Rescue Ready® self-tests indicate a depleted battery, or after the recommended service life (4 years installed / 5 years from manufacture), without waiting for it to fail during an intervention.

Storage and rotation: keep spare batteries in their original packaging, in a cool, dry place; rotate stock to respect manufacturing dates and avoid extreme temperatures that accelerate self-discharge. After replacement, allow the AED to complete its power-on self-test.

Accessories: also check the expiry date of the Powerheart G5 defibrillation electrodes (Ref. AB8823XELAED001A) to keep the device fully operational. For institutional orders, tenders or continuous supply contracts, contact our technical-commercial department stating model, serial number and reference of the installed battery; we will confirm compatibility before formalising the order and offer special volume conditions.

Additional information

Manufacturer reference

XBTAED001A (GUDID/FDA 00812394020133)

Battery type

Primary lithium-sulfur dioxide (LiSO2) battery, non-rechargeable, single-use

Technology

Primary lithium-sulfur dioxide (LiSO2) battery, non-rechargeable

Nominal voltage

12 V DC

Capacity

7.75 Ah (7,750 mAh; ≈ 90 Wh; depending on distribution also 7.5 Ah)

Lithium content

≈ 9.2 g

Standby life

4 years installed (full replacement warranty)

Shelf life

5 years from date of manufacture

Rechargeable

No (single-use primary cell)

Compatibility

Defibrillator Cardiac Science Powerheart G5 (G5A-80A, G5A-80C, G5S-80A, G5S-80C)

Self-tests

Rescue Ready®: daily, weekly and monthly (circuit, software, electrodes and battery)

Installation

Direct replacement in the original compartment, without tools or configuration

Approximate weight

≈ 0.5 kg (verify batch)

Recycling

Management in accordance with WEEE regulations (regulated lithium transport)

Original manufacturer reference

XBTAED001A (Cardiac Science / Zoll Medical)

GUDID registration

00812394020133 (current DI 00812394020904)

Installed standby life

4 years (full replacement warranty, subject to conditions)