Description
1. Commercial Introduction
The rapid antigen test for Mycoplasma pneumoniae (reference IMP-502) is an in vitro diagnostic device for professional use that allows detection in just 15 minutes of the presence of the M. pneumoniae antigen in an oropharyngeal swab sample. It is the ideal screening tool for community pharmacies, general practice and paediatric consultations, emergency departments and primary care, especially during the current season of increased respiratory infections due to mycoplasma in paediatric and adult populations.
With simple handling, no additional equipment required and high accuracy (98.2% vs PCR), this test allows rapid guidance of the diagnosis of atypical pneumonia and other respiratory tract infections caused by Mycoplasma pneumoniae, facilitating therapeutic decision-making and avoiding unnecessary empirical treatments. It is presented in a professional box of 20 determinations with all the materials needed to perform it.
2. Technical explanation of the procedure
The device uses a lateral flow chromatographic immunoassay for the qualitative detection of Mycoplasma pneumoniae antigens in pharyngeal swab samples. Specific antibodies against the M. pneumoniae antigen are immobilised on the reactive membrane of the cassette: the extracted sample reacts with the conjugated antibodies and the mixture migrates by capillary action to the test line (T). If the sample contains the antigen, a coloured line appears in the T region; the control line (C) always validates the correct performance of the assay (adequate sample volume and correct membrane absorption).
Summary procedure: (1) Equilibrate the kit to room temperature (15–30 °C). (2) Collect the sample with the sterile swab by rubbing the back of the pharynx, avoiding contamination with saliva. (3) Insert the swab into the tube with extraction buffer and rotate it for about 10 seconds to release the antigen. (4) Add three drops of the extracted sample (approx. 120 µl) to the well (S) of the cassette. (5) Read the result at 15 minutes; do not interpret after 20 minutes.
3. Technical Specifications
The general characteristics, kit contents, packaging and clinical data of the product are detailed below.
General characteristics
| Parameter | Specification |
|---|---|
| Detection method | Lateral flow immunochromatography (qualitative chromatographic immunoassay) |
| Pathogen detected | Mycoplasma pneumoniae (M. pneumoniae) |
| Sample type | Oropharyngeal swab (sterile swab) |
| Format | Individual test cassette |
| Reading time | 15 minutes (do not interpret after 20 minutes) |
| Reference method | PCR (polymerase chain reaction) |
| Storage | 2 °C to 30 °C (do not freeze; keep in the sealed pouch until use) |
| Expiry | 24 months from date of manufacture |
| Use | In vitro diagnostic for professional use |
| Certification | CE marking · Manufactured under ISO 13485 |
Clinical data
| Parameter | Value vs PCR |
|---|---|
| Relative sensitivity | 94.3% (95% CI: 80.8%–99.3%) |
| Relative specificity | 98.7% (95% CI: 99.6%–100.0%) |
| Accuracy | 98.2% (95% CI: 95.7%–99.4%) |
| Correlation with PCR | Greater than 98% |
Box contents
| Materials included | Quantity |
|---|---|
| Test cassettes | 20 |
| Sterile swabs | 20 |
| Extraction tubes with buffer | 20 |
| Dropper tips | 20 |
| Instructions for use | 1 |
CE declaration of conformity
| CE declaration of conformity | |
|---|---|
| Available on request |
Packaging
| Packaging | |
|---|---|
| Units | 20 tests per box |
| Quantity per carton | 60 boxes per carton // 1,200 tests per carton |
4. Equivalences and Cross-References
Reference IMP-502 corresponds to the Mycoplasma pneumoniae Antigen Rapid Test Cassette (Throat Swab) from the Hangzhou AllTest Biotech family (a range also marketed by ACRO Biotech / Citest Diagnostics), an international standard in lateral flow immunochromatography for the detection of mycoplasma antigens in pharyngeal exudate. This same device is marketed in Europe under equivalent presentations from other distributors with identical technical specifications and a 15-minute reading time, such as the Mycoplasma pneumoniae rapid tests from Biopanda Reagents or the TruQuick format from Meridian Bioscience, as well as by the main European IVD manufacturers (SPINREACT and others), which facilitates supply homologation in purchasing centres.
Within the Cahnos rapid test catalogue, it shares technology, cassette format and presentation with other devices from the same family, allowing a comprehensive point-of-care diagnostic panel to be completed:
- IBP-502 – Rapid antigen test for Bordetella pertussis (whooping cough), same AllTest family, box of 20 tests with swab.
- IST-N502 – Rapid test for group A streptococcal sore throat (Strep A), same professional cassette format with oropharyngeal swab.
- FIG-502 – Rapid test for fetal membrane rupture (iGFBP-1), same manufacturing line.
- IRT-545 / IRT-545-1 – Combined panels SARS-CoV-2 / Influenza A and B / RSV / Adenovirus.
- ICIC-525 – Combined Covid-19 and Influenza A+B rapid antigen test.
- COIF-522 / COIF-522C – Multiplex antigen test Covid | Influenza A and B for professional use.
5. Selection tips and variants of interest
- Professional use: indicated for pharmacies, medical practices, paediatrics, emergency departments and clinical laboratories; it is not a home-use self-test. Like all TROD devices, it should be used as a diagnostic aid together with clinical assessment.
- Result confirmation: in the presence of positive results, the therapeutic regimen can be initiated quickly; negative or doubtful results should be confirmed by PCR, especially in paediatric, elderly or immunocompromised patients, as the antigen load may be below the detection threshold.
- Sample collection: perform the oropharyngeal swab on the posterior pharynx, avoiding the tongue and contamination with saliva, to maximise the antigen concentration.
- Volume and cost per determination: the presentation of 20 tests per box (1,200 tests per carton) optimises the cost per test and storage in centres with medium and high demand. Request special conditions for full-carton orders through our commercial team.
- Storage: keep the cassettes in their sealed packaging between 2 °C and 30 °C, protected from humidity and direct sunlight; do not freeze and respect the indicated expiry date.
- Respiratory differential diagnosis: if your centre needs to cover several aetiologies in a single sample collection, consider the combined panels in the catalogue (IRT-545, ICIC-525, COIF-522); for specific bacterial coverage, combine this test with IBP-502 (whooping cough) and IST-N502 (group A streptococcus).





Reviews