Description
1. Commercial Introduction
The Rapid COVID-19 Antigen Test Operon SIMPLE/STICK Ag SARS-CoV-2 is a single-step in vitro diagnostic device (TROD) designed for the qualitative detection of the SARS-CoV-2 nucleocapsid antigen in nasal swab samples. Manufactured in Spain by OPERON S.A., it provides a fast and reliable screening response at the point of care: clinical laboratories, hospitals, emergency departments, primary care, private practices and healthcare residences. Its simple/stick format, ready to use and requiring no additional equipment, makes it the ideal tool for screening patients with suspected COVID-19 during the first days of symptoms, with results in approximately 10 minutes and with the guarantee of the CE-IVD marking.
2. Technical explanation of the procedure
The device uses a lateral flow chromatographic immunoassay (colloidal method with gold nanoparticles) for the qualitative detection of the nucleocapsid protein (N protein) antigen of SARS-CoV-2. By targeting the nucleocapsid and not the mutation points of the spike protein (S), the test maintains its diagnostic capability against circulating variants of concern. The sample is collected with the included nasal swab, introduced into the extraction buffer and the mixture migrates by capillary action along the test strip: the appearance of a coloured line in the test zone (T) indicates the presence of the virus, while the control line (C) validates the correct functioning of the assay.
Summary procedure:
- Allow the kit to reach room temperature and wash your hands.
- Collect the sample with the included nasal swab, rotating it gently in the nostril.
- Insert the swab into the container with the extraction buffer and mix properly.
- Add the extracted sample to the indicated well of the device.
- Wait the indicated reading time (approximately 10 minutes) and read the result; the control line must always appear to validate the test.
Indication for use (manufacturer’s data sheet): «For the qualitative in vitro detection of the SARS-CoV-2 nucleocapsid antigen in nasal swab (NS) samples from people with suspected COVID-19 within the first 5 days after symptom onset. This test is used by clinical laboratories or healthcare workers for testing in healthcare centres, not for home testing.»
3. Technical Specifications
| Parameter | Specification |
| Product type | Rapid antigen test (TROD) for in vitro diagnostics |
| Analyte detected | SARS-CoV-2 (COVID-19) – nucleocapsid protein (N) antigen |
| Detection method | Lateral flow immunochromatography (colloidal gold), qualitative detection |
| Sample type | Anterior nasal swab (NS); also suitable for oropharyngeal sample according to the manufacturer’s data sheet |
| Reading time | Approximately 10 minutes |
| Interpretation | Visual or automatic with OPERON LF Reader |
| Format | Single-use device, simple/stick format |
| Intended use | Professional: clinical laboratories and healthcare centres (not suitable for self-diagnosis) |
| Detection window | First 5 days from symptom onset |
| Reference method | RT-PCR |
| Presentation | Box of 20 units, nasal swabs included |
| Storage | 2–30 °C, in sealed packaging, protected from moisture |
| Shelf life | 24 months |
| Certification | CE-IVD marking (Directive 98/79/EC and RD 1662/2000) · Manufactured under ISO 13485 |
| Manufacturer | OPERON S.A. (Spain) |
Diagnostic performance (concordance with RT-PCR)
| Parameter | Value |
| Sensitivity (concordance) | 96.4 % |
| Specificity (concordance) | 100 % |
| Total concordance | 99.4 % |
| Limit of detection (LOD) | 6 × 10² TCID50/mL (1.63 × 10⁵ copies/mL) |
4. Equivalences and Cross-References
The commercial reference of Cahnos for this product is 181001. The official OPERON order code for the 20-unit presentation could not be verified in public sources and must be confirmed directly with the manufacturer before use in tenders. There is no cross-reference equivalence table validated by the manufacturer OPERON S.A. relating this reference to codes from other brands; the references indicated below are functional category comparisons verified against technical data sheets and authorised catalogues, and are not interchangeable 1:1. Before any substitution, the device format (stick versus cassette), the authorised sample type, the reading time, the intended use (professional versus self-testing) and the performance data must be checked.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| ACON Biotech – Flowflex™ | SARS-CoV-2 Rapid Antigen Test (professional TROD, cassette) | L031-11815 | Comparable by category (not interchangeable 1:1) | Same analyte (nucleocapsid protein) and same lateral flow technique, but cassette format (not stick), nasal and nasopharyngeal sample, reading in 15 minutes and pack of 25 kits. |
| ACON Biotech – Flowflex™ | SARS-CoV-2 Rapid Antigen Test (cassette, nasal sample) | L031-11845 | Comparable by category (not interchangeable 1:1) | Presentation of 1 test per kit and 20 kits per box; cassette format and nasal sample only; same antigen technique. |
| Multipathogen respiratory panel | Combined Rapid Test SARS-CoV-2 / Influenza A / Influenza B / RSV / Adenovirus | IRT-545 | Comparable by category (not interchangeable 1:1) | Expanded panel that also detects influenza A/B, RSV and adenovirus in the same sample; broader spectrum than the monovalent test, cassette format and reading in ~15 minutes. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
5. Selection tips and variants of interest
- Professional COVID-19 screening: this monovalent test in stick format is the simplest and most direct option when only SARS-CoV-2 coverage is needed at the point of care, with a result in approximately 10 minutes and no additional equipment.
- Optional automatic reading: if your laboratory handles a high volume of determinations, the result is compatible with automatic reading using the OPERON LF Reader, which reduces variability in visual interpretation.
- Variant detection: by detecting the nucleocapsid protein, the test maintains its diagnostic capability against circulating variants; in the event of a positive result, always confirm by RT-PCR before making therapeutic or isolation decisions.
- Seasonal differential diagnosis: if your centre also needs to distinguish influenza A/B, RSV or adenovirus in a single sample, consider the combined panels in the catalogue (e.g. COIF-522 / COIF-522C, IU-COVTESG-8 or IRT-545 / IRT-545-1).
- Volume and budget: the presentation of 20 units per box optimises the cost per determination in screening campaigns; contact our sales team for special conditions on volume orders.
- Storage: keep the kit in its sealed packaging at room temperature (2–30 °C), protected from moisture, and always respect the expiry date indicated.





Reviews