Description
1. Introduction
The Contiplex® D Cannula from B. Braun is the puncture component of the most established continuous regional anaesthesia system on the market. Designed for anaesthesiology services and pain units, it allows placement of a stable local anaesthetic infusion line next to the target plexus or nerve, ensuring prolonged postoperative analgesia with lower doses and reduced opioid consumption. An essential product in enhanced recovery programmes (ERAS) and in continuous block protocols for the brachial plexus, femoral, sciatic and other peripheral nerves.
2. Technical explanation
The Contiplex® D cannula consists of a fully insulated atraumatic puncture needle, with a 15º or 30º bevel and depth markings, covered by a Teflon (PTFE) capillary with integrated contrast lines detectable by X-ray. The assembly is introduced under electrical neurostimulation and/or ultrasound guidance: the cable integrated in the hub is connected to the neurostimulator to confirm proximity to the nerve via muscle contraction; once the optimal point is located, the capillary is advanced, the needle is removed and the Contiplex® catheter is introduced through the capillary to the target structure, with the local anaesthetic being administered continuously or intermittently via the infusion line.
The ergonomic, green and translucent hub integrates two extensions: the connection cable to the neurostimulator and the local anaesthetic administration line, with a standardised female Luer connector compatible with male Luer cone syringes for priming, aspiration and infusion. It is a sterile, single-use and non-resterilisable product, as required by the manufacturer’s IFU.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Product type | Cannula with puncture needle for continuous regional anaesthesia |
| System | Contiplex® D (B. Braun) |
| Cannula size | 1.3 x 55 mm (external Ø 1.3 mm, length 55 mm) |
| Bevel | 15º (ref. 4898205) or 30º (ref. 4898235) |
| Needle | Atraumatic, fully insulated, with depth markings |
| Capillary | Teflon (PTFE) with radiopaque contrast lines, detectable by X-ray |
| Hub | Ergonomic, green and translucent, with neurostimulation cable and local anaesthetic administration line |
| Connections | Standardised female Luer connector; standardised Contiplex® electrode connector |
| Sterilisation | Sterile (ethylene oxide), single-use, non-resterilisable |
| MRI compatibility | Not evaluated for magnetic resonance environment (see IFU) |
| Presentation | Box of 25 units |
4. Equivalences and Cross-References
Within the B. Braun family itself, the official references corresponding to this cannula are 4898205 (15º bevel) and 4898235 (30º bevel), verified against the manufacturer’s technical data sheet. The same cannula is also marketed integrated into the Contiplex® D continuous block sets (refs. 4898305, 4898335, 4898325, 4898308 and 4898311).
There is no equivalence table validated by the manufacturer that cross-references this cannula with products from other brands; consult the technical department before use in tenders. As a category-based reference, other leading European brands in stimulated continuous peripheral block are PAJUNK® (StimuLong® NanoLine and PlexoLong® families) and Vygon, whose specific codes must be checked against their official catalogue according to gauge, length, bevel and components.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| Teleflex (Arrow) | StimuCath™ Continuous Peripheral Nerve Block Set | AB-19608-S | Comparable by category (not interchangeable 1:1) | Complete continuous block system with stimulating catheter (19G) and introducer; it is not a Teflon cannula with integrated needle like Contiplex® D. Verify configuration and references of each presentation before substituting. |
5. Selection tips and variants of interest
- 15º vs. 30º bevel: the 15º bevel offers less trauma in superficial approaches; the 30º one facilitates passage through cutaneous planes and denser tissues. Choose according to technique and anatomical area.
- Cannula length: the 1.3 x 55 mm size is the most common for brachial plexus blocks (interscalene, supraclavicular and axillary approaches), femoral and other superficial approaches; for deep approaches consider the 80 and 110 mm versions of the Contiplex® D system.
- System compatibility: use the cannula with the Contiplex® catheter (400 mm for 55 mm needles; 1000 mm for 80/110 mm needles) and with the Perifix® 0.2 µm connector and filter, as recommended by the manufacturer’s IFU to ensure patient safety.
- Ultrasound guidance: if your usual practice is dual guidance (ultrasound + neurostimulation), consider complementing with echogenic needles from the Contiplex® S Ultra 360 range, keeping the same catheter system and connections.
- Single-use: do not resterilise or reuse; check the integrity of the sterile packaging and the expiry date before each procedure.






Reviews