Description
1. Introduction
The absorbable multifilament polyglactin 910 (PGLA) suture, reference 1L12HRN4010, from the Premisut range by Cahnos, is the ideal surgical closure solution for professionals who demand a flexible, secure thread with a fully predictable absorption profile. Its ½ circle needle with cylindrical body and 40 mm blunt point (HRN 40) penetrates tissues while minimising trauma, making it especially suitable for approximating soft tissues, ligatures and ophthalmic surgery.
It is a synthetic absorbable medical device, sterile and single-use, with an excellent price-quality ratio for B2B procurement by hospitals, ophthalmic clinics, operating theatres and purchasing centres. It is supplied in a box of 12 units with a minimum order of 30 boxes, ensuring continuous and traceable supply for your healthcare facility.
2. Technical explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a grooved finish. This configuration provides great flexibility, low shape memory and a contained friction coefficient that favours thread sliding, manageability and firm, secure knot tying, with high initial tensile strength.
Absorption occurs by hydrolysis, i.e. by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. The thread retains 75% of tensile strength at 2 weeks, 50% at 3 weeks and 30% at 4 weeks, while the tissue recovers its own resistance; an ideal profile for tissues requiring support during 2 to 4 weeks of healing.
The ½ circle needle with cylindrical body and 40 mm blunt (atraumatic) point penetrates tissues by separating the fibres without cutting them, which reduces the risk of tearing and makes it especially recommended in ophthalmic surgery and in the approximation of friable tissues. The violet thread ensures optimal visibility in the surgical field. Sterilisation by ethylene oxide; sterile single-use product with no possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference 1L12HRN4010
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Colour | Violet (also available colourless) |
| Gauge | USP 0 (metric 3.5) |
| Thread length | 75 cm |
| Needle shape | ½ circle |
| Needle body | Cylindrical |
| Needle tip | Blunt (atraumatic) |
| Needle size | HRN 40 mm |
| Absorption mechanism | Hydrolysis |
| Absorption period | 56-72 days |
| Sterilisation | Ethylene oxide (EO), single-use |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference (SKU) | 1L12HRN4010 |
3.2. Tensile strength retention profile
| Time after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
Complete absorption by hydrolysis between 56 and 72 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate. Price per box of 12 units.
4. Equivalences and Cross-References
Polyglactin 910 is the reference synthetic absorbable material in the European market. This suture is clinically equivalent in composition, absorption profile and mechanical behaviour to the braided polyglactin 910 lines of the main manufacturers:
| Manufacturer / Company | Equivalent line |
|---|---|
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil / Surgicryl (braided poliglactin 910, violet) |
| Medtronic / Covidien | Absorbable braided polyglactin 910 range |
| Laboratory Aragó | Polyglactin 910 range with cylindrical needles (HR / RN series) |
Within the Cahnos catalogue, its direct equivalent in the same configuration within the Prime range is reference P1L12HRN4010 (PGLA Prime, same HRN 40 mm blunt-point needle, USP 0, violet, 75 cm) and the alternative in polyglycolic acid with the same needle is reference AM112HRN4010 (PGA, HRN 40 mm, USP 0, violet, 75 cm, absorption 60-90 days). In a reinforced heavy version with 90 cm thread, reference 1L13HRN4010H completes the family. The indicated equivalences are for guidance; for a formal substitution, please consult our technical-commercial department.
5. Selection tips and variants of interest
To choose the most suitable suture, consider the wound support time required by the tissue and the necessary gauge according to the anatomical area. The USP 0 gauge (metric 3.5) is very versatile: suitable for subcutaneous tissue, superficial muscle planes, approximation of soft tissues and ophthalmic surgery procedures, with an excellent balance between initial strength and predictable absorption during 2 to 4 weeks of healing.
| Reference | Variant |
|---|---|
| 1L12HRN4010 | PGLA USP 0, blunt point HRN 40 mm, violet, 75 cm (current product) |
| 1L12HRN3020 | PGLA USP 2/0, blunt point HRN 30 mm, violet, 75 cm (smaller gauge and radius) |
| 1L12HR4810 | PGLA USP 0, conical point HR 48 mm, violet, 75 cm |
| 1L13HRN4010H | PGLA USP 0, reinforced blunt point HRN 40 mm, violet, 90 cm |
| 1L13HRN4810 | PGLA USP 0, blunt point HRN 48 mm, violet, 90 cm (larger needle for deep fields) |
| 1L13HRN4001H | PGLA USP 1, heavy blunt point HRN 40 mm, violet, 90 cm (higher tensile strength) |
| 1L13HRT4010 | PGLA USP 0, trocar point HRT 40 mm, violet, 90 cm (fibrous tissues) |
The blunt point (HRN) is the most atraumatic option for delicate tissues; the conical point (HR) facilitates penetration and the trocar point (HRT) is designed for somewhat denser tissues. For longer force retention needs (180-210 days), the polydioxanone (PDO) family is the recommended alternative. All references are supplied in boxes of 12 units; the minimum order is 30 boxes. For special B2B conditions and volume orders, contact our technical-commercial team.





Reviews