Description
1. Commercial Introduction
The MISSION® Hb haemoglobin and haematocrit analysis system is the ideal point-of-care diagnostic solution for medical practices, health centres, clinical laboratories and emergency services that need a reliable Hb and Hct result immediately. Its compact and portable design allows testing at the patient’s side with just 10 µL of blood and a reading in less than 15 seconds, without requiring sample preparation. A professional tool that streamlines anaemia screening, preoperative assessment and patient follow-up, maintaining a level of accuracy comparable to that of automatic haematology analysers.
2. Technical explanation of the equipment
The MISSION® Hb meter uses the principle of methemoglobin detection by reflectance photometry at a wavelength of 525 nm. When the test strip is inserted and the sample is introduced, the device measures the reflectance generated by the reaction and converts the result into quantitative haemoglobin (Hb) and haematocrit (Hct) values in less than 15 seconds. The system incorporates automatic calibration via code chip, so it does not require specimen preparation or manual calibration. It works with capillary or venous whole blood (only 10 µL), stores up to 1,000 results with date, time and patient identification, and offers auto power-off after 8 minutes to optimise the consumption of the 3 AAA batteries (up to 2,700 tests or 360 hours of use). The use of dry test strips eliminates the drawbacks of transporting and storing liquid reagents.
3. Technical Specifications
The complete technical specifications of the MISSION® Hb system are detailed below:
3.1. Measurement characteristics
| Methodology | Reflectance photometry |
|---|---|
| Detection principle | Methemoglobin |
| Time to result | < 15 seconds |
| Sample type and volume | 10 µL / capillary and venous whole blood |
| Hb measurement range | 4.5 – 25.6 g/dL |
| Hct range | 13 – 75% |
| Wavelength | 525 nm |
| Calibration | Automatic |
| Precision CV (intra-series) Hb | = 3% |
| Precision CV total Hb | = 3% |
| Accuracy | Hb 4.5 – 10 g/dL, ±0.4 g/dL; Hb 10 – 25.6 g/dL ± 4% |
| Memory | 1,000 tests with date/time and ID number |
3.2. Power supply and dimensions
| Power supply | 3 AAA batteries or AC adapter on request |
|---|---|
| Battery life | 2,700 tests or 360 hours |
| Auto power-off | 8 minutes |
| Size (LxWxH) | 127 x 58 x 25 mm |
3.3. Included accessories
- 10 test strips
- 1 code chip
- 3 AAA batteries
- 10 lancets (26 G)
- 2 control strips
- 1 carrying case
- 1 lancet holder
- 10 capillary transfer tubes (plastic)
- User manual in English and Italian
Multilingual box presentation: GB, FR, IT, ES, PT, DE, PL, RO, SE, FI, GR, TR, Arabic. Mini USB output for PC connection.
3.4. Product data
| Brand | MISSION |
|---|---|
| Sales unit | 1 piece |
| Device type | IVD – Professional use |
| CND | W02020699 |
| EAN | 0682607610327 |
| Manufacturer reference | 24022 |
| SKU Cahnos | GM24022 |
| Available references | 24022 |
4. Equivalences and Cross-References
The reference 24022 (SKU Cahnos GM24022) corresponds to the manufacturer’s official factory code within the MISSION® range of portable analysers, dedicated to quantitative in vitro diagnostic (IVD) tests. Within the same MISSION® ecosystem there are other analysers in the brand family, such as the cholesterol meter with 3-in-1 lipid panel (reference 23932) and the MISSION® thermal printer (reference 23929), compatible with the point-of-care diagnostic workflow. There is no manufacturer-validated cross-reference table that would allow this system to be exchanged 1:1 with haemoglobin meters from other manufacturers; any comparison with third-party reflectance photometry equipment must be validated field by field (sample, volume, measurement range, precision and units) before use in tenders or stock replacement.
5. Selection tips and variants of interest
This system is marketed as a 1-piece sales unit (complete device with starter accessories). For continued use, it is recommended to plan for the consumption of test strips, control strips, lancets and capillary tubes, as well as the specific calibration code chip. It is a professional-use IVD device: intended for qualified healthcare personnel in laboratories, practices and care services that require rapid monitoring of Hb and Hct. Always verify that the test strips correspond to the MISSION® Hb range of the device, use the control strips to validate the batch and keep the meter with its 3 AAA batteries or AC adapter. For institutional volumes, multi-unit allocations or specific B2B supply conditions, contact the technical-commercial department of Cahnos.