Description
The Absorbable Multifilament PGLA Suture (Poliglactin 910) PREMISUT, reference 1L12HR1750, is the reference synthetic absorbable solution for the professional who demands predictable behaviour, excellent tensile strength and great knot security. This presentation incorporates a ½ circle needle with cylindrical body and conical Taperpoint point of 17 mm (HR17), USP 5/0 gauge (metric 1) and violet thread of 75 cm.
It is the ideal choice for the approximation of soft tissues, ligatures and ophthalmic surgery, in a professional format of a box of 12 units, ideal for stock management in operating theatres, ophthalmic clinics and purchasing centres, with special B2B conditions for hospitals and volume buyers.
Technical explanation of the material and procedure
Poliglactin 910 (PGLA) is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a grooved surface. This configuration provides a regular surface that favours the sliding of the thread through tissue, smooth handling, low memory and firm, secure knot tying, even in fine gauges such as USP 5/0.
Its absorption occurs by hydrolysis, that is, by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days and independent of the patient’s enzymatic activity. During the first weeks of healing it maintains the necessary tensile strength —75% at 2 weeks, 50% at 3 and 30% at 4—, the precise time for the tissue to recover its own strength.
The ½ circle needle with cylindrical body and conical point (Taperpoint geometry) combines an atraumatic point, which separates the fibres without cutting effect, with a smooth passage through tissue: it is the usual configuration in ophthalmic surgery and in the approximation of delicate tissues. The violet colour of the thread guarantees optimal visibility on the surgical field, highly valued in deep planes and in microsurgery. The product is sterilised by ethylene oxide (EO), is single-use and does not allow re-sterilisation.
Technical specifications
Technical data sheet for reference 1L12HR1750
| Reference (SKU) | 1L12HR1750 |
| Brand / Quality | PREMISUT – Standard |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament (grooved) |
| USP / metric gauge | 5/0 (metric 1) |
| Thread colour | Violet |
| Thread length | 75 cm |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Conical (Taperpoint), atraumatic |
| Needle size | 17 mm (HR17) |
| Absorption mechanism | Hydrolysis (56–72 days) |
| Sterilisation | Ethylene oxide (EO), sterile, single-use |
| Tissue reaction | Minimal |
| Indications | Approximation of soft tissues, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units (minimum order: 30 boxes) |
Tensile strength retention and absorption profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 30% of initial tensile strength |
| Total absorption | 56–72 days, by hydrolysis |
Predictable and controlled absorption profile, with minimal tissue reaction, excellent tensile strength and optimal knot tying. Price per box of 12 units.
Equivalences and cross-references
Poliglactin 910 is one of the most well-established multifilament synthetic absorbable suture raw materials on the market. This reference is clinically equivalent, in chemical composition, absorption profile (56–72 days) and tensile strength retention, to the main international lines:
| Manufacturer / commercial house | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (braided polyglactin 910, violet) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic (Covidien) | Polysorb (absorbable braided range of polyglactin 910) |
| SMI | Surgicryl (braided polyglactin 910) |
| Assut | Assucryl 910 (polyglactin 910) |
| DemeTECH | Vitacryl (coated braided polyglactin 910) |
Within the Cahnos catalogue, its closest equivalent by needle profile and gauge is reference P1L12HR1750 (same configuration in Prime quality), as well as the version with colourless thread 1L28HR1750 and the high-visibility needle version P1L18HR1750VB. The equivalences indicated are for guidance only; for a formal substitution of references, please consult our technical-commercial department.
Selection tips and variants of interest
The USP 5/0 gauge (metric 1) is indicated for very fine and low-tension tissues: mucosa, subcutaneous tissue, delicate planes and precision approximations, with special application in ophthalmic surgery, where aesthetic healing and minimal trauma are sought. The ½ circle needle with conical point of 17 mm is the most atraumatic option for these tissues.
In the same PREMISUT family, variants are available by gauge: 1L12HR1740 (USP 4/0), 1L12HR1730 (USP 3/0) and 1L12HR1720 (USP 2/0), all with HR 17 mm needle and 75 cm violet thread; and by needle, reference 1L12HR1350 with HR 13 mm needle for smaller radius accesses. If you prefer colourless thread, reference 1L28HR1750 maintains the same gauge and needle profile.
As a material alternative, the PGA family (polyglycolic acid), references such as AM112HR1750, offers similar performance with absorption of 60–90 days. If you need a longer retention profile (180–210 days) or monofilament thread, the polydioxanone (PDO) family is the recommended alternative. All references are supplied in boxes of 12 units; the minimum order is 30 boxes to guarantee continuous supply to your healthcare centre. As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B conditions by volume or equivalences from other commercial houses, contact our technical-commercial team.





Reviews