Description
1. Product Presentation
The absorbable multifilament poliglactin 910 (PGLA) suture without needle from the PREMISUT range is the trusted solution for operating theatres, clinics and B2B purchasing centres that need sterile surgical thread in loose format and of great length. Its presentation as a 180 cm strand per unit allows optimal use of the material in ligatures, approximation of deep planes, high-tension closures and manual suturing techniques with a reusable needle holder. The USP 1 gauge, with violet finish, offers excellent visibility on the surgical field and mechanical performance comparable to the major braided absorbable ranges on the market, with an excellent quality-price ratio for volume procurement.
2. Technical Explanation
Poliglactin 910 (PGLA) is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide. Its braided, non-crimped multifilament structure provides great flexibility, smooth passage through tissues and firm, secure knot anchorage. Absorption occurs by hydrolysis, a uniform process independent of tissue enzymes that guarantees a predictable loss of mass between 56 and 72 days, with minimal tissue reaction and maintenance of tensile strength during the first weeks of healing. As it is a needle-free presentation, the thread is ready for mounting on a needle holder and manual suturing techniques, being especially practical for ligatures, internal closures and precision procedures. The USP 1 gauge (metric 4) combines a medium-thick diameter with notable tension retention in tissues subjected to moderate-high load. It is supplied sterile by ethylene oxide, for single use and without re-sterilisation.
3. Technical Specifications
3.1. General characteristics
| Characteristic | Specification |
| Material / composition | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Suture type | Synthetic absorbable multifilament |
| Structure | Braided multifilament, non-ribbed |
| Format | Without needle (loose thread for manual suturing and ligatures) |
| USP size | 1 (metric 4) |
| Approximate diameter | 0.40 – 0.499 mm |
| Thread length | Thread of 180 cm per unit |
| Colour | Violet (colourless on request) |
| Brand / quality | PREMISUT, standard quality |
| Sterilisation | Ethylene oxide (EtO) |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference / SKU | 1L1-18001 |
3.2. Absorption profile and tensile strength retention
| Parameter | Value |
| Absorption type | Hydrolysis |
| Absorption time | 56 – 72 days |
| Tensile strength retention | Approximately 32 – 35 days |
| Tensile strength at 2 weeks | 75 % |
| Tensile strength at 3 weeks | 50 % |
| Tensile strength at 4 weeks | 30 % |
| Knot tying | Excellent safety |
3.3. Size equivalence
| USP size | Approximate diameter | Metric |
| 1 | 0.40 – 0.499 mm | 4 |
Main indications: tissue approximation, medium-thick gauge ligatures, deep planes, moderate-tension closures and ophthalmic surgery. Purchase format: price per box of 12 units, minimum order of 30 boxes.
4. Equivalences and Cross-References
Poliglactin 910 is the most widely used absorbable multifilament material in surgical practice. This needle-free reference behaves as a functional equivalent of the braided poliglactin 910 threads from the main international manufacturers, with absorption profiles (56-72 days) and tensile strength retention equivalent to ranges such as Vicryl / Vicryl Plus (Ethicon / Johnson & Johnson), Safil (B. Braun) and Polysorb (Medtronic). As for loose needle-free strands, it is comparable to the poliglactin 910 range from Laboratorio Aragó.
| Manufacturer | Equivalent range |
| Ethicon / Johnson & Johnson | Vicryl / Vicryl Plus (polyglactin 910) |
| B. Braun | Safil (polyglactin 910) |
| Medtronic | Polysorb (polyglactin 910) |
| Laboratory Aragó | Polyglactin 910 in needle-free strands |
Within the Cahnos catalogue, the warehouse version of this same presentation is handled through our technical-commercial team, and this reference belongs to the PGLA – Poliglactin 910 family of the PREMISUT brand, which includes both needle-free formats and sutures mounted with needles (HR cylindrical, spatulated and reverse-cutting triangular, among others).
5. Selection tips and variants of interest
The needle-free format is the recommended option for ligatures, approximation sutures with a reusable needle holder and procedures where the thread needs to be cut to the technique’s requirements. The USP 1 gauge (metric 4) is indicated for tissues requiring notable initial resistance: deep planes, fascia, aponeurosis, moderate-tension closures and vascular ligatures, while maintaining excellent manageability for general surgery and precision approaches. The 180 cm strand maximises yield per box and reduces waste in the operating theatre. The violet colour facilitates visibility and control of the thread on the surgical field; the colourless version is available on request. For earlier absorption, the rapid-absorption poliglactin 910 range (Rapid) may be considered and, for longer tensile support, the polydioxanone or caprolactone lines. Consult our technical-commercial team at Cahnos to optimise your box order and benefit from special B2B conditions.