Description
1. Product Presentation
The absorbable multifilament suture of Polyglactin 910 (PGLA) from Cahnos, brand Premisut in Prime quality, is the synthetic resorbable solution designed for the professional who demands predictable behaviour, excellent tensile strength and great security in knot tying. Its ½ circle needle with cylindrical body and Ultra Plus conical point of 17 mm (HRP17) provides smooth and controlled penetration in soft tissues, minimising tissue trauma. It is the ideal choice for tissue approximation, ligatures and ophthalmic surgery, in a professional format of a box of 12 units, ideal for stock management in operating theatres, clinics and purchasing centres.
2. Technical explanation of the material and procedure
Polyglactin 910 (PGLA) is a synthetic copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure that provides great flexibility, smooth passage through tissue and excellent knot-tying behaviour. Its absorption occurs by hydrolysis over an interval of 56 to 72 days, allowing a progressive and predictable loss of mass, without depending on the patient’s enzymatic activity. It maintains a tensile strength retention of 75% at 2 weeks, 50% at 3 weeks and 30% at 4 weeks, a sufficient period for the healing of most soft tissues. The thread, violet in colour, guarantees optimal visibility on the surgical field. The ½ circle needle with cylindrical body and Ultra Plus conical point (designation HRP17) is especially indicated for medium and delicate consistency tissues, since its conical profile passes through with less resistance and without cutting effect on adjacent tissue, ideal for ophthalmic procedures and approximation of lax tissues.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference | P1L18HRP1740 |
| Brand / Quality | Premisut – Prime |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament |
| USP size | 4/0 (metric 1.5) |
| Thread length | 70 cm |
| Colour | Violet |
| Needle shape | 1/2 circle |
| Needle body | Cylindrical |
| Needle tip | Conical Ultra Plus (Taperpoint) |
| Needle size | 17 mm (HRP17) |
| Sterilisation | Ethylene oxide (EO) |
| Presentation | Box of 12 units |
| Minimum order | 10 boxes |
Strength and absorption profile
| Parameter | Value |
|---|---|
| Tensile strength at 2 weeks | 75 % |
| Tensile strength at 3 weeks | 50 % |
| Tensile strength at 4 weeks | 30 % |
| Total absorption by hydrolysis | 56 – 72 days |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
4. Equivalences and Cross-References
Polyglactin 910 (PGLA) is one of the most widespread synthetic absorbable sutures on the market, so this Premisut Prime reference is perfectly comparable from a clinical point of view with the Vicryl® line from Ethicon (Johnson \u0026 Johnson), with Safil® from B. Braun, with Polysorb® from Medtronic/Covidien and with the polyglactin 910 ranges from SMI (Surgicryl). As for the needle, the ½ circle format with cylindrical body and 17 mm conical point corresponds to the classic pattern of small-format taper needles used in ophthalmic surgery and delicate tissues. The Cahnos reference P1L18HRP1740 is positioned as a direct alternative in a box of 12 units format for tenders and hospital supply, facilitating homologation in operating theatre protocols.
5. Selection tips and variants of interest
For the same HRP17 needle profile with Ultra Plus conical point, Cahnos offers the variants P1L18HRP1710 (USP 0), P1L18HRP1720 (USP 2/0), P1L18HRP1730 (USP 3/0), P1L18HRP1750 (USP 5/0) and P1L18HRP1760 (USP 6/0), all with violet thread of 70 cm and box of 12 units. If colourless thread is preferred with identical needle geometry, gauge and length, the equivalent reference is P1L28HRP1740. For a shorter needle of the same profile, reference P1L18HRP1340 (Ultra Plus conical point of 13 mm, USP 4/0) is the ideal alternative. For fibrous tissues or those requiring greater penetration, consider the trocar point variant P1L18HRT1740; for maximum visibility of the point, the VisiBlack version P1L18HR1740VB; and for precision ophthalmic surgery, the double needle version P1L18HRP1730-2 (2×70 cm, USP 3/0). It is recommended to confirm gauge and needle configuration according to the specialty and type of tissue before placing the order. Consult your Cahnos advisor for minimum order conditions and B2B rates for purchasing centres and hospitals.




Reviews