Description
1. Commercial Introduction
The absorbable multifilament suture of polyglactin 910 (PGLA) from the Premisut range, reference 1L18HR4802, is the reference surgical closure solution for the professional who demands a flexible, resistant thread with a totally predictable absorption behaviour in the operating theatre. Its ½ circle needle with cylindrical body and conical point of 48 mm (HR 48) penetrates soft tissues with precision and without tearing, making it especially suitable for tissue approximation, ligatures and ophthalmic surgery.
It is a synthetic absorbable medical device, sterile and single-use, with an excellent price-quality ratio for B2B procurement for hospitals, ophthalmic clinics and purchasing centres. It is presented in a box of 12 units with a minimum order of 30 boxes, ensuring continuous supply for your healthcare centre.
2. Technical explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90 % glycolide and 10 % L-lactide, presented in a braided multifilament structure with a grooved finish. This configuration provides a regular surface that facilitates the passage of the thread through tissue, very smooth handling and firm, secure knot tying, even in larger diameter gauges such as USP 2 (metric 5).
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. This profile makes the suture ideal for tissues that require wound support during the first weeks of healing, maintaining the necessary tensile strength while the tissue recovers its own resistance. The ½ circle needle with conical point penetrates without cutting, minimising trauma to soft tissues. The product is supplied sterile by ethylene oxide, single-use and with no possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference 1L18HR4802
| Characteristic | Specification |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Gauge | USP 2 (metric 5) |
| Thread colour | Violet |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Taperpoint |
| Needle size | HR 48 mm |
| Thread length | 70 cm |
| Absorption mechanism | Hydrolysis |
| Absorption period | 56-72 days |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference (SKU) | 1L18HR4802 |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
Complete absorption by hydrolysis between 56 and 72 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a totally predictable absorption rate. Indicated for tissue approximation, ligatures and ophthalmic surgery.
4. Equivalences and Cross-References
Polyglactin 910 (PGLA) is one of the most widely used synthetic absorbable materials on the European market. This reference is clinically comparable, in terms of composition, absorption profile and tension retention, to braided multifilament sutures from other manufacturers:
| Manufacturer / Company | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic / Covidien | Absorbable braided polyglactin 910 range |
| Laboratory Aragó | Polyglactin 910 range with 1/2 circle needles (HR series) |
Within the Cahnos catalogue, its twin in structure and geometry with violet thread of 90 cm is reference 1L13HR4802, and with colourless thread 1L23HR4802. The alternative in polyglycolic acid (PGA) with the same needle is AM113HR4802 (90 cm) or AM118HR4802 (70 cm). The indicated equivalences are for guidance; for a formal substitution of references, consult our technical-commercial department.
5. Selection tips and variants of interest
To choose the most suitable suture, consider the wound support time required by the tissue and the necessary gauge according to the anatomical area. The USP 2 gauge with HR 48 mm needle is a common choice for deep planes, fascia, aponeurosis, high-tension closures and ligatures, with an excellent balance between initial strength and predictable absorption for soft tissues with healing of 2 to 4 weeks.
- 1L13HR4802: PGLA ½ circle cylindrical HR 48 mm, USP 2, violet, 90 cm (same geometry, longer thread length).
- 1L23HR4802: PGLA ½ circle cylindrical HR 48 mm, USP 2, colourless, 90 cm (same geometry, maximum aesthetic discretion).
- 1L18HR4502: PGLA ½ circle cylindrical HR 45 mm, USP 2, violet, 70 cm (intermediate needle).
- 1L18HR4002 / 1L13HR4002: PGLA cylindrical HR 40 mm, USP 2, violet, 70 cm / 90 cm (shorter needle for reduced fields).
- 1L13HRT4820: version with trocar point (tapercut) for somewhat denser tissues.
- 1L13HS4802: version with reverse cutting triangular point (HS 48 mm) for firm tissues.
- AM113HR4802 / AM118HR4802: alternative in polyglycolic acid (PGA) with absorption of 60-90 days and the same needle geometry.
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply for your healthcare centre. For longer force retention needs (180-210 days), the polydioxanone (PDO) family is the recommended alternative. For special B2B conditions and volume orders, contact our technical-commercial team.