Description
1. Product Presentation
The absorbable suture PREMISUT of Polyglactin 910 (PGLA) Prime, reference P1L11VSP840-2, is a high-performance synthetic multifilament suture designed for professionals seeking precision, comfortable handling and predictable clinical outcomes. Its double ¼ circle spatulated needle of 8 mm facilitates atraumatic passage through delicate tissues, while the violet thread of 45 cm in USP 4/0 gauge offers excellent visibility in the surgical field. Each box contains 12 units, making it an efficient solution for operating theatres, ophthalmic clinics and general surgery departments.
2. Technical explanation
This suture is manufactured from a polyglactin 910 copolymer (90 % glycolide and 10 % L-lactide), structured as a coated braided multifilament that combines flexibility and surface smoothness, minimising tissue drag. It is absorbed by hydrolysis over a range of 56 to 72 days, maintaining tensile strength during the critical healing period: 75 % at 2 weeks, 50 % at 3 weeks and 30 % at 4 weeks, with tensile strength retention for 32-35 days. The Micropoint spatulated needle, non-grooved, reduces trauma to fragile tissues (cornea, conjunctiva and eyelids), and its double-needle format allows surgical time to be optimised in continuous sutures or multiple stitches. The product is sterilised by ethylene oxide and is indicated for ophthalmic surgery, soft tissue approximation, subcutaneous closures and ligatures in general, digestive, gynaecology and urology surgery.
3. Technical Specifications
| Composition | Copolymer of 90% glycolide and 10% L-lactide (Polyglactin 910 / PGLA) |
| Structure | Coated braided multifilament |
| Gauge | USP 4/0 |
| Needle | Double needle ¼ circle spatula Micropoint (2 x VSP), non-cutting |
| Needle size | 8 mm |
| Thread length | 45 cm |
| Colour | Violet (non-barbed) |
| Absorption | By hydrolysis, between 56 and 72 days |
| Tensile strength retention | 32-35 days |
| Sterilisation | Ethylene oxide |
| Presentation | Box of 12 units |
| Reference | P1L11VSP840-2 |
Tensile strength retention after implantation:
| Period | Percentage of retained strength |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
Absorption profile:
| Mechanism | Value |
| Absorption mechanism | Hydrolysis |
| Practically complete absorption | 56-72 days |
| Tissue reaction | Minimal |
| Use | Single-use · Do not re-sterilize |
4. Equivalences and Cross-References
Polyglactin 910 (PGLA) is the direct equivalent of the Vicryl line (Ethicon) and of the Safil (B.Braun), Surgicryl 910 (SMI) and polyglactin 910 ranges from other manufacturers, with an absorption profile of 56-70 days and similar strength retention (75 % / 50 % / 25 % at 14, 21 and 28 days). In the internal reference catalogue (Laboratorio Aragó), this configuration corresponds to the Polyglactin 910 range in a presentation with a spatulated needle. As an alternative in polyglycolic acid (PGA), the same needle format and gauge is found in reference AMP111VSP840-2. Compared with polyglecaprone (monofilament, absorption 90-110 days) or PGA (absorption 60-90 days), PGLA offers a more predictable medium absorption for soft tissues and precision microsurgery.
5. Selection tips and variants of interest
The USP 4/0 gauge with an 8 mm Micropoint spatulated needle is especially suitable for ophthalmic surgery (cornea, conjunctiva, eyelids), microsurgery and approximation of delicate tissues where minimal trauma is required. For tissues with higher strength requirements, larger gauges can be considered, and for finer microsurgery, gauges 5/0 and 6/0. In the same 1/4 circle spatulated format, the references P1L11VSP850-2 (USP 5/0), P1L11VSP860-2 (USP 6/0) and P1L11VSP6,560-2 (6.5 mm needle, USP 6/0) are available; in 1/2 circle spatulated, P1L11HSP860-2 (USP 6/0 violet) and P1L21HSP860-2 (USP 6/0 colourless). It is recommended to check with the technical-commercial service the minimum order conditions (10 boxes) and the availability of colourless presentations or with other thread lengths.